Utah Admin. Code R313-28-55 - Radiographic Systems - Standby Radiation From Capacitor Discharge Equipment
Radiation emitted from the x-ray tube when the system is fully charged and the exposure switch or timer is not activated shall not exceed a rate of 0.516 uC/kg (two milliroentgens) per hour at five centimeters from accessible surfaces of the diagnostic source assembly, with the beam-limiting device fully open.