023-3 Wyo. Code R. §§ 3-18 - Meter Testing Program
(a) Each utility
shall develop a meter testing program for the calibration, recertification,
care and maintenance of meters, recording devices, field testing equipment and
meter calibration equipment (hereinafter referred to in this section as
"equipment") in order to keep the equipment in proper working condition. The
utility shall have access to a meter laboratory, standard meters, instruments,
meter calibration equipment and facilities necessary to carry out its meter
testing program. The facilities and equipment shall be available at reasonable
times for inspection by any authorized representative of the
Commission.
(b) Each utility shall
submit its meter testing program to the Commission for approval.
(c) At a minimum, the meter testing program
shall include:
(i) Identification of all
equipment requiring calibration or recertification;
(ii) Identification of the nationally
recognized standards for the proposed testing and calibration, testing
intervals and degree of accuracy for all equipment;
(iii) A plan for proper care and maintenance
of equipment;
(iv) A requirement
that equipment used as a reference or transfer standard shall have a higher
degree of accuracy than the item being calibrated and the calibration is
traceable to NIST;
(v) A
requirement that equipment used as a reference or transfer standard to certify
other equipment shall be kept in temperature-stable environments and calibrated
annually;
(vi) A requirement that
equipment used as a reference or transfer standard shall only be used for
calibration purposes and shall not be used for trouble shooting, corrective
maintenance or any other activity that might jeopardize the instrument's
calibration accuracy;
(vii) For
natural gas utilities, a requirement that bell provers, transfer provers or
sonic nozzles shall be recertified at not more than five-year intervals and
after being moved or dimensionally altered. Recertification shall be
accomplished by use of standard-cubic foot bottles, by strapping or by optical
measurement techniques, the latter being the preferable method;
(viii) For natural gas utilities, a
requirement that calibrating meters equate a cubic foot of gas with the amount
of gas that occupies one cubic foot, dry, at 60° F at 14.73 pounds per
square inch absolute; and
(ix) For
electric utilities, a requirement that instrument transformers shall have its
installation verified and be tested for ratio and burden performance at the
same intervals as the associated meter.
(d) All meter calibration equipment shall
bear the last calibration date.
(e)
Each electric utility furnishing metered electric service shall provide or have
available portable indicating electrical testing instruments of suitable range
and type for testing and confirming equipment installations are performing
properly.
(f) Each water utility
furnishing metered water service using portable test meters to determine the
accuracy of meters in service shall recalibrate the portable test meters at
sufficiently frequent intervals to ensure correct registration at the specified
rates of flow.
Notes
State regulations are updated quarterly; we currently have two versions available. Below is a comparison between our most recent version and the prior quarterly release. More comparison features will be added as we have more versions to compare.
No prior version found.