A. If inventory contro l is used to meet the
requirements of
R18-12-241, it shall be used in
conjunction with tank tightness testing described in subsection (C). Inventory
control shall be conducted monthly in accordance with
R18-12-281(O) to
detect a release of at least 1.0% of flow-through plus 130 gallons on a monthly
basis in the following manner:
1. Inventory
volume measurements for regulated substance inputs, withdrawals, and the amount
still remaining in the tank are recorded each operating day;
2. The equipment used is capable of measuring
the level of the regulated substance over the full range of the tank's vertical
dimension to the nearest 1/8 of an inch;
3. The regulated substance inputs are
reconciled with delivery receipts by measurement of the tank inventory volume
before and after delivery;
4.
Deliveries of regulated substances are made through a drop tube that extends to
within one foot of the tank bottom;
5. Dispensing of regulated substances is
metered and recorded within the standards established by the entity with
jurisdiction. If no standards are established, dispensing which meets an
accuracy of six cubic inches for every five gallons of regulated substance
withdrawn shall be used;
6. The
measurement of any water level in the bottom of the tank is made to the nearest
1/8 of an inch at least once a month;
7. Inventory control shall not be utilized as
the sole method of release detection.
B. Manual tank gauging used to meet the
requirements of
R18-12-241 shall meet all of the
following requirements:
1. Tank liquid level
measurements are taken at the beginning and ending of a period equal to the
appropriate minimum duration of test in the table below during which no liquid
is added to or removed from the UST system;
2. Level measurements are based on an average
of two consecutive stick readings at both the beginning and ending of the
period;
3. The equipment used is
capable of measuring the level of regulated substance over the full range of
the tank's vertical dimension to the nearest 1/8 of an inch;
4. A release is suspected and subject to the
requirements of A.R.S. §
49-1004 and the rules promulgated
thereunder if the variation between beginning and ending measurements exceeds
the weekly or monthly standards in the following table:
|
Nominal Tank Capacity
|
Minimum duration of test
|
Weekly standard (1 test)
|
Monthly standard (average of 4 tests)
|
|
550 gallons or less
|
36 hours
|
10 gallons
|
5 gallons
|
|
551-1,000 gallons (when tank diameter is 64
inches)
|
44 hours
|
9 gallons
|
4 gallons
|
|
551-1,000 gallons (when tank diameter is 48
inches)
|
58 hours
|
12 gallons
|
6 gallons
|
|
551-1,000 gallons (also requires periodic tank
tightness testing)
|
36 hours
|
13 gallons
|
7 gallons
|
|
1,001-2,000 gallons (also requires periodic tank
tightness testing)
|
36 hours
|
26 gallons
|
13 gallons
|
5.
Manual tank gauging may be used as the sole method of release detection only
for tanks of 550 gallons or less capacity and tanks with a nominal capacity of
551 to 1,000 gallons that meet the tank diameter criteria in the table in
subsection (B)(4) of this Section. Manual tank gauging may be used in place of
inventory control in subsection (A) for all other tanks of 551 to 2,000
gallons. This method shall not be used to meet the requirements of
R18-12-241 for tanks of greater
than 2,000 gallons capacity.
C. Tank tightness testing shall be capable of
detecting a 0.1 gallon per hour leak rate from any portion of the tank that
routinely contains a regulated substance while accounting for the effects of
thermal expansion or contraction of the regulated substance, vapor pockets,
tank deformation, evaporation or condensation, and the location of the water
table.
D. Equipment for automatic
tank gauging that tests for the loss of regulated substance and conducts
inventory control used to meet the requirements of
R18-12-241 shall meet all of the
following requirements:
1. The automatic
regulated substance level monitor test shall be performed at least monthly and
be capable of detecting a 0.2 gallon per hour leak rate from any portion of the
tank that routinely contains regulated substance,
2. The automatic tank gauging equipment shall
meet the inventory control (or other test of equivalent performance)
requirements of subsection (A), and
3. The test shall be performed with the
system operating in one of the following modes:
a. In-tank static testing conducted at least
once monthly; or
b. Continuous
in-tank leak detection operating on an uninterrupted basis or operating within
a process that allows the system to gather incremental measurements to
determine the leak status of the tank at least once monthly.
E. Testing or monitoring
for vapors within the soil gas of the excavation zone used to meet the
requirements of
R18-12-241 shall be conducted at
least monthly and shall meet all of the following requirements:
1. In the UST excavation zone, the site is
assessed to ensure that the leak detection method will comply with the
requirements in subsections (E)(2) through (6);
2. The leak detection system is constructed
and designed so that the number and positioning of monitoring wells will detect
releases into the excavation zone from any portion of the system which
routinely contains a regulated substance within 30 days from the date of
commencement of a release;
3. The
stored regulated substance, or a tracer compound placed in the UST system, will
produce a vapor level that is detectable by the monitoring devices in the
monitoring wells within 30 days from the date of commencement of a release from
the UST system;
4. The materials
used as backfill will allow diffusion of vapors from releases into the
excavation area such that a release is detected within 30 days from the date of
commencement of a release from the UST system;
5. The groundwater, rainfall, soil moisture,
or other known interferences will not render the measurement of vapors by the
monitoring device inoperable so that a release could go undetected by the
monitoring devices in the monitoring wells for more than 30 days from the date
of commencement of the release from the UST system;
6. The level of background contamination at
the site will not interfere with the method used to detect releases from the
tank system;
7. The vapor monitors
are designed and operated to detect any significant increase in concentration
above a documented background level of the regulated substance stored in the
tank system, a component or components of that substance, or a volatile tracer
compound placed in the tank system;
8. Monitoring wells are clearly marked and
secured to avoid unauthorized access and tampering.
F. Testing or monitoring for liquids on the
groundwater used to meet the requirements of
R18-12-241 shall be conducted
monthly and meet the following requirements:
1. Within and immediately below the UST
excavation zone, the site is assessed to ensure that the leak detection method
will comply with the requirements in subsections (F)(2) through (7);
2. The leak detection system shall be
constructed and designed so that the number and positioning of monitoring wells
or devices will detect releases into the excavation zone from any portion of
the system which routinely contains a regulated substance;
3. The regulated substance stored is
immiscible in water and has a specific gravity of less than 1;
4. Groundwater is never more than 20 feet
from the ground surface and the hydraulic conductivity of the material between
the UST system and the monitoring wells or devices is not less than 0.01
centimeters per second (e.g., the soil should consist of gravels, coarse to
medium sands, coarse silts or other permeable materials);
5. Monitoring wells or devices intercept the
excavation zone or are as close to it as is technically feasible;
6. The slotted portion of the monitoring well
casing shall be designed to prevent migration of natural soils or filter pack
into the well and to allow entry of regulated substance on the water table into
the well under both high and low ground-water conditions;
7. Monitoring wells shall be sealed from the
ground surface to the top of the filter pack;
7
8. The continuous
monitoring devices or manual methods used can detect the presence of at least
1/8 of an inch of free product on top of the groundwater in the monitoring
wells;
8.
Monitoring wells shall be sealed from the ground
surface to the top of the filter pack;
9. Monitoring wells are clearly marked and
secured to avoid unauthorized access and tampering.
G. Interstitial monitoring between the UST
system and a secondary barrier immediately around or beneath it which is used
to meet the requirements of
R18-12-241 shall be conducted at
least monthly and shall be designed, constructed and installed to detect a leak
from any portion of the UST system that routinely contains a regulated
substance, and shall meet one of the following requirements:
1. For double-walled UST systems, the
sampling or testing method shall be able to detect a leak through the inner
wall in any portion of the UST system that routinely contains a regulated
substance.
2. For UST systems with
a secondary barrier within the excavation zone, characteristics of the site and
system components shall be designed and constructed to detect a leak between
the UST system and the secondary barrier and shall meet all of the following
requirements:
a. The secondary barrier around
or beneath the UST system shall be constructed of synthetic materials which are
sufficiently thick and impermeable to prevent structural weakening of the
secondary barrier as a result of contact with any released regulated substance.
The rate of permeability shall not exceed 10-6
centimeters per second for the regulated substance stored. In addition, the
secondary barrier shall be capable of directing any leak to the monitoring
point and permit its detection;
b.
The barrier is compatible with the regulated substance stored so that a leak
from the UST system will not cause a deterioration of the barrier allowing a
release to pass through undetected;
c. For cathodically protected UST systems,
the secondary barrier shall be installed so that it does not interfere with the
proper operation of the cathodic protection system;
d. The groundwater, soil moisture, or
rainfall will not render the testing or sampling method used inoperative so
that a release could go undetected for more than 30 days;
e. The characteristics of the UST site are
assessed to ensure that the secondary barrier is always above the groundwater
and not in a 25-year flood plain, unless the barrier and monitoring designs are
for use under such conditions;
f.
Monitoring wells are clearly marked and secured to avoid unauthorized access
and tampering.
3. For
tanks with an internally fitted liner, an automated device shall be able to
detect a leak between the inner wall of the tank and the liner, and the liner
shall be compatible with the substance stored.
H. Statistical inventory reconciliation.
Release detection methods based on the application of statistical principles to
inventory data similar to those described in
R18-12-243(A)
shall meet the following requirements:
1.
Report a quantitative result with a calculated leak rate;
2. Be capable of detecting a leak rate of 0.2
gallon per hour or a release of 150 gallons within 30 days; and
3. Use a threshold that does not exceed
one-half the minimum detectible leak rate.
H
I. Any other type of
release detection method, or combination of methods, may be used to meet the
requirements of
R18-12-241 if all of the following
requirements are met:
1. The monitoring is
conducted at least monthly;
2. The
Department determines that the method meets either of the following
requirements:
a. The method can detect a 0.2
gallon per hour leak rate or a release of 150 gallons within 30 days with
probability of detection and probability of false alarm in accordance with
R18-12-240(A)(4);
b. The owner and operator can demonstrate
that the method is able to detect a release as effectively as any of the
methods allowed in subsections (C) through (H). In comparing methods, the
Department shall consider the size of release that the method can detect and
the frequency and reliability with which it can be detected. If the method is
approved, the owner and operator shall comply with any conditions imposed by
the Department on its use to ensure the protection of human health and the
environment.