Wis. Admin. Code Department of Health Services DHS 157.80 - Computed tomography x-ray systems
(1) EQUIPMENT
REQUIREMENTS. A computed tomography (CT) x-ray system shall meet all of the
following requirements, as applicable:
(a)
Termination of exposure.
1.
Means shall be provided to terminate the x-ray exposure automatically by either
de-energizing the x-ray source or shuttering the x-ray beam in the event of
equipment failure affecting data collection. The termination shall occur within
an interval that limits the total scan time to no more than 110% of its preset
value through the use of either a back-up timer or devices which monitor
equipment function. A visible signal shall indicate when the x-ray exposure has
been terminated. The operator shall be able to terminate the x-ray exposure at
any time during a scan, or series of scans under CT x-ray system control, of
greater than one-half second duration.
2. For systems that allow high voltage to be
applied to the x-ray tube continuously and that control the emission of x-ray
with a shutter, the radiation emitted may not exceed 0.88 mGy (100mR) in 1 hour
at any point 5 centimeters outside the external surface of the housing of the
scanning mechanism when the shutter is closed.
3. Compliance shall be determined by
measurements averaged over an area of 100 square centimeters with no linear
dimension greater than 20 centimeters.
(b)
Tomographic plane indication and
alignment.
1. A single tomogram
system shall allow for visual determination of the tomographic plane or a
reference plane offset from the tomographic plane.
2. A multiple tomogram system shall allow for
visual determination of the location of a reference plane.
Note: The reference plane may be offset from the location of the tomographic planes.
3. If a device using a light source is used
to satisfy the requirements in subd. 2., the light source shall provide
illumination levels sufficient to permit visual determination of the location
of the tomographic plane or reference plane under ambient light conditions of
up to 500 lux.
(c)
Beam-on and shutter status indicators and control switches.
1. The CT x-ray control and gantry shall
provide visual indication whenever x-rays are produced and, if applicable,
whether the shutter is open or closed. Each emergency button or switch shall be
clearly labeled as to its function.
2. For systems that allow high voltage to be
applied to the x-ray tube continuously and that control the emission of x-ray
with a shutter, the radiation emitted may not exceed 0.88 mGy (100 mR) in one
hour at any point 5 cm outside the external surface of the housing of the
scanning mechanism when the shutter is closed.
3. Compliance shall be determined by
measurements averaged over an area of 100 square cm with no linear dimension
greater than 20 cm.
(d)
Indication of CT conditions of operation. A CT x-ray system
shall be designed such that the CT conditions of operation to be used during a
scan or a scan sequence shall be indicated prior to the initiation of the scan
or a scan sequence. On equipment having all or some of these conditions of
operation at fixed values, this requirement may be met by permanent markings.
Indication of CT conditions of operation shall be visible from any position
from which scan initiation is possible.
(e)
Maximum surface CTDI100
identification. The angular position where the maximum surface CTDI100
occurs shall be identified to allow for reproducible positioning of a CT
dosimetry phantom.
(f)
CT
x-ray systems containing a gantry manufactured after September 3,
1985. A computed tomography x-ray system containing a gantry that was
manufactured after September 3, 1985, shall meet all the following
requirements:
1. The total error in the
indicated location of the tomographic plane or reference plane may not exceed 5
millimeters.
2. If the x-ray
production period is less than one-half second, the indication of x-ray
production shall be actuated for at least one-half second. Indicators at or
near the gantry shall be discernible from any point external to the patient
opening where insertion of any part of the human body into the primary beam is
possible.
3. The deviation of
indicated scan increment versus actual increment may not exceed plus or minus
one millimeter with any mass from 0 to 100 kilograms resting on the support
device. The patient support device shall be incremented from a typical starting
position to the maximum incremented distance or 30 centimeters, whichever is
less, and then returned to the starting position. Measurement of actual versus
indicated scan increment may be taken anywhere along this patient support
device movement distance.
4.
Premature termination of the x-ray exposure by the operator shall necessitate
resetting of the CT conditions of operation prior to the initiation of another
scan.
5. Two-way aural
communication shall exist between the patient and the operator at the control
panel.
6. A viewing window, or
viewing system, such as closed circuit television or an equivalent, shall be
installed to permit continuous observation of the patient during irradiation.
The window or viewing system shall be installed such that the operator, located
at the control position, can continuously observe the patient during
irradiation. When the primary viewing system is electronic, an alternative
system shall be available for use in the event of failure of the primary
viewing system.
(2) OPERATING PROCEDURES.
(a)
1. A CT
x-ray system for human use may only be operated for diagnostic procedures by an
American registry of radiologic technologists certified person who is licensed
as a radiographer by the State of Wisconsin or has met the radiographer license
exemptions and has been specifically trained in its operation.
2. A CT x-ray system for veterinary use may
only be operated for diagnostic procedures by a person who is certified by the
American registry of radiological technologists or has completed training
equivalent to the requirements of ch. DHS 157 Appendix L and has been
specifically trained in its operation.
3. Combination systems which combine CT with
another imaging device that uses radioactive material shall be operated by a
person qualified by training in the safe use of radioactive materials and who
meets the training requirements of ch. DHS 157 Appendix L.
(b) Information shall be available at the
control panel regarding the operation and calibration of the system. The
information shall include all of the following components:
1. Dates of the latest calibration and spot
checks and the location within the facility where the results of those tests
may be obtained.
2. Instructions on
the use of the CT dosimetry phantom including a schedule of spot checks
appropriate for the system, allowable variations for the indicated parameters
and the results of at least the most recent spot checks conducted on the
system.
3. The distance in
millimeters between the tomographic plane and the reference plane if a
reference plane is utilized.
4. A
current technique chart or list of protocols available at the control panel,
which specifies for each routine examination the CT conditions of operation and
the number of scans per examination including body part size and correct kV/mA
for that body part. The technique chart or a list of protocols shall be used to
adjust techniques based on the body part being examined.
(c) Calibration and spot check measurements
shall be made at a frequency recommended by the manufacturer or established by
a medical physicist. If the calibration or spot check of the CT x-ray system
identifies that a system operating parameter has exceeded a tolerance
established by the medical physicist, use of the CT x-ray system on patients
shall be limited to those uses permitted by established written instructions of
the medical physicist.
(d) A
facility shall follow the manufacturer's daily start up routines and
preventative maintenance schedules for a specific computed tomography x-ray
system.