034-4 Wyo. Code R. §§ 4-5 - Radiograph Use

Inspections of x-ray machines are required as follows:

(a) Inspectors. Any individual who performs x-ray machine assembly, installation or service shall meet the following educational and experience requirements:
(i) Completion of a structured educational program that includes training in radiation machine safety, assembly, installation and service, including, but not limited to:
(A) A baccalaureate degree in electrical engineering with specialized training in radiation producing devices;
(B) A one-year associate degree in biomedical equipment repair;
(C) Equivalent manufacturer, military or other technical school training; and
(ii) At least six (6) months of supervised, documented training on inspection and calibration of the applicable x-ray machine.
(b) Inspections. Inspections of x-ray equipment shall be conducted in accordance with relevant national standards. All machines shall be inspected at the time of installation and thereafter every five (5) years for x-ray machines and every three (3) years for computed tomography machines.
(c) Waiver. Licensees may be granted a six (6) month waiver to the inspection requirement at the discretion of the Board.
(d) Compliance and Documentation. Failure to abide with the requirements of this section shall be considered "unprofessional conduct" within the meaning of Chapter 9 and shall constitute grounds for disciplinary action by the Board.
(i) Inspector should provide evidence for dentists to place on machines indicating the last inspection date and next inspection due date.
(ii) Evidence of compliance to the Board is not required unless specifically requested.

Notes

034-4 Wyo. Code R. §§ 4-5
Adopted, Eff. 1/14/2022. Amended, Eff. 5/29/2024.