Or. Admin. Code § 410-122-0730 - Continuous Glucose Monitoring Equipment and Supplies
(1) Indications and Limitations of Coverage
and Medical Appropriateness:
(a) The Division
shall cover personal, real-time continuous glucose monitoring (CGM) equipment
and supplies for clients with a diagnosis of type 1 diabetes mellitus, type 2
diabetes mellitus, or gestational diabetes when criteria in the Prioritized
List Guideline note 108 are met;
(b) Continuous glucose monitoring equipment
and supplies (E2102, E2103, A4238, A4239, A9276, A9277, A9278) require prior
authorization with clinical documentation to support that all criteria in OAR
Ch 410 Div. 122 DMEPOS rules, and Table 122-XXXX are met;
(c) Initial authorization shall be limited to
6 months from the date of Division's authorization. Continued use beyond the
initial 6 months authorization requires the provider send to the Division
current medical documentation of a client follow up visit with the treating
practitioner which supports that there continues to be a medical need, that the
requirements of this rule are met, and the client is regularly utilizing the
CGM and the data from the CGM to adjust insulin; regularly utilizing means that
the client used the device for at least 50% of the time since the date of the
initial authorization;
(d) CGM
equipment must be FDA approved for the age of the client;
(e) Insulin pumps with integrated CGM
function require prior authorization and must meet criteria in this rule and
OAR 410-122-0525;
(f) The Division considers all other
applications of real-time CGM to be investigational as the effectiveness of its
use has not been established in peer-reviewed professional
literature;
(2)
Therapeutic/Non-adjunctive CGM devices (E2103 and A4239) are used to make
treatment decisions and replaces standard blood glucose monitoring and related
supplies. Standard blood glucose monitoring and supplies (E0607, E2100. E2101,
A4233, A4234, A4236, A4244. A4245, A4246, A4247, A4250, A4253. A4255. A4257,
A4258, A4259) cannot be billed separately to the Division on any day that falls
within the CGM authorization time period when the client is using
non-adjunctive CGM. This is a duplication of services;
(3) Non-therapeutic/Adjunctive CGM devices
(E2102 and A4238) do not replace standard blood glucose monitoring. These
supplies may be billed separately;
(4) Prior authorization requests for CGM
systems that are FDA approved but are not considered DME (i.e., CGM systems
that use smart phone exclusively for monitor/receiver) will be reviewed on a
case-by-case basis. Clinical documentation from the treating practitioner,
specific to the client, must demonstrate that all other covered CGM
alternatives have been considered, tried and failed, or reasonably expected to
fail;
(5) Providers may not use
A9276, A9277, or A9278 to bill supplies for any CGM device coded as an
adjunctive CGM (E2102) or non-adjunctive CGM (E2103) device;
(6) See Table 122-0730 for billing
codes.
Notes
To view attachments referenced in rule text, click here to view rule.
Statutory/Other Authority: ORS 413.042 & ORS 414.065
Statutes/Other Implemented: ORS 414.065
State regulations are updated quarterly; we currently have two versions available. Below is a comparison between our most recent version and the prior quarterly release. More comparison features will be added as we have more versions to compare.
No prior version found.