Utah Code R313-35-111 - General Requirements for Veterinary X-ray Systems
(1) In addition to the requirements of
Section R313-35-110 veterinary x-ray
systems shall meet the requirements of Subsections R313-35-111(2) and
R313-35-111(3).
(2) Warning label.
The control panel containing the main power switch shall bear the warning
statement, legible and accessible to view: "WARNING: This x-ray unit may be
dangerous to patient and operator unless safe exposure factors and operating
instructions are observed."
(3)
Leakage radiation from the diagnostic source assembly. The leakage radiation
from the diagnostic source assembly measured at a distance of one meter in any
direction from the source may not exceed 25.8 uC/kg, 100 milliroentgens, in one
hour if the x-ray tube is operated at its leakage technique factors.