The requirements
obligations of this rule do not apply to
radiation-generating equipment used for dental intraoral or panoral,
mammography, bone densitometry, computed tomography, fluoroscopy or spot film
imaging, and equipment used in radiation therapy.
(A) General purpose radiographic equipment
shall
will
meet the following equipment standards:
(1) A
means shall
will be provided for limiting the x-ray beam to the
image receptor and area of clinical interest;
(2) For radiographic equipment having a
variable x-ray field limitation device, the limitation device
shall
will
have means for independent stepless adjustment of both the length and width of
the x-ray field;
(3) For
radiographic equipment that employs a light field for visually defining the
perimeter of the x-ray field, the light source shall
will be
functional and the total misalignment of the edges of the visually defined
field with the respective edges of the x-ray field along either the length or
width of the visually defined field shall
will not exceed
two per cent of the source-to-image distance (SID);
(4) In addition to the
requirements
obligations of paragraphs (A)(1), (A)(2) and (A)(3) of
this rule stationary radiographic equipment having a variable x-ray field
limitation device
shall
will have:
(a) A
means provided to indicate when the axis of the x-ray beam is perpendicular to
the plane of the image receptor;
(b) A means provided to align the center of
the x-ray field with respect to the center of the image receptor to within two
per cent of the SID, when the x-ray beam is perpendicular to the plane of the
image receptor;
(c) A means to
indicate the SID to within two per cent. If it is a fixed SID, the distance
shall
will be
indicated with a permanent marking;
(d) The beam limiting device indicate
numerically the field size in the plane of the image receptor to which it is
adjusted and be accurate within two per cent of the SID; and
(e) Compliance measurements made to discrete
SID's and image receptor dimensions in common clinical use, or at any other
specific dimensions at which the beam-limiting device or its associated
diagnostic x-ray system is uniquely designed to operate;
(5) General purpose radiographic equipment
designed with only one image receptor size at a fixed SID
shall
will
have:
(a) A means to limit the field at the
plane of the image receptor to dimensions no greater than those of the image
receptor, and to align the center of the x-ray field with the center of the
image receptor within two per cent of the SID; or
(b) A means to align the field such that the
x-ray field at the plane of the image receptor shall
will not extend
beyond any edge of the image receptor;
(6) General purpose radiographic equipment
designed with multiple removable or selectable fixed apertures
shall
will
have:
(a) A means to limit the x-ray field in
the plane of the receptor so that such field does not exceed each dimension of
the image receptor by more than two per cent of the SID when the axis of the
x-ray beam is perpendicular to the plane of the image receptor;
(b) A means to align the center of the x-ray
field with the center of the image receptor to within two per cent of the SID
or for equipment uniquely designed where the beam axis is intended to be offset
from the center of the image receptor, the x-ray field at the plane of the
image receptor shall
will not extend beyond the image receptor;
and
(c) Paragraphs (A)(6)(a) and
(A)(6)(b) of this rule may be met with either:
(i) An assortment of removable,
fixed-aperture, beam limiting devices sufficient to meet the
requirement
obligation for each combination of image receptor size
and SID for which the unit is designed with each such device having clear and
permanent markings to indicate the image receptor size and SID for which it is
designed; or
(ii) A beam-limiting
device having multiple fixed apertures sufficient to meet the
requirement
obligation for each combination of image receptor size
and SID for which the unit is designed. Permanent, clearly legible markings
shall
will
indicate the image receptor size and SID for which each aperture is designed
and shall
will
indicate which aperture is in position for use;
(7) If a positive beam limitation (PBL)
device is used, it
shall
will meet the following additional
requirements
obligations:
(a) The
PBL
shall
will
prevent the production of x-rays when any of the following conditions are met:
(i) Either the length or width of the x-ray
field in the plane of the image receptor differs from the corresponding image
receptor dimensions by more than three per cent of the SID, except as permitted
by paragraph (A)(7)(c) of this rule;
(ii) The sum of the length and width
differences as stated in paragraph (A)(7)(a)(i) of this rule without regard to
sign exceeds four per cent of the SID; or
(iii) The beam-limiting device is at an SID
for which PBL is not designed for sizing;
(b) Compliance with paragraph (A)(7)(a) of
this rule
shall
will be determined:
(i) When the equipment indicates that the
beam axis is perpendicular to the plane of the image receptor; and
(ii) No sooner than five seconds after
insertion of the image receptor;
(c) The PBL system shall
will be capable
of operation, at the discretion of the operator, such that the size of the
field may be made smaller than the size of the image receptor through stepless
adjustment of the field size. The minimum field size at a SID of one hundred
centimeters shall
will be equal to or less than five centimeters by five
centimeters;
(d) The PBL system
shall
will be
designed such that if a change in image receptor does not cause an automatic
return to PBL function as described in paragraph (A)(7)(a) of this rule, then
any change of image receptor size or SID must
will cause an
automatic return; and
(e) The PBL
system
shall
will function as described in paragraph (A)(7) of this
rule whenever all the following conditions are met:
(i) The image receptor is inserted into a
permanently mounted cassette holder;
(ii) The image receptor length and width are
less than fifty centimeters;
(iii)
The x-ray beam axis is within plus or minus three degrees of vertical in any
direction and the SID is ninety to one hundred thirty centimeters inclusive; or
the x-ray beam axis is within plus or minus three degrees of horizontal and the
SID is ninety to two hundred five centimeters inclusive;
(iv) The x-ray beam axis is perpendicular to
the plane of the image receptor to within plus or minus three degrees;
and
(v) Neither tomographic nor
stereoscopic radiography is being performed;
(8) A device shall
will be provided
to terminate the exposure at a preset time interval, preset product of current
and time, preset number of pulses, or preset radiation exposure to the image
receptor;
(9) For radiographic
equipment that provides manual exposure control, the operator
shall
will be
able to terminate the exposure at any time unless:
(a) The exposure is 0.5 second or less;
or
(b) During serial radiography,
means are provided to permit completion of any single exposure of the series in
progress;
(10) In the
case of radiographic equipment that provides automatic exposure control:
(a) The control panel
shall
will
indicate when this mode of operation is selected;
(b) The density setting and automatic
exposure control detector positions that are selected prior to the exposure
shall
will be
indicated; and
(c) A visible signal
shall
will
indicate when an exposure has been terminated at the back-up limit. Manual
resetting shall
will be required
obligated
before further automatic timed exposures can be made;
(11) The x-ray control panel
shall
will
provide visual indication when x-rays are produced and an audible signal
shall
will
indicate when the exposure has terminated;
(12) The exposure control switch
shall
will
meet the following
requirements
obligations:
(a) The
switch shall
will be a "dead-man switch;"
(b) It shall
will not be
possible to initiate an exposure when the timer is set to the "zero" or "off"
position if either position is provided; and
(c) The switch shall
will be
permanently mounted in a protected area so that it cannot be operated outside
the protected area except for portable, mobile, or veterinary
radiation-generating equipment.
(B) In addition to the applicable quality
assurance
requirements
obligations of Chapter 3701:1-66 of the Administrative
Code, handlers of general purpose radiographic equipment
shall
will
comply with the following:
(1) The kilovoltage
peak (kVp) accuracy shall
will be within plus or minus ten per cent of the
indicated value;
(2) The accuracy
of the timing device shall
will be within plus or minus ten per cent of the
indicated setting. The timing device shall
will be tested
at a minimum of two settings within the operative range of fifty milliseconds
to one thousand milliseconds;
(3)
The coefficient of variation:
(a) Of the kVp
reproducibility for at least four consecutive exposures
shall
will not
exceed 0.05;
(b) Of the timing
device reproducibility for at least four consecutive exposures
shall
will not
exceed 0.05; and
(c) Of radiation
exposure reproducibility for at least four consecutive exposures
shall
will not
exceed 0.05 for any specific combination of selected technique
factors;
(4) For
radiographic equipment having independent selection of x-ray tube current (mA),
the average ratios of exposure to the indicated mA-seconds (mAs) product
obtained at any two consecutive tube current settings
shall
will not
differ by more than ten per cent of their sum;
(5) For radiographic equipment having a
combined x-ray tube current-exposure time product, or mAs selector, but not a
separate tube current, or mA selector, the average ratios of exposure to the
indicated milliampere-seconds product (milligray/mAs) values obtained at any
two consecutive mAs selector settings shall
will not differ
by more than ten per cent of their sum;
(6) The average exposure ratio for paragraphs
(B)(4) and (B)(5) of this rule
shall
will be expressed as follows:
|X1 - X2| <
0.10(X1+X2)
Where the value of X1 and
X2 are the average milligray/mAs values obtained at each
of the two consecutive tube mA or mAs settings, or at two settings differing by
no more than a factor of two where the mA or mAs selector provides continuous
selection.
(C)
In addition to the applicable radiation safety rules in Chapter 3701:1-38 and
rules
3701:1-66-02 and
3701:1-66-04 of the
Administrative Code, the operator of general purpose radiographic equipment
shall
will
limit the useful beam to the area of clinical interest, not to exceed the size
of the image receptor by more than two per cent of the source-to-image
distance.
(D) Handlers of mobile or
portable radiation-generating equipment
shall
will not be
required
obligated to comply with the
requirements
obligations of paragraph (H) of rule
3701:1-66-02 of the
Administrative Code, and
shall
will comply with the following:
(1) Mobile and portable radiation-generating
equipment which are:
(a) Used continuously for
greater than one week in the same location, such as a room or suite,
shall
will
have the x-ray control permanently mounted in
a
behind a protected
area
barrier
so that the operator is required to
can remain in
that
entirely behind the protected
area
barrier and
see the patient without leaving the protective barrrier during the entire
exposure; or
(b) Used for less than
one week at the same location shall
will be provided with either a protective barrier at
least 6.5 feet high for operator protection that allows
the operator to see the patients without leaving the protective barrier
during exposures, or means shall
will be provided to allow the operator to be at least
six feet from the tube housing assembly during exposures and the operator
shall
will
wear a protective apron of not less than 0.25 millimeter lead equivalent when
making exposures;
(2)
Radiation emitted from the x-ray tube when a capacitor energy storage system is
fully charged and the exposure switch, timer, or any discharge mechanism is not
activated shall
will not exceed an air kerma of 0.26 microgray in one
minute at five centimeters from any accessible surface of the diagnostic source
assembly, with the beam-limiting device fully open; and
(3) A tube stand or other mechanical support
shall
will be
used so that the x-ray tube housing assembly shall
will not be
hand-held during exposures.
(E) Handlers of stationary veterinary
radiation-generating equipment
shall
will not be
required
obligated to
comply with the
requirements
obligations of paragraph (H)(4) of rule
3701:1-66-02 of the
Administrative Code. However, stationary veterinary radiation-generating
equipment
shall
will be provided with either a 6.5 foot high
protective barrier for operator protection during exposures, or
shall
will be
provided with means to allow the operator to be at least six feet from the tube
housing assembly during exposures. If the operator or assistant is not behind
the protective barrier, a lead apron of not less than 0.25 millimeter lead
equivalent
shall
will be worn when making exposures.
(F) Handlers of mobile or portable veterinary
radiation-generating equipment
shall
will not be
required
obligated to
comply with the
requirements
obligations of paragraph (H) of rule
3701:1-66-02 of the
Administrative Code. However, mobile or portable veterinary
radiation-generating equipment
shall
will be provided with either a 6.5 foot high
protective barrier for operator protection during exposures, or
shall
will be
provided with means to allow the operator to be at least six feet from the tube
housing assembly during exposures. If the operator or assistant is not behind
the protective barrier, a lead apron of not less than 0.25 millimeter lead
equivalent
shall
will be worn when making exposures.
(G) Handlers of certified
veterinary radiation-generating equipment
specifically designed by the manufacturer to be hand-held during radiographic
exposures
shall
will not be
required
obligated to
comply with the
requirements
obligations of paragraph (H) of rule
3701:1-66-02 of the
Administrative Code. The handler
shall
will develop and implement safe operating procedures
as part of the quality assurance program specified in rule
3701:1-66-04 of the
Administrative Code, which
shall
will address and document at least the following:
(1) Hand-held radiation-generating equipment
shall
will be
used for intraoral, extremity or small animal purposes only;
(2) Examination specific source-to-image
distances shall
will be developed and implemented to assure the useful
beam is limited to the area of clinical interest or no larger than the image
receptor;
(3) Operators of the
hand-held radiation-generating equipment and individuals participating in the
x-ray procedure shall
will be protected from direct scatter radiation by
protective aprons of not less than 0.25 millimeter lead equivalent
material;
(4) If the hand-held
radiation-generating equipment is designed with a back scatter shield, the
backscatter shield shall
will be in place during all radiographic
exposures;
(5) Storage and security
procedures shall
will be developed and implemented to assure hand-held
radiation-generating equipment is secured against unauthorized use or removal
when not under the control and constant surveillance of the handler;
(6) Hand-held radiation generating equipment
shall
will not
be used in hallways or waiting rooms; and
(7) Operator training
shall
will
include documented specific instruction to the x-ray operator regarding:
(a)
The
prohibition on
Not placing any part of
their body into the useful beam unless protected by not less than 0.5
millimeter lead equivalent material;
(b) The proper use of source-to-image
distance for the examination to assure the size of the radiation beam is no
larger than necessary;
(c)
Ensure
Ensuring that all individuals
required
obligated for the examination are wearing the
appropriate lead equivalent aprons and no bystanders are in the
vicinity;
(d) The areas of use,
proper storage and security procedures for the hand-held radiation-generating
equipment; and
(e) The use of the
dead-man switch and software safety devices such as locks and
sensors.
Notes
Ohio Admin. Code
3701:1-66-05
Effective:
12/13/2024
Five Year Review (FYR) Dates:
9/25/2024 and
12/13/2029
Promulgated
Under: 119.03
Statutory
Authority: 3748.04
Rule
Amplifies: 3748.01,
3748.02,
3748.04,
3748.05,
3748.06,
3748.07,
3748.12,
3748.121,
3748.13,
3748.14,
3748.15,
3748.17,
3748.18,
3748.19,
3748.20,
3748.22,
3748.99
Prior
Effective Dates: 02/15/2001, 02/10/2006, 04/01/2012, 07/01/2014,
12/20/2019