Ohio Admin. Code 3701:1-66-07 - Medical fluoroscopic equipment
(A) Fluoroscopic equipment
shall
will
meet the following standards:
(1) Unless the
United States food and drug administration (FDA) has granted a variance for
specific fluoroscopic equipment, the source-to-skin distance (SSD) for
fluoroscopy equipment shall
will not be less
than:
(a) Thirty-eight centimeters on
stationary fluoroscopic equipment unless a particular procedure application
prohibits
prevents that distance, in which case the SSD
shall
will not
be less than twenty centimeters;
(b) Thirty centimeters on mobile fluoroscopic
equipment unless a particular procedure prohibits
prevents
that distance, in which case it shall
will not be less than twenty centimeters;
and
(c) Nineteen centimeters for
c-arm type fluoroscopic equipment having a maximum source-to-image distance
(SID) less than forty-five centimeters unless a particular procedure
prohibits
prevents that distance, in which case it
shall
will not
be less than ten centimeters. Such systems shall
will be used for
extremity or dental purposes only;
(2) For c-arm fluoroscopic equipment equipped
with a removable spacer cone, the spacer cone shall
will be attached
to the x-ray source during use at all times unless it interferes with the
clinical procedure;
(3) The
fluoroscopic imaging assembly shall
will be provided with a primary protective barrier
which intercepts the entire cross-section of the useful beam at any SID and
shall
will
prevent further exposures when the primary barrier is not in the path of the
entire x-ray beam;
(4) All
fluoroscopic equipment shall
will provide intensified imaging. As used in this rule
"intensified imaging" will include the use of digital image
receptors;
(5) Fluoroscopic
equipment shall
will meet the following field limitation
specifications:
(a) For fluoroscopic equipment
manufactured before June 10, 2006, the following applies:
(i) Neither the length nor the width of the
x-ray field in the plane of the image receptor shall
will exceed that
of the visible area of the image receptor by more than three per cent of the
SID. The sum of the excess length and the excess width
shall
will be
no greater than four per cent of the SID; and
(ii) For rectangular x-ray fields used with
circular image receptors, the error in alignment shall
will be
determined along the length and width dimensions of the x-ray field which pass
through the center of the visible area of the image receptor;
(b) For fluoroscopic equipment
with a circular image receptor manufactured on or after June 10, 2006, the
maximum area of the x-ray field in the plane of the circular image receptor
shall
will
conform with one of the following requirements
obligations:
(i) When
any linear dimension of the visible area of the image receptor measured through
the center of the visible area is less than or equal to thirty-four centimeters
in any direction, at least eighty per cent of the area of the x-ray field
shall
will
overlap the visible area of the image receptor; or
(ii) When any linear dimension of the visible
area of the image receptor measured through the center of the visible area is
greater than thirty-four centimeters in any direction, the x-ray field measured
along the direction of greatest misalignment with the visible area of the image
receptor shall
will not extend beyond the edge of the visible area of
the image receptor by more than two centimeters;
(c) For fluoroscopic equipment with a
rectangular image receptor manufactured on or after June 10, 2006, the
following applies:
(i) Neither the length nor
the width of the x-ray field in the plane of the image receptor
shall
will
exceed that of the visible area of the image receptor by more than three per
cent of the SID. The sum of the excess length and the excess width
shall
will be
no greater than four per cent of the SID; and
(ii) The error in alignment
shall
will be
determined along the length and width dimensions of the x-ray field which pass
through the center of the visible area of the image receptor;
(d) If the fluoroscopic x-ray
field size is adjusted automatically as the SID or image receptor size is
changed, a capability may be provided for overriding the automatic adjustment
in case of system failure. If it is so provided, a signal visible at the
operator's position shall
will indicate whenever the automatic field adjustment
is overridden. Each such system failure override switch
shall
will be
clearly labeled as follows:
"For X-ray Field Limitation System Failure;"
(e) Beam-limiting devices
shall
will be
provided with a means for stepless adjustment of the x-ray field; and
(f) Stepless adjustment
shall
will, at
the greatest SID, provide continuous field sizes from the maximum obtainable to
a field size of five centimeters by five centimeters or less;
(6) Timers
shall
will
meet the following specifications:
(a) A means
shall
will be
provided to preset the cumulative on-time timer of the fluoroscopic tube. The
maximum cumulative time of the timer shall
will not exceed
five minutes without resetting;
(b)
The timer shall
will terminate the exposure or emit a signal audible
to the operator when the exposure time reaches a maximum of five minutes. The
signal shall
will continue to sound while x-rays are produced until
the timer is reset;
(c) For x-ray
controls manufactured on or after June 10, 2006, there
shall
will be
provided for each fluoroscopic tube:
(i) A
display of the fluoroscopic irradiation time at the operator's working
position. This display shall
will function independently of the audible signal
described in paragraph (A)(6)(c)(ii) of this rule. The following
requirements
obligations apply:
(a) When the x-ray tube is activated, the
fluoroscopic irradiation time in minutes and tenths of minutes
shall
will be
continuously displayed and updated at least once every six seconds;
(b) The fluoroscopic irradiation time
shall
will
also be displayed within six seconds of termination of an exposure and remain
displayed until reset; and
(c)
Means shall
will be provided to reset the display to zero prior to
the beginning of a new examination or procedure;
(ii) A signal audible to the operator
shall
will
sound for each passage of five minutes of fluoroscopic irradiation time during
an examination or procedure. The signal shall
will sound until
manually reset or, if automatically reset, for at least two seconds;
(7) X-ray production in
the fluoroscopic mode shall
will be
controlled by a device which requires continuous pressure by the operator for
the entire time of any exposure. When recording serial fluoroscopic images, the
operator shall
will be able to terminate the x-ray exposure at any
time, but means may be provided to permit completion of any single exposure of
the series in progress;
(8)
Fluoroscopic systems shall
will meet the following air kerma rate limits:
(a) Fluoroscopic equipment provided with only
automatic exposure rate control, or provided with both automatic exposure rate
control and manual mode capabilities, shall
will not exceed
an air kerma rate of eighty-eight milligray per minute (ten roentgens per
minute exposure rate) in either mode at any combination of tube potential and
current, at the measurement point specified in paragraph (C)(6) of this
rule;
(b) Fluoroscopic equipment
provided with only manual mode capabilities shall
will not exceed
an air kerma rate of forty-four milligray per minute (five roentgens per minute
exposure rate) at any combination of tube potential and current, at the
measurement point specified in paragraph (C)(6) of this rule; and
(c) For fluoroscopic equipment that is
provided with high-level control, and the high-level control is activated, the
air kerma rate shall
will not exceed one hundred seventy-six milligray per
minute (twenty roentgens per minute exposure rate) at any combination of tube
potential and current, at the measurement point specified in paragraph (C)(6)
of this rule;
(i) For all fluoroscopy
equipment that is provided with high-level control, special means of activation
of high level control, such as manual pressure applied continuously by the
operator, shall
will be required
needed to
avoid accidental use; and
(ii) A
continuous signal audible to the operator shall
will indicate
that high level control is being employed;
(9) During fluoroscopy and cinefluorography
the x-ray tube potential and current shall
will be
continuously indicated;
(10) For
undertable fluoroscopic equipment, a shielding device of at least 0.25
millimeter lead equivalent shall
will cover the bucky-slot;
(11) For undertable fluoroscopic equipment,
protective drapes, or other devices, at least 0.25 millimeter lead equivalent
shall
will be
provided between the patient and the individual operating the fluoroscopic
equipment to intercept scattered radiation which would otherwise reach the
fluoroscopist and others near the x-ray unit, except when such drapes or other
devices would compromise the sterile field. Such devices
shall
will not
substitute for wearing required
obligated protective apparel;
(12) Radiography using the fluoroscopic
imaging assembly shall
will meet the following specifications:
(a) A means shall
will be provided
between the source and the patient which will automatically limit the x-ray
field at the time the exposure is initiated to no more than the portion of the
image receptor selected by the operator for spot films or radiographic images.
If the x-ray field size is less than the size of the selected portion of the
image receptor, the field size shall
will not open automatically to the size of the
selected portion of the image receptor unless the operator has selected such a
mode of operation;
(b) Neither the
length nor the width of the x-ray field in the plane of the image receptor
shall
will
differ from the corresponding dimensions of the selected portion of the image
receptor by more than three per cent of the SID when adjusted for full coverage
of the selected portion of the image selector;
(c) The center of the x-ray field in the
plane of the image receptor shall
will be aligned with the center of the selected
portion of the image receptor to within two per cent of the SID; and
(d) Means shall
will be provided
to reduce the x-ray field size in the plane of the image receptor to a size
smaller than the selected portion of the image receptor. The minimum field size
at the greatest SID shall
will not exceed five centimeters by five
centimeters;
(13)
Fluoroscopic equipment manufactured on or after June 10, 2006,
shall
will
display at the operator's working position the air kerma rate (AKR) and
cumulative air kerma in accordance with the following
requirements
obligations:
(a) When
the x-ray tube is activated and the number of images produced per unit time is
greater than six images per second, the AKR in milligrays per minute
shall
will be
continuously displayed and updated at least once every second;
(b) The cumulative air kerma in units of
milligrays shall
will be displayed either within five seconds of
termination of an exposure or displayed continuously and updated at least once
every five seconds;
(c) The display
of the AKR shall
will be clearly distinguishable from the display of
the cumulative air kerma;
(d) The
AKR and cumulative air kerma shall
will represent the value for conditions of free-in-air
irradiation at one of the following reference locations specified according to
the type of fluoroscope;
(i) For fluoroscopes
with x-ray source below the x-ray table, x-ray source above the table, or of
lateral type, the reference location shall
will be the
respective locations specified in paragraph (C)(6)(a), (C)(6)(b) or (C)(6)(d)
of this rule; or
(ii) For C-arm
fluoroscopes, the reference location shall
will be fifteen
centimeters from the isocenter toward the x-ray source along the beam axis.
Alternatively, the reference location shall
will be at a
point specified by the manufacturer to represent the location of the
intersection of the x-ray beam with the patient's skin;
(e) Means shall
will be provided
to reset to zero the display of cumulative air kerma prior to the commencement
of a new examination or procedure; and
(f) The displayed AKR and cumulative air
kerma shall
will not deviate from the actual values by more than
plus or minus thirty-five per cent;
(14) Fluoroscopic equipment manufactured on
or after June 10, 2006 shall
will be equipped with means to display a last image
hold (LIH) image following termination of the fluoroscopic exposure:
(a) For a LIH image obtained by retaining
pre-termination fluoroscopic images, if the number of images and method of
combining images are selectable by the user, the selection
shall
will be
indicated prior to initiation of the fluoroscopic exposure;
(b) For a LIH image obtained by initiating a
separate radiographic exposure at termination of the fluoroscopic imaging, the
technique factors for the LIH image shall
will be
selectable prior to the fluoroscopic exposure, and the combination selected
shall
will be
indicated prior to initiation of the fluoroscopic exposure; and
(c) Means shall
will be provided
to clearly indicate to the user whether a displayed image is the LIH radiograph
or fluoroscopy. Display of the LIH radiograph shall
will be replaced
by the fluoroscopic image concurrently with re-initiation of the fluoroscopic
exposure unless separate displays are provided.
(B) In addition to other applicable radiation
safety rules adopted pursuant to Chapter 3748. of the Revised Code, handlers of
fluoroscopic radiation-generating equipment shall
will comply with
the following:
(1) Any individual who is in
the room during the fluoroscopic procedure shall
will be
adequately protected by standing behind a whole body protective barrier or
shall be required to
will wear a protective lead apron of not less than
0.25 millimeter lead equivalent. If a handler's
radiation expert includes documented evidence and specifies in the quality
assurance program a distance and time interval in the room from the source at
which an individual is unlikely to receive a total effective dose equivalent of
greater than two millirem in any one hour or one hundred millirem in a year,
the handler can forgo the use of a protective barrier or lead apron for that
time at that distance;
(2)
Protective lead or lead equivalent gloves shall
will be used by
individuals who are required
obligated to have their hands in
or near the useful beam; and
(3) Handlers of fluoroscopic equipment used
for interventional or cardiac procedures or on pediatric or pregnant patients
shall
will
maintain a record of:
(a) Cumulative air kerma
or dose area product used for each examination, if the display of either is
available on the fluoroscopic equipment; or
(b) The following items if the cumulative air
kerma or dose area product is not displayed on the fluoroscopic equipment:
(i) Mode of operation such as high-level or
pulsed mode;
(ii) Cumulative
fluoroscopic exposure time; and
(iii) Number of radiographs and number of
acquisitions.
(C) In addition to other applicable quality
assurance requirements
obligations of Chapter 3701:1-66 of the Administrative
Code, handlers of fluoroscopic equipment shall
will comply with
the following:
(1) Handlers
shall
will
designate and utilize a radiation expert who shall
will develop in
writing and perform fluoroscopic image quality evaluations appropriate for the
fluoroscopic equipment including written procedures to include time intervals
and system conditions for the evaluation of image quality;
(2) On new installations or reinstallations
of existing equipment prior to patient exposure, handlers
shall
will
utilize a radiation expert to perform the following:
(a) Radiographic device tests to determine
compliance with allowable limits as specified in paragraph (A)(12) of this
rule;
(b) Fluoroscopic image
quality evaluations as specified in paragraph (C)(1) of this rule;
(c) Air kerma rate tests as specified in
paragraph (C)(6) of this rule;
(d)
High contrast and low contrast resolution evaluations in both fluoroscopic and
radiographic modes;
(e) Five minute
timer evaluations; and
(f)
Evaluation of the accuracy of technique factor indicators and integrated
radiation dose displays;
(3) After initial evaluations of fluoroscopic
equipment have been performed, the test and evaluations in paragraph (C)(2) of
this rule shall
will be performed by a radiation expert
annually;
(4) After repair or
replacement of any component of the fluoroscopic equipment which may alter the
radiation output or image quality, prior to patient use, a radiation expert
shall
will
perform and document measurements of air kerma rates as specified in paragraph
(C)(6) of this rule and image quality as specified in paragraph (C)(1) of this
rule unless in the documented determination of a radiation expert, the repair
or replacement will not cause a significant change in radiation output or
significant degradation of image quality as specified in the quality assurance
program;
(a) The radiation expert may
designate qualified individuals to perform and document the measurements
specified in paragraphs (C)(6) and (C)(1) of this rule;
(b) The radiation expert
shall
will
provide the criteria for qualifying these designees in the quality assurance
program; and
(c) The radiation
expert's approval of the designee's test results shall
will be
documented within thirty days;
(5) The results of all tests performed in
accordance with paragraphs (C)(2) to (C)(4) of this rule
shall
will:
(a) Include the technique factors used in
determining such results;
(b)
Include the name of the individual performing the measurements;
(c) Include the date the measurements were
performed; and
(d) Be maintained by
the IRRP between inspections for review by the department;
(6) Compliance with air kerma rate allowable
limits in paragraph (A)(8) of this rule shall
will be
determined as follows:
(a) If the source is
below the x-ray table, the air kerma rate shall
will be measured
at one centimeter above the tabletop or cradle;
(b) If the source is above the x-ray table,
the air kerma rate shall
will be measured at thirty centimeters above the
tabletop with the end of the beam limiting device or spacer positioned as
closely as possible to the point of measurement;
(c) For c-arm type fluoroscopic equipment,
the air kerma rate shall
will be measured at thirty centimeters from the input
surface of the image receptor with the source positioned at any SID;
(d) For fixed SID lateral fluoroscopes
attached to the x-ray table, the maximum air kerma rate
shall
will be
measured at a point fifteen centimeters from the centerline of the x-ray table
and in the direction of the x-ray source with the end of the beam-limiting
device or spacer positioned as closely as possible to the point of measurement.
If the table top is moveable, it shall
will be positioned as closely as possible to the
lateral x-ray source, with the end of the beam-limiting device or spacer no
closer than fifteen centimeters to the centerline of the table;
(e) For c-arm type fluoroscopic equipment
having a SID less than forty-five centimeters, the air kerma rate
shall
will be
determined at the minimum SSD; and
(f) The maximum air kerma rate
shall
will be
determined with the kVp, mA and/or other selectable parameters adjusted to
those settings which give the maximum air kerma rate. X-ray systems that
incorporate automatic exposure control shall
will have
sufficient attenuative material placed in the useful beam to produce the
maximum exposure rate of the system.
(D) Handlers of mobile fluoroscopic equipment
shall
will not
be required
obligated to comply with the
requirements
obligations of paragraphs (A)(10), and (A)(11) of this
rule and paragraph (H) of rule
3701:1-66-02 of the
Administrative Code.
(E) Handlers
of c-arm fluoroscopic equipment having a maximum SID less than forty-five
centimeters shall
will not be required
obligated to
comply with the requirements
obligations of paragraphs (A)(5)(e), (A)(5)(f),
(A)(10), (A)(11), and (A)(12) of this rule and paragraph (H) of rule
3701:1-66-02 of the
Administrative Code. In addition, if a radiation expert has specified in the
registrant's quality assurance program that an individual is unlikely to
receive a total effective dose equivalent of greater than two millirem in any
one hour or one hundred millirem in a year, the handler
shall
will not
be required
obligated to comply with the
requirements
obligations of paragraph (B)(1) of this
rule.
(F) All individuals operating
fluoroscopic equipment, and individuals likely to receive an annual effective
dose equivalent in excess of one millisievert (one hundred millirem) from
participating in fluoroscopic procedures, shall
will receive at
least two hours of radiation protection training specific to fluoroscopy in
addition to the training required
obligated by rule
3701:1-38-10 of the
Administrative Code prior to performing or participating in fluoroscopic
procedures. Additionally, each individual shall
will receive one
hour of re-training whenever the individual receives in excess of thirty per
cent of the allowable occupational dose measured over one calendar
year.
(G) The training
required
obligated by paragraph (F) of this rule
shall
will be
approved by the registrant's designated radiation expert, and be specific to
the type of fluoroscopic equipment used. Documentation of receiving the
training required
obligated by paragraph (F) of this rule
shall
will be
retained by the registrant and be available for review upon inspection. At a
minimum, training topics shall
will include, but not be limited to:
(1) Principles and operation of the
fluoroscopic equipment to be used;
(2) Fluoroscopic and radiographic outputs of
each mode of operation, including high-level control options clinically
used;
(3) Dose management,
including dose reduction techniques for fluoroscopic equipment;
(4) Safe operating procedures of each piece
of fluoroscopic equipment that may be used by each individual.
(5) Units of measurement and dose, including
dose-area product values and air kerma;
(6) Radiation protection methods for patient
and staff;
(7) Basic properties of
radiation; and
(8) Biological
effects of radiation.
(H) Fluoroscopic equipment used for radiation
therapy procedures is regulated pursuant to rule
3701:1-67-09 of the
Administrative Code.
(I) Computed
tomography scanners equipped with fluoroscopic capabilities are regulated
pursuant to rule
3701:1-66-10 of the
Administrative Code.
Notes
Promulgated Under: 119.03
Statutory Authority: 3748.04
Rule Amplifies: 3748.01, 3748.02, 3748.04, 3748.05, 3748.06, 3748.07, 3748.12, 3748.121, 3748.13, 3748.14, 3748.15, 3748.17, 3748.18, 3748.19, 3748.20, 3748.22, 3748.99
Prior Effective Dates: 02/15/2001, 12/01/2003, 04/15/2009, 07/01/2014, 12/20/2019