Mich. Admin. Code R. 333.5717 - Quality control program

Rule 717.

(1) A CT facility shall establish and implement a quality control program under the supervision of the medical physicist. The documented program shall include evaluation of all of the following:
(a) Image quality.
(b) Patient radiation dose.
(c) Personnel radiation protection.
(d) Compliance with the provisions of this part.
(e) Ongoing quality control.
(2) Evaluations and tests shall be performed following written procedures and methods. Corrective action shall be taken and documented according to instructions provided by the medical physicist if the results of an evaluation or test fall outside the control limits.
(3) The medical physicist shall determine the frequency of each test and who may perform the test. An on-site CT radiologic technologist shall be identified to be responsible for the ongoing quality control testing. The tests shall be performed by this technologist or by other personnel qualified by training and experience following written procedures and methods under subrule (2) of this rule.