Minn. R. agency 144, ch. 4731, MEDICAL USE OF RADIOACTIVE MATERIAL, pt. 4731.4519 - SPOT CHECK RECORDS; TELETHERAPY UNITS

A. A licensee must retain a record of each periodic spot check for teletherapy units required under part 4731.4472 for three years. The record must include:
(1) the date of the spot check;
(2) the unit manufacturer's name;
(3) the model number and serial number for the teletherapy unit;
(4) the model number, serial number, and identity of the source;
(5) instruments used to measure the output of the teletherapy unit;
(6) an assessment of time linearity and constancy;
(7) the calculated on-off error;
(8) a determination of the coincidence of the radiation field and the field indicated by the light beam localizing device;
(9) the determined accuracy of each distance measuring and localization device;
(10) the difference between the anticipated output and the measured output;
(11) notations indicating the operability of each entrance door electrical interlock, electrical or mechanical stop, source exposure indicator light, and viewing and intercom system and doors;
(12) the name of the individual who performed the periodic spot check; and
(13) the signature of the authorized medical physicist who reviewed the record of the spot check.
B. A licensee must retain a copy of the procedures required under part 4731.4472, subpart 2, until the licensee no longer possesses the teletherapy unit.