Or. Admin. Code § 333-123-0070 - Control Panel Functions
A control panel must be designed to provide:
(1) An indication of whether electrical power is available at the control panel and if activation of the electronic brachytherapy source is possible;
(2) An indication of whether X-rays are being produced;
(3) A means for displaying electronic brachytherapy source accelerating voltage(kV), beam current(µA), pre-set value of radiation dose(cGy) and real time display of dose delivered;
(4) The means for terminating an exposure at any time; and
(5) An access locking control device that can prevent unauthorized use of the electronic brachytherapy device.
Notes
Stat. Auth.: ORS 453.605 - 453.807
Stats. Implemented: ORS 453.605 - 453.807
State regulations are updated quarterly; we currently have two versions available. Below is a comparison between our most recent version and the prior quarterly release. More comparison features will be added as we have more versions to compare.
No prior version found.