S.C. Code Regs. § 61-65.II.RHC 2.5 - Vendor Registration and Obligation
2.5.1 Each person
who is a) engaged in the business of selling, leasing, or installing particle
accelerators or machine components; or b) offering to sell, lease, or install
particle accelerators or machine components; or c) engaged in the business of
furnishing or offering to furnish any equipment services in South Carolina shall
apply for registration as a vendor with the Department within 30 days following the
effective dates of these regulations or thereafter prior to furnishing or offering
to furnish any such services.
2.5.1.1 In-house
personnel employed by a registered facility or corporation shall be exempt from the
registration requirement, provided such personnel:
2.5.1.1.1 Shall meet the education, training, and
experience requirements for the appropriate vendor Class and;
2.5.1.1.2 Shall exclusively service one facility
or corporation.
2.5.1.2
Documentation of education, training, and experience for in-house service personnel
shall be maintained by the facility or corporation and available for Department
review.
2.5.2 Any person who
sells, leases, transfers, lends, moves, assembles or installs particle accelerators
in South Carolina shall notify the Department of the following within 30 days of the
transaction:
2.5.2.1 The name and address of
persons who have received the machine;
2.5.2.2 The manufacturer, control model and serial
number, and tube(s) model and serial number of each particle accelerator
transferred; and
2.5.2.3 The date of
transfer of each particle accelerator .
2.5.2.4 Notification to the Department shall be
made on DHEC Form 823. A DHEC 823 form shall be submitted to the Department each
month by Class I and Class II vendors, as outlined in Part II "Registration of X-Ray
Machines and Services" of R.61-64, X-Rays (Title B), regardless of whether a
particle accelerator was sold that month.
2.5.3 No person shall make, sell, lease, transfer,
lend, maintain, repair, assemble, reassemble, reinstall or install particle
accelerators or the supplies used in connection with such machines unless such
supplies and equipment, when properly placed in operation and used, meet the
requirements of these regulations. Each vendor shall ensure that the facility it is
providing with services or supplies is registered with the Department prior to
providing services or supplies.
2.5.4
Each vendor shall maintain records for review by the Department . These records shall
include, at a minimum:
2.5.4.1 All information
required by RHC 2.5;
2.5.4.2 Tests
performed at the time of installation to ensure that the equipment complies with
these regulations. A copy of these results shall be provided to the registrant at
the time of installation;
2.5.4.3
Records of any routine maintenance, repair, alterations, or reassembly of particle
accelerators. Records of maintenance, repair, alterations, or reassemblies shall
include the date that the service was performed and the legible signature of the
person performing the service. A copy of these records shall be provided to the
registrant at the time the service is provided;
2.5.4.4 Names of all employees and their dates of
employment with the vendor. Records shall also be maintained of training provided to
the employees during their term of employment.
2.5.5 All records required by this Part shall be
maintained by the vendor for review by the Department . Training records shall be
retained for personnel currently acting in any role as described in this Part. All
other records shall be retained for five years. All records shall be accurate and
factual.
2.5.6 Each vendor shall
maintain sufficient calibrated and operable instruments to perform the testing
appropriate to the class in which the vendor is registered. Instruments must be
calibrated with sources consistent with the conditions under which they are used.
Records shall be maintained of the calibrations performed on instrumentation used
for testing.
2.5.6.1 Survey meters used for
radiation area surveys shall be calibrated at intervals not to exceed 12 months and
after each instrument servicing.