Or. Admin. Code § 333-122-0030 - General Requirements for All RGD: Warning Devices

(1) Open-beam configurations shall be provided with a readily discernible indication of:
(a) X-ray tube "on-off" status located near the radiation source housing, if the primary beam is controlled in this manner; and
(b) Shutter "open-closed" status located near each port on the radiation source housing, if the primary beam is controlled in this manner.
(2) Warning devices shall be labeled so that their purpose is easily identified. On equipment installed after January 1, 1978, warning devices shall have fail-safe characteristics.

Notes

Or. Admin. Code § 333-122-0030
PH 16-2021, adopt filed 04/21/2021, effective 4/21/2021

Statutory/Other Authority: ORS 453.605 - 453.807

Statutes/Other Implemented: ORS 453.605 - 453.807

State regulations are updated quarterly; we currently have two versions available. Below is a comparison between our most recent version and the prior quarterly release. More comparison features will be added as we have more versions to compare.


No prior version found.