S.C. Code Regs. § 61-63.II.RHA 2.29 - Registration of Sealed Sources and Devices Containing Sealed Sources
2.29.1 Any manufacturer or initial
distributor of a sealed source or device containing a sealed source may submit
a request to the Department for evaluation of radiation safety information
about its product and for its registration.
2.29.2 The request for review must be sent to
the Department. The request for a review of a sealed source or a device must
include sufficient information about the design, manufacture, prototype
testing, quality control program, labeling, proposed uses and leak testing and
for a device, the request must also include sufficient information about
installation, service and maintenance, operating and safety instructions, and
its potential hazards, to provide reasonable assurance that the radiation
safety properties of the source or device are adequate to protect health and
minimize danger to life and property.
2.29.3 The Department normally evaluates a
sealed source or a device using radiation safety criteria in accepted industry
standards. If these standards and criteria do not readily apply to a particular
case, the Department formulates reasonable standards and criteria with the help
of the manufacturer or distributor. The Department shall use criteria and
standards sufficient to ensure that the radiation safety properties of the
device or sealed source are adequate to protect health and minimize danger to
life and property.
RHA
2.20 of this part includes specific criteria that
apply to certain exempt products and
RHA 2.4 includes specific
criteria applicable to certain generally licensed devices.
RHA 2.7 includes specific
provisions that apply to certain specifically licensed items.
2.29.4 After completion of the evaluation,
the Department issues a certificate of registration to the person making the
request. The certificate of registration acknowledges the availability of the
submitted information for inclusion in an application for a specific license
proposing use of the product, or concerning use under an exemption from
licensing or general license as applicable for the category of
certificate.
2.29.5 The person
submitting the request for evaluation and registration of safety information
about the product shall manufacture and distribute the product in accordance
with:
2.29.5.1 The statements and
representations, including quality control program, contained in the request;
and
2.29.5.2 The provisions of the
registration certificate.
2.29.6 Authority to manufacture or initially
distribute a sealed source or device to specific licensees may be provided in
the license without the issuance of a certificate of registration in the
following cases:
2.29.6.1 Calibration and
reference sources containing no more than:
2.29.6.1.1 37 MBq (1 mCi), for beta and/or
gamma emitting radionuclides; or
2.29.6.1.2 0.37 MBq (10 µCi), for alpha
emitting radionuclides; or
2.29.6.2 The intended recipients are
qualified by training and experience and have sufficient facilities and
equipment to safely use and handle the requested quantity of radioactive
material in any form in the case of unregistered sources or, for registered
sealed sources contained in unregistered devices, are qualified by training and
experience and have sufficient facilities and equipment to safely use and
handle the requested quantity of radioactive material in unshielded form, as
specified in their licenses; and
2.29.6.2.1
The intended recipients are licensed under
RHA 2.8, or comparable
regulation; or
2.29.6.2.2 The
recipients are authorized for research and development; or
2.29.6.2.3 The sources and devices are to be
built to the unique specifications of the particular recipient and contain no
more than 740 GBq (20 Ci) of tritium or 7.4 GBq (200 mCi) of any other
radionuclide.
2.29.7 After the certificate is issued, the
Department may conduct an additional review as it determines is necessary to
ensure compliance with current regulatory standards. In conducting its review,
the Department will complete its evaluation in accordance with criteria
specified in this section. The Department may request such additional
information as it considers necessary to conduct its review and the certificate
holder shall provide the information as requested.
2.29.8 Inactivation of certificates of
registration of sealed sources and devices.
2.29.8.1 A certificate holder who no longer
manufactures or initially transfers any of the sealed source(s) or device(s)
covered by a particular certificate issued by the Department shall request
inactivation of the registration certificate. Such a request must be made to
the Department and must normally be made no later than two years after initial
distribution of all of the source(s) or device(s) covered by the certificate
has ceased. However, if the certificate holder determines that an initial
transfer was in fact the last initial transfer more than two years after that
transfer, the certificate holder shall request inactivation of the certificate
within 90 days of this determination and briefly describe the circumstances of
the delay.
2.29.8.2 If a
distribution license is to be terminated in accordance with
RHA 2.11, the licensee shall
request inactivation of its registration certificates associated with that
distribution license before the Department will terminate the license. Such a
request for inactivation of certificate(s) must indicate that the license is
being terminated and include the associated specific license number.
2.29.8.3 A specific license to manufacture or
initially transfer a source or device covered only by an inactivated
certificate no longer authorizes the licensee to initially transfer such
sources or devices for use. Servicing of devices must be in accordance with any
conditions in the certificate, including in the case of an inactive
certificate.
Notes
State regulations are updated quarterly; we currently have two versions available. Below is a comparison between our most recent version and the prior quarterly release. More comparison features will be added as we have more versions to compare.
No prior version found.