Minn. R. agency 144, ch. 4731, MEDICAL USE OF RADIOACTIVE MATERIAL, pt. 4731.4477 - TELETHERAPY AND GAMMA STEREOTACTIC RADIOSURGERY UNITS; FULL-INSPECTION SERVICING
Subpart 1.
Inspection and servicing required. A licensee must have each
teletherapy unit and gamma stereotactic radiosurgery unit fully inspected and
serviced during source replacement to assure proper functioning of the source
exposure mechanism and other safety components. The interval between each
full-inspection servicing must not exceed five years for each teletherapy unit,
and must not exceed seven years for each gamma stereotactic radiosurgery unit.
Subp. 2.
Qualified
inspectors. The inspection and servicing must be performed by persons
specifically licensed to do so by the commissioner, the NRC, or an agreement
state.
Subp. 3.
Record
retention.
A licensee must keep a record of the inspection and servicing according to part 4731.4524.
Notes
Statutory Authority: MS s 144.1202; 144.1203