Fla. Admin. Code Ann. R. 64E-5.1103 - Radiation Survey Instruments
(1) The
licensee or registrant shall maintain sufficient calibrated and operable
radiation survey instruments at each field station and temporary jobsite to
make physical radiation surveys as required by this part and by Part III.
Instrumentation shall be capable of measuring 0.1 milliroentgen (0.001 mSv) per
hour through at least 50 milliroentgens (0.5 mSv) per hour.
(2) Radiation survey instruments used to
establish dose rates shall be calibrated:
(a)
At energies and geometries appropriate for use;
(b) At intervals not to exceed 6 months, and
after each instrument servicing;
(c) Such that accuracy within plus or minus
20 percent can be demonstrated; and,
(d) For linear scale instruments, at two
points located approximately 1/3 and 2/3 of full-scale on each scale; for
logarithmic scale instruments, at midrange of each decade, and at two points of
at least one decade; and for digital instruments, at appropriate
points.
(3) Records of
survey instrument calibrations shall be maintained for 3 years after the
calibration date for inspection by the Department.
Notes
Rulemaking Authority 404.051, 404.061, 404.081, 404.22 FS. Law Implemented 404.022, 404.051(1), (4), 404.061(2), 404.081(1), 404.22 FS.
New 7-17-85, Amended 4-4-89, Formerly 10D-91.1205, Amended 10-8-00.