(a) Each owner or
operator subject to the provisions of Sections R315-265-1030 through
R315-265-1035 shall comply with the test methods and procedures requirements
provided in Section R315-265-1034.
(b) If a closed-vent system is tested for
compliance with no detectable emissions, as required in Subsection
R315-265-1033(k), the
test shall comply with the following requirements:
(1) Monitoring shall comply with Reference
Method 21 in 40 CFR part
60.
(2)
The detection instrument shall meet the performance criteria of Reference
Method 21.
(3) The instrument shall
be calibrated before use on each day of its use by the procedures specified in
Reference Method 21.
(4)
Calibration gases shall be:
(i) Zero air,
less than 10 ppm of hydrocarbon in air.
(ii) A mixture of methane or n-hexane and air
at a concentration of approximately, but less than, 10,000 ppm methane or
n-hexane.
(5) The
background level shall be determined as set forth in Reference Method
21.
(6) The instrument probe shall
be traversed around all potential leak interfaces as close to the interface as
possible as described in Reference Method 21.
(7) The arithmetic difference between the
maximum concentration indicated by the instrument and the background level is
compared with 500 ppm for determining compliance.
(c) Performance tests to determine compliance
with Subsection R315-265-1032(a) and with the total organic compound
concentration limit of Subsection R315-265.1033(c) shall comply with the
following:
(1) Performance tests to determine
total organic compound concentrations and mass flow rates entering and exiting
control devices shall be conducted and data reduced in accordance with the
following reference methods and calculation procedures:
(i) Method 2 in 40 CFR part
60 for velocity
and volumetric flow rate.
(ii)
Method 18 or Method 25A in 40 CFR part
60, appendix A, for organic content. If
Method 25A is used, the organic HAP used as the calibration gas shall be the
single organic HAP representing the largest percent by volume of the emissions.
The use of Method 25A is acceptable if the response from the high-level
calibration gas is at least 20 times the standard deviation of the response
from the zero calibration gas if the instrument is zeroed on the most sensitive
scale.
(iii) Each performance test
shall consist of three separate runs; each run conducted for at least one hour
under the conditions that exist if the hazardous waste management unit is
operating at the highest load or capacity level reasonably expected to occur.
For the purpose of determining total organic compound concentrations and mass
flow rates, the average of results of all runs shall apply. The average shall
be computed on a time-weighted basis.
(iv) Total organic mass flow rates shall be
determined by the following equation:
(A) For
sources utilizing Method 18 he equation found in
40 CFR
264.1034(c)(1)(iv)(A), is
adopted and incorporated by reference.
Where:
Eh = Total organic mass flow rate, kg/h.
Q2sd = Volumetric flow rate of gases entering or exiting
control device, as determined by Method 2, dscm/h.
n = Number of organic compounds in the vent gas.
Ci = Organic concentration in ppm, dry basis, of compound i
in the vent gas, as determined by Method 18.
MWi = Molecular weight of organic compound i in the vent gas,
kg/kg-mol.
0.0416 = Conversion factor for molar volume, kg-mol/m3 (@ 293
K and 760 mm Hg).
10-6 = Conversion from ppm.
(B) For sources utilizing Method 25A.
Eh = (Q)(C)(MW)(0.0416)(10-6).
Where:
Eh = Total organic mass flow rate, kg/h.
Q = Volumetric flow rate of gases entering or exiting control
device, as determined by Method 2, dscm/h.
C = Organic concentration in ppm, dry basis, as determined by
Method 25A.
MW = Molecular weight of propane, 44.
0.0416 = Conversion factor for molar volume, kg-mol/m3 (@ 293
K and 760 mm Hg).
10-6 = Conversion from ppm.
(v) The annual total organic emission rate
shall be determined by the following equation:
EA = (Eh) (H).
where:
EA = Total organic mass emission rate, kg/y.
Eh = Total organic mass flow rate for the process vent,
kg/h.
H = Total annual hours of operations for the affected unit,
h.
(vi) Total organic
emissions from all affected process vents at the facility shall be determined
by summing the hourly total organic mass emission rates, Eh, as determined in
Subsection R315-265-1034(c)(1)(iv), and by summing the annual total organic
mass emission rates, EA, as determined in Subsection R315-265-1034(c)(1)(v),
for all affected process vents at the facility.
(2) The owner or operator shall record such
process information as may be necessary to determine the conditions of the
performance tests. Operations during periods of startup, shutdown, and
malfunction shall not constitute representative conditions for the purpose of a
performance test.
(3) The owner or
operator of an affected
facility shall provide, or cause to be provided,
performance testing facilities as follows:
(i) Sampling ports adequate for the test
methods specified in Subsection R315-265-1034(c)(1).
(ii) Safe sampling platforms.
(iii) Safe access to sampling
platforms.
(iv) Utilities for
sampling and testing equipment.
(4) For the purpose of making compliance
determinations, the time-weighted average of the results of the three runs
shall apply. In the event that a sample is accidentally lost or conditions
occur in which one of the three runs must be discontinued because of forced
shutdown, failure of an irreplaceable portion of the sample train, extreme
meteorological conditions, or other circumstances beyond the owner or
operator's control, compliance may, upon the Director's approval, be determined
using the average of the results of the two other runs.
(d) To show that a process vent associated
with a
hazardous waste distillation, fractionation, thin-film evaporation,
solvent extraction, or air or steam stripping operation is not subject to the
requirements of Sections R315-265-1030 through R315-265-1035, the owner or
operator shall make an initial determination that the time-weighted, annual
average total organic concentration of the
waste managed by the
waste
management unit is less than 10 ppmw using one of the following two methods:
(1) Direct measurement of the organic
concentration of the
waste using the following procedures:
(i) The owner or operator shall take a
minimum of four grab samples of waste for each waste stream managed in the
affected unit under process conditions expected to cause the maximum waste
organic concentration.
(ii) For
waste generated onsite, the grab samples shall be collected at a point before
the waste is exposed to the atmosphere such as in an enclosed pipe or other
closed system that is used to transfer the waste after generation to the first
affected distillation fractionation, thin-film evaporation, solvent extraction,
or air or steam stripping operation. For waste generated offsite, the grab
samples shall be collected at the inlet to the first waste management unit that
receives the waste provided the waste has been transferred to the facility in a
closed system such as a tank truck and the waste is not diluted or mixed with
other waste.
(iii) Each sample
shall be analyzed and the total organic concentration of the sample shall be
computed using Method 9060A, incorporated by reference under Section
R315-260-11, of "
Test
Methods for Evaluating Solid
Waste, Physical/Chemical Methods," EPA Publication
SW-846; or analyzed for its individual organic constituents.
(iv) The arithmetic mean of the results of
the analyses of the four samples shall apply for each waste stream managed in
the unit in determining the time-weighted, annual average total organic
concentration of the waste. The time-weighted average is to be calculated using
the annual quantity of each waste stream processed and the mean organic
concentration of each waste stream managed in the unit.
(2) Using knowledge of the waste to determine
that its total organic concentration is less than 10 ppmw. Documentation of the
waste determination is required. Examples of documentation that shall be used
to support a determination under this provision include production process
information documenting that no organic compounds are used, information that
the waste is generated by a process that is identical to a process at the same
or another facility that has previously been demonstrated by direct measurement
to generate a waste stream having a total organic content less than 10 ppmw, or
prior speciation analysis results on the same waste stream where it can also be
documented that no process changes have occurred since that analysis that could
affect the waste total organic concentration.
(e) The determination that distillation
fractionation, thin-film evaporation, solvent extraction, or air or steam
stripping operations manage hazardous wastes with time-weighted annual average
total organic concentrations less than 10 ppmw shall be made as follows:
(1) By the effective date that the facility
becomes subject to the provisions of Sections R315-265-1030 through
R315-265-1035 or by the date when the waste is first managed in a waste
management unit, whichever is later; and
(2) For continuously generated waste,
annually; or
(3) If there is a
change in the waste being managed or a change in the process that generates or
treats the waste.
(f) If
an owner or operator and the Director do not agree on whether a distillation,
fractionation, thin-film evaporation, solvent extraction, or air or steam
stripping operation manages a hazardous waste with organic concentrations of at
least 10 ppmw based on knowledge of the waste, the dispute may be resolved
using direct measurement as specified at Subsection
R315-265-1034(d)(1).