15 Miss. Code. R. 22-5.10 - Quality Assurance Measures for Cannabis Testing Entities
5.10.1 The cannabis testing entity shall
develop and implement a Quality Assurance (QA) program to assure the
reliability and validity of the analytical data produced by the cannabis
testing entity. The QA program shall, at minimum, include a written QA manual
that addresses the following:
A. Quality
control procedures;
B. Cannabis
testing entity organization and employee training and responsibilities,
including good laboratory practice (GLP);
C. QA objectives for measurement
data;
D. Traceability of data and
analytical results;
E. Instrument
maintenance, calibration procedures, and frequency;
F. Performance and system audits,
G. Corrective action procedures;
H. Steps to change processes when
necessary;
I. Record retention and
document control;
J. Test procedure
standardization; and
K. Method
validation;
L. Chain of custody
protocols;
M. Premise and sample
security;
N. Sample handling,
including sample receipt, identification, rejection, storage and
destruction;
O. Contingency plans
for data that is not within control limits, or is otherwise unacceptable for
analysis; and
P. Disposal of
cannabis and laboratory waste.
5.10.2 The supervisory or management cannabis
testing entity employee shall annually review, amend if necessary, and approve
the QA program and manual both when they are created and when there is a change
in methods, testing entity equipment, or the supervisory or management testing
entity employee.
5.10.3 The
cannabis testing entity's standard operating procedures for testing methods
shall include the following:
A. The name of
the testing method;
B. A list of
all analytes used in the testing method;
C. The applicable matrix or
matrices;
D. Sample receipt and
acceptance;
E. Method
sensitivity;
F. Potential
interferences;
G. Analytical
instrument and equipment used;
H.
Consumable supplies, reagents, and standards;
I. Sample preservation and hold
time;
J. Type, frequency, and
acceptable criteria for quality control samples;
K. Type, frequency, and acceptable criteria
for calibration standards;
L.
Procedures for analyzing batch samples;
M. Data quality assessment and acceptance
criteria;
N. Calibration of
results; and,
O. Reagent solution
and reference material preparation.
P. Current step-by-step instructions with
sufficient detail to perform the assay to include equipment operation and any
abbreviated versions used by a testing analyst.
5.10.4 Each cannabis testing entity shall
maintain a consumables log or inventory for all reagents, reference standards
and media purchased and received. All reagents and reference standards,
including any working standards, must be:
A.
Labeled to indicate identity, batch number, date received or prepared,
expiration date, and where applicable, concentration or purity, and date
opened;
B. Stored under appropriate
conditions to minimize degradation or deterioration of the material;
C. Within their expiration or
re-qualification dates at the time of use; and,
D. Documented on records for each
analysis.
5.10.5 Each
cannabis testing entity shall calibrate and maintain its equipment as specified
below, and the calibration, verification and/or check and maintenance must be
documented.
A. Trend testing space
temperatures and humidity daily using NIST-certified temperature devices.
Record corrective action if temperatures are out-of-range.
B. Check autoclaves performance with
bioindicator monthly and use heat-indicating tape with each cycle.
C. Check automatic Pipettes or Micropipettors
and Pipette Tips dispensing accuracy and precision quarterly and calibrate
annually.
D. Check balances daily
with a documented zero before use and service and recalibrate
annually.
E. Inspect Biosafety
cabinet airflow with each use and have certified annually.
F. Clean blenders as required by manufacturer
after each homogenization of submitted cannabis or cannabis products.
G. Verify centrifuge speeds and temperatures
daily and have certified annually.
H. Calibrate conductivity monthly.
I. Trend freezer and refrigerator
temperatures daily using NIST-certified temperature devices. Record corrective
action if temperatures are out-of-range.
J. Inspect glassware for chemistry cannabis
testing entity with each use for cleanliness, chips, and etching with each use.
Use class A when specified by the approved method and keep certificate of
conformance per each piece of class A glassware. If class B or class A without
a certificate of conformance, perform verification check upon purchase or prior
to first use.
K. Inspect glassware
for microbiological cannabis testing entity with each use for cleanliness,
chips, and etching.
L. Trend
incubator temperatures daily using NIST-certified temperature devices. Record
corrective action if temperatures are out-of-range.
M. Trend water bath temperatures daily using
NIST-certified temperature devices. Record corrective action if temperatures
are out-of-range.
N. Trend Laminar
Flow Hoods daily and service annually.
O. Clean Microscope optics and stage daily
and check alignment with each use. Service annually.
P. Follow Microwave digestors manufacturer's
instructions.
Q. Check Muffle
furnaces temperature accuracy at least annually.
R. Standardize pH meters with at least 2
buffer solutions daily before use.
S. Check Spectrophotometers
wavelength.
T. Check Timers and
stop watches at least annually.
U.
Certify reference weights annually.
V. Follow Analytical Instrumentation
manufacturer's instructions for cleaning and maintenance and document all
cleaning, calibrations, maintenance, and repairs.
W. Maintain all service records for the life
of equipment.
5.10.6 The
cannabis testing entities shall develop, implement, and validate test methods
for the analyses of samples as follows:
A. To
the extent practicable, methods shall compart with the following
guidelines:
B. The Bacteriological
Analytical Manual (BAM), 2019, which is incorporated by reference, includes no
future editions or amendments, and is available at
https://www.fda.gov/food/laboratory-methods-food/bacteriological-analytical-manualbam;
C. AOAC Official Methods of Analysis, 21st
Edition, 2019, which is incorporated by reference, includes no future editions
or amendments, and is available at
https://www.aoac.org/official-methods-of-analysis-21st-edition-2019;
and
D. To the extent practicable,
methods shall be validated in accordance with the following guidelines:
1. AOAC - Appendix J: Guidelines for
Validation of Microbiological Methods for Food and Environmental Surfaces,
2012, which is incorporated by reference, includes no future editions or
amendments, and is available at http://www.eoma.aoac.org/app_j.pdf [File Link
Not Available];
2. AOAC - Appendix
K: Guidelines for Dietary Supplements and Botanicals, 2013, which is
incorporated by reference, includes no future editions or amendments, and is
available at http://www.eoma.aoac.org/app_k.pdf [File Link Not
Available];
3. ICH - Validation of
Analytical Procedures: Text and Methodology Q2(R1) 2005, which is incorporated
by reference, includes no future editions or amendments, and is available at
https://database.ich.org/sites/default/files/Q2_R1__Guideline.pdf [File Link
Not Available] or Unofficial version of the Rules in 9 A.A.C. 17, effective
September 8, 2022 Page 115
https://www.fda.gov/regulatory-information/search-fda-guidance-documents/q2-r1-validation-analytical-procedures-text-and-methodology.
E. Method validation should, at a
minimum, verify accuracy, precision, analytical sensitivity, analytical
specificity, limit of detection, limit of quantification, reportable range and
the identification of interfering substances.
F. Methods adopted from a matrix specific
standard method, inclusivity and exclusivity do not require a comprehensive
reassessment, provided that there were no modifications to the methods,
including, but not limited to, all of the following:
1. Referenced media.
2. Primers.
3. Probes.
4. Antibodies.
5. Critical chemistries that were not
modified.
6. Microbial methods
shall include environmental monitoring and quality control of all buffers,
media, primers, and incubators.
G. The licensed laboratory shall generate a
validation report for each test method. Each validation report shall include
the following information:
1. Instrument
calibration data, if any;
2. Raw
data, including instrument raw data scanned as a PDF, for each test method, if
any;
3. Cannabis reference
materials or certified reference material results;
4. Data and calculations pertaining to LOD
and LOQ determinations, if any;
5.
Quality Control Sample report;
6.
Worksheets, forms, pictures, or copies of laboratory notebook pages
H. The laboratory director shall
review, approve, sign, and date the validation report for each test
method.
I. Validations shall be
submitted to the agency for approval with an acceptable and graded external
proficiency test by a third party, where all required analytes are shown to
have passed.
J. Upon new test
methods or altered test methods being used in the laboratory, the new
validation report shall be submitted to the Department within 5 business
days.
Notes
State regulations are updated quarterly; we currently have two versions available. Below is a comparison between our most recent version and the prior quarterly release. More comparison features will be added as we have more versions to compare.
No prior version found.