Minn. R. agency 144, ch. 4731, MEDICAL USE OF RADIOACTIVE MATERIAL, pt. 4731.4513 - BRACHYTHERAPY SOURCE CALIBRATION RECORDS
A licensee must maintain a record of the calibrations of brachytherapy sources required under part 4731.4455 for three years after the last use of the source. The records must include:
A. the date of the calibration;
B. the manufacturer's name, the model number
and serial number for the source, and instruments used to calibrate the
source;
C. the source output or
activity;
D. the source positioning
accuracy within the applicators; and
E. the signature of the authorized medical
physicist.