Ohio Admin. Code 3701-72-04 - Licensee authorizations and requirements
(A) Each radiologic licensee shall operate
ionizing radiation-generating equipment or use radiopharmaceuticals in
accordance with the documented quality assurance and radiation protection
programs required under the rules adopted pursuant to Chapter 3748. of the
Revised Code. The director may at any time during business hours evaluate
operator competency in regard to the safe operation of ionizing
radiationgenerating equipment or use of radioactive sources to determine
compliance with Chapter 4773. of the Revised Code or any rules adopted pursuant
to Chapter 3748. of the Revised Code.
(B) Radiographers, nuclear medicine
technologists, radiation therapists, and general x-ray machine operators who
are registered nurses providing occupational health nursing services in an
industrial workplace shall work under general supervision.
(C) General x-ray machine operators, other
than those registered nurses who provide occupational health nursing services
in an industrial workplace, shall work under the direct supervision of a
licensed practitioner.
(D) The
following radiologic procedure tasks shall be performed only by an Ohio
licensed practitioner, radiographer, nuclear medicine technologist, radiation
therapist or general x-ray machine operator within his/her scope of practice:
(1) Adjust or set technique
factors;
(2) Move or position
equipment;
(3) Move or position the
patient;
(4) Expose the patient to
radiation;
(5) Reset the audible
signal on the cumulative radiation exposure timer;
(6) Assure adequate radiation protection to
the patient and individuals in the procedure room from unnecessary radiation;
and
(7) Process and post process
digital image at the time of the procedure.
(E) Provided that all radiation safety
training required by rules adopted pursuant to Chapter 3748. of the Revised
Code have been completed, registered nurses, registered cardiovascular invasive
specialists, and general x-ray machine operators licensed prior to September
30, 1995, who have been continuously licensed in accordance with this chapter
to practice in the area of fluoroscopy pursuant to the conditional license
requirements of section
4773.05 of the Revised Code, may
perform, under the personal supervision of a licensed practitioner present in
the room, the following tasks of a cardiac catheterization fluoroscopic
procedure:
(1) Move or position the
patient;
(2) Move or position the
fluoroscopic unit; and
(3) Reset
the audible signal on the cumulative radiation exposure timer.
(F) General x-ray machine
operators licensed after August 1, 2008, shall complete one or as many of the
following competency-based clinical training modules from an accredited
competency-based clinical training program needed to meet the requirements to
work within the scope of practice of their supervising practitioner:
(1) Chest and abdomen,
(2) Extremities,
(3) Skull and sinus,
(4) Spine,
(5) Podiatric radiography, or
(6) Bone densitometry.
(G) Effective August 1, 2008, general x-ray
machine operators changing employment to a different licensed practitioner
within a new area of clinical practice, shall, prior to practicing in the new
area, complete the appropriate competency-based clinical training module as
required by paragraph (F) of this rule and submit evidence of completion of the
clinical training to the department.
(H) General x-ray machine operators initially
licensed prior to August 1, 2008, who want to perform standard, diagnostic
radiologic procedures using digital equipment, shall complete the corresponding
competency-based clinical training module as listed in paragraph (F) of this
rule and submit evidence of completion of the clinical training to the
department.
(I) A licensee shall
make his or her license available immediately upon the request of any
person.
(J) Each licensee shall
notify the department in writing within thirty days if there is a change in the
licensee's name, email or mailing
street address or email
address.
(K) Positron
emission tomography-computed tomography and single photon emission computed
tomography-computed tomography fusion procedures shall be performed by a
licensed nuclear medicine technologist or a licensed radiographer, who has obtained training and demonstrated
competency with the equipment and procedures. Radiopharmaceuticals shall be
injected only by a licensed nuclear medicine technologist or other competent
licensed individuals acting within their scope of practice as authorized by the
Revised Code. If the computed tomography scanner is
operated for any other diagnostic radiographic procedures, the computed
tomography scanner shall be operated by a licensed
radiographer.
(L)
Radiographers
Licensed radiographers or licensed radiation therapists, who have obtained
training and demonstrated competency with the computerized tomography (CT)
equipment and CT protocols, may operate CT units for radiation therapy
simulation procedures. CT simulation procedures shall be overseen by a
radiation therapist. If the CT equipment is operated
for any other diagnostic radiographic procedures, the CT equipment shall be
operated by a radiographer.
(M)
A licensed
nuclear medicine technologist or radiation therapist who is certified in
computed tomography by a national certifying agency approved by the director
may perform computed tomography procedures if the licensed nuclear medicine
technologist or radiation therapist has received the training described in
paragraphs (C)(1) and (C)(2) of rule
3701-72-03 of the Administrative
Code.
A licensed nuclear medicine technologist or radiation therapist may perform computed tomography procedures as part of a documented prerequisite clinical experience program leading to certification in computed tomography by a national certifying agency approved by the director provided:
(1)
The computed tomography procedures and program are
conducted under personal supervision of, and documented by an individual fully
qualified to perform computed tomography procedures; and
(2)
The computed
tomography clinical experience program is completed within twenty four months
from the initial start date of the documented prerequisite clinical experience
program.
(N)
A licensee shall perform computed tomography procedures
in accordance with acceptable professional standards.
(O)
A radiologic
licensee shall not expose an individual to the useful beam unless ordered by a
licensed practitioner, as defined in Chapter 4773. or 3748. of the Revised
Code, acting within his or her scope-of-practice for dental, medical or
radiation therapy purposes. Exposing an individual for training, demonstration
or other purposes is prohibited unless otherwise specified in rules promulgated
under Chapter 4773. or 3748. of the Revised Code.
Notes
Promulgated Under: 119.03
Statutory Authority: 4773.08
Rule Amplifies: 4773.06, 4773.061, 4773.08
Prior Effective Dates: 10/16/1995, 10/13/1997, 09/04/2000, 05/17/2004, 08/01/2007, 04/01/2012, 06/01/2018