Mich. Admin. Code R. 333.5717 - Quality control program
Rule 717.
(1) A
CT facility shall establish and implement a quality control program under the
supervision of the medical physicist. The documented program shall include
evaluation of all of the following:
(a) Image
quality.
(b) Patient radiation
dose.
(c) Personnel radiation
protection.
(d) Compliance with the
provisions of this part.
(e)
Ongoing quality control.
(2) Evaluations and tests shall be performed
following written procedures and methods. Corrective action shall be taken and
documented according to instructions provided by the medical physicist if the
results of an evaluation or test fall outside the control limits.
(3) The medical physicist shall determine the
frequency of each test and who may perform the test. An on-site CT radiologic
technologist shall be identified to be responsible for the ongoing quality
control testing. The tests shall be performed by this technologist or by other
personnel qualified by training and experience following written procedures and
methods under subrule (2) of this rule.