Utah Admin. Code R313-28-57 - Radiographic Systems Other Than Fluoroscopic, or Dental Intraoral Systems - mA/mAs Linearity
The following requirements apply when the equipment is operated on a power supply as specified by the manufacturer for fixed x-ray tube potentials within the range of 40 percent to 100 percent of the maximum rated potentials.
(1) Equipment having
independent selection of x-ray tube current, mA. Where the tube current is
continuous, the average ratios of exposure to the indicated milliampere-seconds
product, C/kg/mAs or mR/mAs, obtained at two consecutive tube current settings
or at two settings differing by no more than a factor of two shall not differ
by more than 0.10 times their sum.
(2) Equipment having a combined x-ray tube
current-exposure time product, mAs, selector, but not a separate tube current,
mA, selector. Where the tube current is continuous, the average ratios of
exposure to the indicated milliampere-seconds product, C/kg/mAs or mR/mAs,
obtained at two consecutive milliampere-seconds settings or at two settings
differing by no more than a factor of two shall not differ by more than 0.10
times their sum.