Minn. R. agency 144, ch. 4731, MEDICAL USE OF RADIOACTIVE MATERIAL, pt. 4731.4468 - DOSIMETRY EQUIPMENT
Except for low dose-rate remote afterloader sources where the source output or activity is determined by the manufacturer, a licensee must have a calibrated dosimetry system available for use. To satisfy this requirement:
A licensee must have a dosimetry system available for spot check output measurements, if applicable. To satisfy this requirement, the system may be compared with a system that has been calibrated according to subpart 1. The comparison must have been performed within the previous year and after each servicing that may have affected system calibration. The spot check system may be the same system used to meet the requirement under subpart 1.
A licensee must retain a record of each calibration, intercomparison, and comparison according to part 4731.4517.