(a)
Definitions. For the purposes of this section, the definitions set forth in
section
93108(a) shall
apply unless otherwise specified below:
(1)
"Administrator" means the Administrator of the United States Environmental
Protection Agency (or the implementing agency in accordance with any delegation
of authority to approve alternatives from the U. S. Environmental Protection
Agency).
(2) "Back-draft
valve/chamber exhaust stream" is the air stream which results from collection
of ethylene oxide-contaminated air which may be removed from the sterilizer
through a back-draft valve or rear chamber exhaust system during unloading of
the sterilized materials.
(3)
"Baseline temperature" means the range of temperatures at the outlet point of a
catalytic oxidation control device or at the exhaust point from the combustion
chamber for a thermal oxidation control device established during the
performance test at which the unit achieves at least 99 percent control of
ethylene oxide emissions.
(4)
"Manifolding emissions" means combining ethylene oxide emissions from two or
more vent types for the purpose of controlling these emissions with a single
control device.
(5) "Maximum
ethylene glycol concentration" means the concentration of ethylene glycol in
the scrubber liquor of an acid-water scrubber control device established during
a performance test when the scrubber achieves at least 99 percent control of
ethylene oxide emissions.
(6)
"Maximum liquor tank level" means the level of scrubber liquor in the
acid-water scrubber liquor recirculation tank established during a performance
test when the scrubber achieves at least 99 percent control of ethylene oxide
emissions.
(7) "Modification" means
either
(A) any physical change in, method of
operation of, or addition to, an existing permit unit that requires an
application for a permit to construct and/or operate. Routine maintenance
and/or repair shall not be considered a physical change. A change in the method
of operation of equipment, unless previously limited by an enforceable permit
condition, shall not include:
1. an increase
in the production rate, unless such increases will cause the maximum design
capacity of the equipment to be exceeded; or
2. an increase in the hours of operation;
or,
3. a change in ownership of a
source; or,
(B) the
addition of any new permit unit at an existing source; or,
(C) the replacement of components if the
fixed capital cost of the components exceeds 50 percent of the fixed capital
cost that would be required to construct a comparable new
source.
(8) "Oxidation
temperature" means the temperature at the outlet point of a catalytic oxidation
device or at the exhaust point from the combustion chamber for a thermal
oxidation device.
(9) "Parametric
monitoring" means monitoring of a specific operating parameter or parameters of
a control device established to demonstrate that the control device is
operating under conditions that meet a performance
standard.
(b)
Applicability. Effective January 28, 1999, any person who owns or operates a
commercial sterilizer or an aerator using 2,000 pounds or more of ethylene
oxide in any 12 consecutive month period after December 6, 1996, must comply
with this section.
(c) Initial
Notification. Any person subject to this regulation must provide the following
information, in writing, within 30 days after the source becomes subject to the
regulation, to the district and the Administrator unless the Administrator has
waived this requirement:
(1) The name(s) and
address of the owner and operator of the facility;
(2) The location of the facility;
(3) The number of sterilizers and aerators at
the facility;
(4) An estimate of
the facility-wide pounds of ethylene oxide used per year;
(5) A brief description of the nature, size,
design, design operating capacity, expected control efficiency, and method of
operation of the source, and control equipment, including operating design
capacity, bypass valves, and an identification of each point of
emission;
(6) Facilities complying
with this regulation with a control technology other than acid-water scrubbers
or catalytic or thermal oxidizers must provide information describing the
design and operation of the air pollution control system including
recommendations for the operating parameters to be monitored that will indicate
proper operation and maintenance, where the site specific operating, reporting
and monitoring parameters will be determined during the performance
test;
(7) A statement of whether
the source is a major or area source to the Administrator (If the source is a
new major source or a major source undergoing modification, it must receive
written approval in advance from the Administrator. The source may use the
"Application for Construction or Modification" in Appendix 2 to satisfy the
initial notification requirements); and
(8) An identification of the relevant
standard, or other requirement, that is the basis of the notification and the
source's compliance date.
(d) Requirements. No person subjected to
these standards shall operate a sterilizer or aerator, unless all of the
following requirements are satisfied:
(1) all
ethylene oxide released from the sterilizer and aerator shall be controlled to
meet the requirements shown in Table I for the applicable control category;
Table I
Emissions Standards for Commercial
Facilities
Control
Category
|
Requirements for
Ethylene Oxide Sterilizer Facilities
|
(Facility-wide Pounds of Ethylene Oxide
used per 12 consecutive months) |
(a)
Emission Streams to be Controlled |
(b)
Emission Streams to be Tested |
(c) Control
Efficiency (%) or Outlet Concentration |
equal to or greater than 2,000 and less than
5,000 |
Sterilizer |
Sterilizer |
99.9 |
|
Aerator |
Aerator |
95.0 |
|
Back-draft Valve |
|
* |
|
|
|
|
|
Aeration Only |
|
95.0 |
|
|
|
|
equal to or greater than 5,000 and less than
20,000 |
Sterilizer |
Sterilizer |
99.9 |
|
Aerator |
Aerator |
99.0 |
|
Sterilizer Door Hood |
|
* |
|
& |
|
|
|
Back-draft Valve |
|
* |
|
Aeration Only |
|
95.0 |
|
|
|
|
Equal to or more than
20,000 |
Sterilizer |
Sterilizer |
99.9 |
|
Aerator |
Aerator |
99.0
or |
|
|
|
1 ppm max |
|
Sterilizer Door Hood |
|
* |
|
Back-draft Valve |
|
99.0* |
|
Aeration Only |
|
99.0 |
__________ |
|
|
|
* Sources may show compliance by manifolding emissions to
control device used to comply with sterilizer or aerator
requirement.
(2) the exhaust
systems and EtO supply including, but not limited to, any piping, ducting,
fittings, valves, or flanges, through which ethylene oxide is conveyed to and
from the sterilizer, aerator and the control device shall be leak-free;
and
(3) Facilities must obtain a
title V permit from the Administrator.
(e) Compliance Procedures.
(1) Compliance Testing Notification
The facility shall notify the Administrator 60 days before
the date and time of any performance tests and monitoring system evaluations.
In the event the source is unable to conduct the test on the date specified in
the notification, the source shall notify the Administrator within 5 days prior
to the scheduled performance test date.
(2) Compliance Testing
(A) Source testing conducted for the purpose
of demonstrating compliance must be according to ARB Test Method 431 (title 17,
CCR, section
94143) and the method evaluations
cited therein or an acceptable source test method approved by the district with
the concurrence of the Executive Officer of the Air Resources Board, and the
Administrator. Before conducting a required source test, the source shall
develop a site-specific test program summary, the test schedule, data quality
objectives, and both an internal and external quality assurance
program.
(B) The following
procedures shall be used to determine the monitored parameters for acid-water
scrubbers:
1. For determining the ethylene
glycol concentration, the facility owner or operator shall establish the
maximum ethylene glycol concentration as the ethylene glycol concentration
averaged over three test runs; the sampling and analysis procedures in ASTM D
3695-88, Standard Test Method for Volatile Alcohols in Water by Direct Aqueous-
Injection Gas Chromatography (1988).
2. For determining the scrubber liquor tank
level, the sterilization facility owner or operator shall establish the maximum
liquor tank level based on a single measurement of the liquor tank level during
one test run.
(C) The
following procedures shall be used to demonstrate the baseline temperature for
catalytic oxidation units or thermal oxidation units and to continuously
monitor the oxidation temperature as required by this measure.
1. The baseline temperature for the
sterilization chamber vent shall be the temperature for the catalytic oxidation
unit or oxidation temperature at the exhaust point from the thermal oxidation
unit averaged over three test runs using the procedures in Test Method 431, and
subsection (f)(2)(A).
2. The
baseline temperature for the aeration room vent shall be the temperature for
the catalytic oxidation unit or the oxidation temperature at the exhaust point
from the thermal oxidation unit averaged over three test runs using the
procedures in Test Method 431, and subsection (f)(2)(B).
3. The baseline temperature for the chamber
exhaust vent shall be the temperature for the catalytic oxidation unit or
oxidation temperature at the exhaust point from the thermal oxidation unit
averaged over three test runs using the procedures in Test Method 431, and
subsection (f)(2)(C).
(D)
A facility seeking to demonstrate compliance with the standards with a control
device other than an acid-water scrubber or catalytic or thermal oxidation unit
shall submit: a description of the device; tests results collected in
accordance with the test method cited within or an approved method verifying
the performance of the device for controlling ethylene oxide emissions to the
levels required by the applicable standards; the appropriate operating
parameters that will be monitored; and the frequency of measuring and recording
to establish continuous compliance with the standards. The monitoring plan is
subject to the Administrator's approval. The owner or operator of the
sterilization facility shall install, calibrate, operate, and maintain the
monitor(s) approved by the Administrator based on the information submitted by
the owner or operator. The owner or operator shall include in the information
submitted to the Administrator proposed performance specifications and quality
assurance procedures for their monitors.
(E) A facility seeking to demonstrate
compliance with the standards with a monitoring device or procedure other than
a gas chromatograph shall provide to the Administrator information describing
the operation of the monitoring device or procedure and the parameter(s) that
would indicate proper operation and maintenance of the device or
procedure.
(3) Compliance
Testing Report
(A) The facility shall send the
district and the Administrator an initial statement of compliance and test
results within 60 days following the performance test.
(B) The facility shall submit (before a title
V permit is issued) to the Administrator:
1.
The methods that were used to determine compliance;
2. The results of any performance tests,
continuous monitoring system (CMS) performance evaluations, and/or other
monitoring procedures or methods that were conducted;
3. The methods that will be used for
determining continuing compliance, including a description of monitoring and
reporting requirements and test methods; and
4. A statement by the owner or operator of
the affected existing, new, or modified source as to whether the source has
complied with the relevant standard or other
requirements.
(f) Monitoring Requirements. The owner or
operator of a sterilizer or aerator shall monitor the parameters of the control
system specified in this section to show compliance with the provisions of this
regulation. If continuous monitoring systems are required, Appendix 1 shall be
consulted for their application. All monitoring equipment shall be installed
such that representative measurements of emissions or process parameters which
affect emissions from the source are obtained. For monitoring equipment
purchased from a vendor, verification of the operational status of the
monitoring equipment shall include, at a minimum, completion of the
manufacturer's written specifications or recommendations for installation,
operation, maintenance, and calibration of the system.
(1) For sterilization facilities complying
with the emissions standard through the use of an acid-water scrubber, the
owner or operator shall either:
(A) Sample the
scrubber liquor and analyze and record once per week the ethylene glycol
concentration using the test procedures in subsection (e)(2)(B)1. Monitoring is
required only if the scrubber unit has been operated during that week;
or
(B) Measure and record once per
week the level of the scrubber liquor in the recirculation tank. The owner or
operator shall install, maintain, calibrate, and use a liquid level indicator
to measure the scrubber liquor tank level (i.e., a visible depth gauge, a
dipstick, a magnetic indicator, etc.).
(C) Operation of the facility with an
ethylene glycol concentration in the scrubber liquor in excess of the maximum
liquor tank level shall constitute a violation of the chamber exhaust vent
standard for sources using 20,000 pounds or more of ethylene oxide per 12
consecutive months.
(2)
For sterilization facilities complying with the emissions standards through the
use of catalytic oxidation or thermal oxidation, the owner or operator shall
continuously monitor and record the oxidation temperature at the outlet to the
catalyst bed or at the exhaust point from the thermal combustion chamber using
a temperature monitor. The temperature monitor shall be installed, calibrated,
operated, and maintained to an accuracy within
±5.6°C(±10°F). The owner or operator shall verify the
accuracy of the temperature monitor twice each calendar year with a reference
temperature monitor (traceable to National Institute of Standards and
Technology (NIST) standard, or with an independent temperature measurement
device dedicated for this purpose). During accuracy checking, the probe of the
reference device shall be at the same location as that of the temperature
monitor being tested.
For sources using 20,000 pounds or more of ethylene oxide
per 12 consecutive months, operation of the facility with the oxidation
temperature, averaged over the cycle, more than 5.6°C (10°F) below the
baseline temperature shall constitute a violation of the chamber exhaust vent
standard.
(A) For the sterilization
chamber vent, a data acquisition system for the temperature monitor shall
compute and record an average oxidation temperature over the length of the
cycle (based on the length of the cycle used during the performance test) and a
three-cycle block average every third cycle.
(B) For the aeration room vent, a data
acquisition system for the temperature monitor shall compute and record an
average oxidation temperature each hour and a 3-hour block average every third
hour.
(C) For the back draft valve
(chamber exhaust vent), a data acquisition system for the temperature monitor
shall compute and record an average oxidation temperature over the length of
the cycle (based on the length of the cycle used during the performance
test).
(3) For
sterilization facilities complying with the emission standards with the use of
a control device other than acid-water scrubbers or catalytic or thermal
oxidizers, the owner or operator shall monitor the parameters as approved by
the Administrator.
(4) For
facilities continuously measuring the ethylene oxide concentration from the
aeration room (after a control device) or in the sterilization chamber
immediately prior to the operation of the chamber exhaust, the owner or
operator shall follow either paragraph (A) or (B) of this subsection:
(A) Measure and record once per hour the
ethylene oxide concentration at the outlet to the atmosphere from the aeration
room vent after any control device. The owner or operator shall compute and
record a 3-hour average every third hour. The owner or operator will install,
calibrate, operate, and maintain a gas chromatograph to measure ethylene oxide.
The daily calibration requirements are required only on days when ethylene
oxide emissions are vented to the control device from the aeration room
vent.
(B) Measure and record the
ethylene oxide concentration in the sterilization chamber immediately before
the chamber exhaust is activated. The owner or operator shall install,
calibrate, operate, and maintain a gas chromatograph to measure ethylene oxide
concentration. The daily calibration requirements are required only on days
when the chamber exhaust is activated.
(5) At facilities using 20,000 pounds or more
of ethylene oxide per consecutive 12 months, seeking to comply with the
standard by manifolding emissions from the chamber exhaust vent to a control
device controlling emissions from another vent type (sterilization chamber vent
and/or aeration room vent), shall monitor the control device to which emissions
from the chamber exhaust vent are manifolded.
(g) Recordkeeping.
(1) The owner or operator of a sterilizer or
aerator subject to the emissions standards in subsection (d) Table I shall
maintain records of all reports and notifications (including compliance
notifications) in a form suitable and readily available for expeditious
inspection and review. The files shall be retained for at least 5 years
following the date of each occurrence, measurement, maintenance, corrective
action, report or record. At a minimum the most recent 2 years of data shall be
retained on site. The files shall contain:
(A)
The occurrence and duration of each malfunction of the air pollution control
equipment;
(B) All required
measurements needed to demonstrate compliance with the standard (including, but
not limited to, 15-minute averages of CMS data, raw performance testing
measurements, and raw performance evaluation measurements, that support data
that the source is required to report);
(C) All measurements as may be necessary to
determine the conditions of performance tests and performance evaluations;
and
(D) Any information
demonstrating whether a source is meeting the requirements for a waiver of
recordkeeping or reporting requirements.
(2) The source may apply for a waiver of
recordkeeping or reporting requirements by submitting a written application to
the Administrator. Until the waiver is granted, the source remains subject to
the requirements of this section. The application must contain at a minimum:
(A) A request for an extension of compliance
(if applicable);
(B) All required
compliance progress reports or compliance status reports;
(C) Any excess emissions and CMS performance
report; and
(D) Information to
convince the administrator that a waiver of recordkeeping or reporting is
warranted.
(h)
Reporting. Any person who owns or operates a sterilizer shall furnish the
following written report to the Administrator and to the district within thirty
days after the date specified by the district.
(1) An annual report that demonstrates that
the facility is a major or area source. The report shall contain at a minimum;
(A) the number of sterilizer cycles and the
pounds of ethylene oxide used per cycle for each sterilizer during the
consecutive 12-month reporting period from the district permit; or
(B) the total pounds of sterilant gas and the
total pounds of ethylene oxide purchased, used, and returned in the consecutive
12-months from the date of the permit.
(2) Facilities shall provide semi-annual
compliance reports to the Administrator that contain information on the
compliance status of the source. This report should also contain the summary
report in Appendix 1, (i). The report shall be signed by the responsible
official who shall certify its accuracy.
(i) Construction or Modification.
The requirements of this section apply to sources subject
to the emission standards in Table I. No person may construct or modify a
source, without obtaining written approval, in advance, from the district and
from the Administrator. For major sources, the application for approval of
construction or modification may be used to fulfill the notification
requirements. For specific requirements, see Appendix 2. In lieu of complying
with requirements in Appendix 2, a facility may fulfill these requirements by
complying with the permitting agency's new source review rule or policy,
provided similar information is obtained.