Or. Admin. Code § 257-030-0140 - Determining Agreement of Breath Samples Within a Testing Sequence
Agreement between two valid breath samples within a testing sequence is established when the subject sample measurements agree within plus or minus ten percent of their mean, inclusive of the upper and lower bounds of the range. If the instrument establishes agreement, the lower breath sample measurement shall be truncated to two decimal places and reported as the chemical test result. If the subject sample measurements do not agree, the instrument will abort the testing sequence and display "Sample Correlation Failure".
Notes
Stat. Auth.: ORS 183.335 & 813.160
Stats. Implemented: ORS 813.160
State regulations are updated quarterly; we currently have two versions available. Below is a comparison between our most recent version and the prior quarterly release. More comparison features will be added as we have more versions to compare.