Utah Code R313-35-111 - General Requirements for Veterinary X-ray Systems

(1) In addition to the requirements of Section R313-35-110 veterinary x-ray systems shall meet the requirements of Subsections R313-35-111(2) and R313-35-111(3).
(2) Warning label. The control panel containing the main power switch shall bear the warning statement, legible and accessible to view: "WARNING: This x-ray unit may be dangerous to patient and operator unless safe exposure factors and operating instructions are observed."
(3) Leakage radiation from the diagnostic source assembly. The leakage radiation from the diagnostic source assembly measured at a distance of one meter in any direction from the source may not exceed 25.8 uC/kg, 100 milliroentgens, in one hour if the x-ray tube is operated at its leakage technique factors.