Minn. R. agency 144, ch. 4731, MEDICAL USE OF RADIOACTIVE MATERIAL, pt. 4731.4469 - TELETHERAPY UNITS; FULL CALIBRATION
A licensee authorized to use a teletherapy unit for medical use must perform full calibration measurements on each teletherapy unit:
To satisfy subpart 1, full calibration measurements must include determination of:
A licensee must use the dosimetry system described in part 4731.4468, subpart 1, to measure the output for one set of exposure conditions. The remaining radiation measurements required under subpart 2 may be made using a dosimetry system that indicates relative dose rates.
A licensee must make full calibration measurements required under subpart 1 according to published protocols accepted by nationally recognized bodies.
A licensee must mathematically correct the outputs determined in subpart 2, item A, for physical decay for intervals not exceeding one month for cobalt-60, six months for cesium-137, or at intervals consistent with one percent decay for all other nuclides.
Full calibration measurements required under subpart 1 and physical decay corrections required under subpart 5 must be performed by the authorized medical physicist.
A licensee must retain a record of each calibration according to part 4731.4518.