Minn. R. agency 144, ch. 4731, MEDICAL USE OF RADIOACTIVE MATERIAL, pt. 4731.4518 - CALIBRATION RECORDS; TELETHERAPY, REMOTE AFTERLOADER, AND GAMMA STEREOTACTIC RADIOSURGERY UNITS
A licensee must maintain a record of the teletherapy unit, remote afterloader unit, and gamma stereotactic radiosurgery unit full calibrations required under parts 4731.4469 to 4731.4471 for three years. The record must include:
A. the date of
the calibration;
B. the unit
manufacturer's name;
C. the model
number and serial number of the teletherapy, remote afterloader, and gamma
stereotactic radiosurgery unit;
D.
the model number, serial number, and identity of the source;
E. the model number, serial number, and
identity of the source instruments used to calibrate the units;
F. the results and an assessment of the full
calibrations;
G. the results of the
autoradiograph required for low dose-rate remote afterloader units;
and
H. the signature of the
authorized medical physicist who performed the full calibration.