Ohio Admin. Code 3701:1-67-05 - Standards for therapy equipment operating at less than one megavolt (MV)
As used in this rule, "therapy equipment" means therapy equipment operating below one megavolt (MV). In addition to the rules in Chapters 3701:1-38 and 3701:1-67 of the Administrative Code, handlers of therapy equipment shall comply with the following:
(A) When the x-ray tube is operated at its
maximum rated tube current for the maximum kilovoltage (kV), the leakage air
kerma rate shall not exceed the value specified at the distance specified for
that classification of therapy equipment.
(1)
For five kV to fifty kV systems, the leakage air kerma rate measured at any
position five centimeters from the tube housing assembly shall not exceed one
milligray (one hundred millirad) in any one hour.
(2) For greater than fifty kV and less than
five hundred kV
one
MV systems, the leakage air kerma rate measured at a distance of one
meter from the target in any direction shall not exceed one centigray (one rad)
in any one hour. This air kerma rate measurement may be averaged over areas no
larger than one hundred square centimeters. In addition, the air kerma rate at
a distance of five centimeters from the surface of the tube housing assembly
shall not exceed thirty centigrays (thirty rad) per hour.
(3) For each piece of therapy equipment, the
handler shall determine, or obtain from the manufacturer, the leakage radiation
existing at the positions specified in paragraphs (A)(1) and (A)(2) of this
rule for the specified operating conditions. Records on leakage radiation
measurements shall be maintained at the installation for inspection by the
department.
(B)
Permanent diaphragms or cones used for limiting the useful beam shall provide
at least the same degree of attenuation as required for the tube housing
assembly.
(C) Requirements for
adjustable or removable beam limiting devices include:
(1) All adjustable or removable beam limiting
devices, diaphragms, cones or blocks shall not transmit more than five per cent
of the useful beam for the most penetrating beam used; and
(2) When adjustable beam limiting devices are
used, the position and shape of the radiation field shall be indicated by a
light beam.
(D) The
filter system shall be so designed that:
(1)
Filters can not be accidentally displaced at any possible tube
orientation;
(2) An interlock
system prevents irradiation if the proper filter is not in place;
(3) The air kerma rate escaping from the
filter slot shall not exceed one centigray (one rad) per hour at one meter
under any operating conditions; and
(4) Each filter shall be marked as to its
material of construction and its thickness.
(E) Requirements for tube immobilization
include:
(1) The x-ray tube shall be so
mounted that it cannot accidentally turn or slide with respect to the housing
aperture; and
(2) The tube housing
assembly shall be capable of being immobilized for stationary portal treatments
unless the unit is designed to be hand-held and the peak tube potential of the
system does not exceed fifty kV.
(F) The tube housing assembly shall be so
marked that it is possible to determine the location of the source to within
five millimeters, and such marking shall be readily accessible for use during
calibration procedures.
(G) Contact
therapy tube housing assemblies shall have a removable shield of material,
equivalent in attenuation to 0.5 millimeters of lead at one hundred kV, which
can be positioned over the entire useful beam exit port during periods when the
beam is not in use.
(H) A suitable
irradiation control device shall be provided to terminate the irradiation after
a pre-set time interval.
(1) A timer with a
display shall be provided by the treatment control system. The timer shall have
a pre-set time selector and an elapsed time or time remaining
indicator;
(2) The timer shall be a
cumulative timer that activates with an indication of "beam-on" and retains its
reading after irradiation is interrupted or terminated. After irradiation is
terminated and before irradiation can be reinitiated, it shall be necessary to
reset the elapsed time indicator;
(3) The timer shall terminate irradiation
when a pre-selected time has elapsed, if any dose monitoring system present has
not previously terminated irradiation;
(4) The timer shall permit accurate
pre-setting and determination of exposure times as short as one
second;
(5) The timer shall not
permit an exposure if set at zero;
(6) The timer shall not activate until the
shutter is opened if irradiation is controlled by a shutter mechanism unless
calibration includes a timer error correction to compensate for mechanical lag;
and
(7) The timer shall be accurate
to within one per cent of the selected value or one second, whichever is
greater.
(I) The control
system, in addition to the displays required by other provisions in this rule,
shall have:
(1) An indication of whether
electrical power is available to the control system and if activation of the
x-ray tube is possible;
(2) An
indication of whether x-rays are being produced;
(3) A means for indicating x-ray tube
potential and current;
(4) The
means for terminating an exposure at any time;
(5) A locking device which will prevent
unauthorized use of the therapy equipment;
(6) A positive display of specific filter(s)
in the beam; and
(7) A warning
label which cautions individuals that radiation is produced when the therapy
equipment is energized.
(J) When a control system may energize more
than one x-ray tube:
(1) It shall be possible
to activate only one x-ray tube at any time;
(2) There shall be an indication at the
control panel identifying which x-ray tube is activated
selected to
enable irradiation; and
(3)
There shall be an indication at the tube housing assembly when that tube is
energized
selected
to enable irradiation.
(K) There shall be a means of determining the
central axis target-to-skin distance (TSD) to within one centimeter and of
reproducing this measurement to within two millimeters thereafter.
(L) Unless it is possible to bring the x-ray
output to the prescribed exposure parameters within five seconds after the
x-ray "on" switch is energized, the beam shall be attenuated by a shutter. In
addition, after the unit is at operating parameters, the shutter shall be
controlled by the operator from the control panel. An indication of shutter
position shall appear at the control panel.
(M) Therapy equipment having a beryllium or
other low-filtration window shall be clearly labeled as such upon the tube
housing assembly and shall be provided with a permanent warning device on the
control panel that is activated when no additional filtration is present, to
indicate that the dose rate is very high.
Notes
Promulgated Under: 119.03
Statutory Authority: 3748.04
Rule Amplifies: 3748.01, 3748.02, 3748.04, 3748.05, 3748.06, 3748.07, 3748.12, 3748.121, 3748.13, 3748.14, 3748.15, 3748.17, 3748.18, 3748.19, 3748.20, 3748.22, 3748.99
Prior Effective Dates: 02/15/2001, 09/01/2005, 01/14/2011, 06/01/2013, 03/01/2016