Wis. Admin. Code Department of Health Services DHS 157.39 - Leak testing and replacement of sealed sources
(1) AUTHORIZATION.
(a) The replacement of any sealed source
fastened to or contained in a radiographic exposure device and leak testing of
any sealed source shall be performed by persons authorized to do so by the
department, the nuclear regulatory commission or another agreement
state.
(b) The opening, repair or
modification of any sealed source shall be performed by persons specifically
authorized to do so by the department, the nuclear regulatory commission or
another agreement state.
(2) LEAK TESTING AND RECORD KEEPING
REQUIREMENTS.
(a) A licensee who uses a sealed
source shall have the source tested for leakage at intervals not to exceed 6
months. The leak testing of the source shall be performed using a method
approved by the department, the nuclear regulatory commission or by another
agreement state. The wipe sample should be taken from the nearest accessible
point to the sealed source where contamination might accumulate. The wipe
sample shall be analyzed for radioactive contamination. The analysis shall be
capable of detecting the presence of 185 becquerel (0.005 mCi) of radioactive
material on the test sample and shall be performed by a person specifically
authorized by the department, the US nuclear regulatory commission or another
agreement state to perform the analysis.
(b) A licensee shall maintain records of the
leak tests for 3 years.
(c) Unless
a sealed source is accompanied by a certificate from the transferor that shows
that it has been leak tested within 6 months before the transfer, it may not be
used by the licensee until tested for leakage. Sealed sources that are in
storage and not in use do not require leak testing, but shall be tested before
use or transfer to another person if the interval of storage exceeds 6 months.
(d) Any leak test that reveals the
presence of 185 becquerel (0.005 mCi) or more of removable radioactive material
shall be considered evidence that the sealed source is leaking. The licensee
shall immediately withdraw the equipment involved from use and shall have it
decontaminated and repaired, or disposed of. A report must be filed with the
department within 5 days of any test with results that exceed the threshold in
this paragraph, describing the equipment involved, the test results and the
corrective action taken.
(e)
1. Each exposure device using depleted
uranium shielding and an "S" tube configuration shall be tested for DU
contamination at intervals not to exceed 12 months. The analysis shall be
capable of detecting the presence of 185 becquerel (0.005 mCi) of radioactive
material on the test sample and shall be performed by a person specifically
authorized by the department, the nuclear regulatory commission or another
agreement state to perform the analysis. If the testing reveals the presence of
DU contamination, the exposure device shall be removed from use until an
evaluation of the wear of the S-tube has been made. If the evaluation reveals
that the S-tube is worn through, the device may not be used again.
2. DU shielded devices need not be tested for
DU contamination while not in use and in storage. If the DU shielded device has
not been used and in storage for more than 12 months, the device shall be
tested for DU contamination before using or transferring the device. A record
of the DU leak-test shall be retained for 3 years.