Minn. R. agency 144, ch. 4731, MEDICAL USE OF RADIOACTIVE MATERIAL, pt. 4731.4424 - POSSESSION OF SEALED SOURCES AND BRACHYTHERAPY SOURCES; REQUIREMENTS
A. A licensee in
possession of any sealed source or brachytherapy source must follow the
radiation safety and handling instructions supplied by the
manufacturer.
B. A licensee in
possession of a sealed source must:
(1) test
the source for leakage before its first use unless the licensee has a
certificate from the supplier indicating that the source was tested within six
months before transfer to the licensee; and
(2) test the source for leakage at intervals
not to exceed six months or at other intervals approved by the NRC or an
agreement state in the sealed source and device registry.
C. To satisfy the leak test requirements
under item B, a licensee must measure the sample so that the leak test can
detect the presence of 0.005 microcurie (185 Bq) of radioactive material on the
sample.
D. A licensee must retain
leak test records according to part 4731.4504, subpart 1.
E. If the leak test reveals the presence of
0.005 microcurie (185 Bq) or more of removable contamination, the licensee
must:
(1) immediately withdraw the sealed
source from use and store, dispose, or cause it to be repaired according to
parts 4731.2000 to 4731.2950 and 4731.3000 to 4731.3175; and
(2) file a report within five days of the
leak test according to part 4731.4527.
F. A licensee need not perform a leak test
on:
(1) sources containing only radioactive
material with a half-life of less than 30 days;
(2) sources containing only radioactive
material as a gas;
(3) sources
containing 100 microcuries (3.7 MBq) or less of beta- or gamma-emitting
material or ten microcuries (0.37 MBq) or less of alpha-emitting
material;
(4) seeds of iridium-192
encased in nylon ribbon; or
(5)
sources stored and not being used. The licensee must, however, test each source
under this subitem for leakage before any use or transfer, unless it has been
leak-tested within six months before the date of use or transfer.
G. A licensee in possession of
sealed sources or brachytherapy sources, except for gamma stereotactic
radiosurgery sources, must conduct a semiannual physical inventory of all such
sources in the licensee's possession. The licensee must retain each inventory
record according to part 4731.4504, subpart 2.