Tenn. Comp. R. & Regs. 0400-20-07-.77 - FULL-INSPECTION SERVICING FOR TELETHERAPY AND GAMMA STEREOTACTIC RADIOSURGERY UNITS
(1) A licensee
shall have each teletherapy unit and gamma stereotactic radiosurgery unit fully
inspected and serviced during source replacement to assure proper functioning
of the source exposure mechanism and other safety components. The interval
between each full-inspection servicing shall not exceed five years for each
teletherapy unit and shall not exceed seven years for each gamma stereotactic
radiosurgery unit.
(2) This
inspection and servicing may only be performed by persons specifically licensed
to do so by the Division, the U.S. Nuclear Regulatory Commission, or an
Agreement State.
(3) A licensee
shall keep a record of the inspection and servicing in accordance with Rule
0400-20-07-.110.