In addition to other requirements of Rule R313-28, each
diagnostic x-ray system shall meet the requirements in Section
R313-28-35.
(1) Warning label. The control panel
containing the main power switch shall bear the warning statement, legible and
accessible to view: "WARNING: This x-ray unit may be dangerous to patient and
operator unless safe exposure factors and operating instructions are
observed."
(2) Battery charge
indicator. On battery powered generators visual means shall be provided on the
control panel to show whether the battery is in a state of charge adequate for
proper operation.
(3) Leakage
radiation from the diagnostic source assembly. The leakage radiation from the
diagnostic source assembly measured at a distance of one meter in any direction
from the source may not exceed 25.8 uC/kg, 100 milliroentgens, in one hour if
the x-ray tube is operated at its leakage technique factors.
(4) Radiation from components other than the
diagnostic source assembly. The radiation emitted by a component other than the
diagnostic source assembly may not exceed 0.516 uC/kg, two milliroentgens, in
one hour at five centimeters from accessible surfaces of the component if it is
operated in an assembled x-ray system under the conditions that it was designed
to operate under. Compliance shall be determined by measurements averaged over
an area of 100 square centimeters with no linear dimension greater than 20
centimeters.
(5) Beam quality.
(a) The half-value layer of the useful beam
for a given x-ray tube potential may not be less than the values shown in
Section
R313-28-35, Table 1. If it is
necessary to determine the half-value layer at an x-ray tube potential that is
not listed in Table 1, linear interpolation or extrapolation may be made.
|
Table 1
|
|
Design Operating Range (Kilovolts Peak)
|
Measured Potential (Kilovolts Peak)
|
Dental Intra-Oral Manufactured Before August 1,
1974 and On or Before December 1, 1980
|
All Other Diagnostic X-Ray Systems
|
|
Below 51
|
30
|
(use prohibited)
|
0.3
|
|
|
40
|
(use prohibited)
|
0.4
|
|
|
50
|
1.5
|
0.5
|
|
|
51
|
1.5
|
1.2
|
|
|
60
|
1.5
|
1.3
|
|
|
70
|
1.5
|
1.5
|
|
Above 70
|
71
|
2.1
|
2.1
|
|
|
80
|
2.3
|
2.3
|
|
|
90
|
2.5
|
2.5
|
|
|
100
|
2.7
|
2.7
|
|
|
110
|
3.0
|
3.0
|
|
|
120
|
3.2
|
3.2
|
|
|
130
|
3.5
|
3.5
|
|
|
140
|
3.8
|
3.8
|
|
|
150
|
4.1
|
4.1
|
(b)
For capacitor discharge equipment compliance with the requirements of
Subsection
R313-28-35(5)(a)
shall be determined with the system fully charged and a setting of 10 mAs for
exposures.
(c) The required minimal
half-value layer of the useful beam shall include the filtration contributed by
materials that are permanently present between the focal spot of the tube and
the patient.
(d) Filtration
control. For x-ray systems that have variable kVp and variable filtration for
the useful beam a device shall link the kVp selector with the filters and shall
prevent an exposure unless the minimum amount of filtration necessary to
produce the HVL required by Subsection
R313-28-35(5)(a)
is in the useful beam for the given kVp that has been selected.
(6) Multiple tubes. If two or more
radiographic tubes are controlled by one exposure switch the tube or tubes that
have been selected shall be clearly shown before initiation of the exposure.
For equipment manufactured after August 1, 1974, indications shall be both on
the x-ray control panel and at or near the tube housing assembly that has been
selected.
(7) Mechanical support of
tube head. The tube housing assembly supports shall be adjusted so that the
tube housing assembly will remain stable during an exposure unless the tube
housing movement during exposure is a designed function of the x-ray
system.
(8) Technique indicators.
(a) The technique factors to be used during
an exposure shall be shown before the exposure begins, except when automatic
EXPOSURE controls are used, in that case the technique factors that are set
before the exposure shall be indicated.
(b) On equipment having fixed technique
factors the requirements in Subsection
R313-28-35(8)(a)
may be met by permanent markings. Indication of technique factors shall be
visible from the operator's position except in the case of spot films made by
the fluoroscopist.
(9)
Maintaining compliance. Diagnostic x-ray systems and their associated
components certified pursuant to the provisions of 21 CFR Part
1020 (2006)
shall be maintained in compliance with applicable requirements of that
standard.
(10) Locks. Each position
locking, holding, and centering device on x-ray system components and systems
shall function as intended.
(11)
X-ray systems that have been granted a variance by the Director, Center for
Devices and Radiological Health, Food and Drug Administration (Director), from
the performance standards for ionizing radiation emitting products in
accordance with 21 CFR
1010.4 (2006) shall be accepted as satisfying
the requirements in Rule R313-28 that correspond to the variance granted by the
Director. The registrant shall insure that labeling pursuant to
21 CFR
1010.5(f) (2006) remains
legible and visible on the x-ray system.