In addition to other requirements of this part, all
diagnostic x-ray systems shall meet the following requirements:
(1) Warning Label. The main control panel and
all auxiliary control panels of the x-ray system shall bear the equivalent
warning statement, legible and accessible to view, "WARNING: This x-ray unit
may be dangerous to patient and operator unless safe exposure factors and
operating instructions are observed."
(2) Battery Charge Indicator. Visual means
shall be provided on the control panel of battery-powered x-ray generators to
indicate whether the battery is in a state of charge adequate for proper
operation.
(3) Leakage Radiation
from the Diagnostic Source Assembly. The leakage radiation from the diagnostic
source assembly measured at a distance of 1 meter in any direction from the
source shall not exceed 100 milliroentgens (25.8 uC per kg) in 1 hour when the
x-ray tube is operated at its leakage technique factors. Compliance shall be
determined by measurements averaged over an area of 100 square centimeters with
no linear dimension greater than 20 centimeters.
(4) Radiation from Components Other than the
Diagnostic Source Assembly. The radiation emitted by a component other than the
diagnostic source assembly shall not exceed 2 milliroentgens (0.516 uC per kg)
in 1 hour at 5 centimeters from any accessible surface of the component when it
is operated in an assembled x-ray system under any conditions for which it was
designed. Compliance shall be determined by measurements averaged over an area
of 100 square centimeters with no linear dimension greater than 20
centimeters.
(5) Beam Quality.
(a) Half-value Layer. The half-value layer
(HVL) of the useful beam for a given x-ray tube potential shall not be less
than the values shown below. If it is necessary to determine such half-value
layer at an x-ray tube potential which is not listed below, linear
interpolation or extrapolation may be made.
|
Design operating range (Kilovolts peak)
|
Measured potential (Kilovolts peak)
|
Half-value layer (Millimeters of aluminum)
|
|
Below 50
|
30
|
0.3
|
|
|
40
|
0.4
|
|
|
49
|
0.5
|
|
50 to 70
|
50
|
1.2
|
|
|
60
|
1.3
|
|
|
70
|
1.5
|
|
Above 70
|
71
|
2.1
|
|
|
80
|
2.3
|
|
|
90
|
2.5
|
|
|
100
|
2.7
|
|
|
110
|
3.0
|
|
|
120
|
3.2
|
|
|
130
|
3.5
|
|
|
140
|
3.8
|
|
|
150
|
4.1
|
1. The above
HVL criteria will be considered to have been met if it can be demonstrated that
the aluminum equivalent of the total filtration in the primary beam is not less
than that shown below:
|
Filtration Required vs. Operating
Voltage
|
|
Operating Voltage (kVp)
|
Total Filtration (inherent plus added) (millimeters
aluminum equivalent)
|
|
Below 50
|
0.5 millimeters
|
|
50-70
|
1.5 millimeters
|
|
Above 70
|
2.5 millimeters
|
2.
Beryllium window tubes shall have a minimum of 0.5 millimeters aluminum
equivalent filtration permanently mounted in the useful beam.
3. For capacitor energy storage equipment,
compliance shall be determined with the maximum quantity of charge per
exposure.
4. The required minimum
aluminum equivalent filtration shall include the filtration contributed by all
materials which are always present between the focal spot of the tube and the
patient.
5. In addition to the
requirements of paragraph (5)(a)1., above, all intraoral dental radiographic
systems manufactured on and after December 1, 1980, shall have a minimum
half-value layer not less than 1.5 millimeters aluminum equivalent filtration
permanently installed in the useful beam.
(b) Filtration Controls. For x-ray systems
which have variable kVp and changeable filters, and which are used for low
filtration techniques, a positive means shall be provided that will prevent an
exposure unless the minimum required amount of filtration is in the useful beam
for the selected kVp.
(6)
Aluminum Equivalent of Material between Patient and Image Receptor. The
aluminum equivalent of each of the items listed below, which are used between
the patient and image receptor, shall not exceed the indicated limits. This
requirement is applicable to the front panel of cassette holders and film
changers provided by the manufacturer for purposes of patient support or to
prevent foreign object intrusions. It does not apply to such items as a screen
and its associated mechanical support panel or grids.
|
Item
|
Maximum Aluminum Equivalent (millimeters)
|
|
Front panel of cassette holder (total of all)
|
1.0
|
|
Front panel of film changer (total of all)
|
1.0
|
|
Stationary tabletop
|
1.0
|
|
Moveable tabletop (including stationary
subtop)
|
1.5
|
|
Cradle
|
2.0
|
(7)
Multiple Tube Heads. Where two or more radiographic tubes are controlled by one
exposure switch, the tube or tubes which have been selected shall be clearly
indicated prior to initiation of the exposure. This indication shall be both on
the x-ray control panel and at or near the selected tube housing
assembly.
(8) Mechanical Support of
Tube Head. The tube housing assembly supports shall be adjusted such that the
tube housing assembly will remain stable during an exposure unless the tube
housing movement is a designed function of the x-ray system.
(9) Technique Indicators.
(a) Each x-ray system shall be equipped with
devices, such as labeled control settings or meters, correctly indicating the
physical factors and modes of operation used for exposures. X-ray systems
utilizing arbitrary number or letter designators for kVp, time and
milliamperage shall be accompanied by a chart giving the value of physical
factors for each arbitrary designator.
(b) The technique factors to be used during
an exposure shall be indicated before the exposure begins, except when
automatic exposure controls are used, in which case the technique factors which
are set prior to the exposure shall be indicated.
(c) On equipment having fixed technique
factors, the requirement in paragraph (9)(a), above, may be met by permanent
markings. Indication of technique factors shall be visible from the operator's
position except in the case of fluoroscopy or spot films made by the
fluoroscopist.
(d) Reproducible
technique factor indicators may be relabeled to meet the accuracy requirements
of this part. Such relabeling shall be conspicuous and clearly legible and
shall be utilized by the registrant in setting technique
factors.
(10) Accuracy of
Technique Factors. Meters, labeled control settings, exposure time selectors
and other physical factor indicators shall be accurate within the following
tolerances:
(a) Milliamperage (mA) -
Radiographic ± 10%
(b) Milliamperage (mA) - Fluoroscopic
± 0.2 mA
(c)
Kilovolt peak (kVp) ± 5 percent
(d) Timer at settings:
1. Greater than 10 seconds
± 1 second
2.
Ten seconds or less ± ten
percent
(11)
Timer Reproducibility. When four timer tests are performed at the same timer
settings, the average time period (T) shall be greater than or equal to 12
times the maximum time period (Tmax) less the minimum
time period (Tmin). T shall be less than or equal to 0.5
seconds. Expressed mathematically, T [GREATER THAN EQUAL TO]= 12
(Tmax - Tmin).
(12) Exposure Reproducibility. The x-ray
exposure produced by radiographic systems shall be reproducible to within the
following criteria: When all technique factors are held constant and four or
more exposures at the same technique factors are made, the value of the average
exposure (E) shall be greater than or equal to 12 times the quantity of maximum
exposure (Emax) minus the minimum exposure
(Emin). Expressed mathematically, E [GREATER THAN EQUAL
TO]= 12 (Emax - Emin).
(13) Exposure Linearity. The x-ray output
produced by radiographic systems utilizing means other than automatic exposure
controls shall be linear to within the following criteria:
(a) When a choice of two or more current
settings (mA) or current-time product settings (mAs) may be selected where
X
1 and X
2 are the average mR per
mAs values obtained from four exposures on each of two mA or mAs settings at a
fixed tube potential (kVp) setting, within the range of 40 to 100 percent of
the maximum tube rating, the average of four exposures (mR) for a given
milliampere-second (mAs) product in mR per mAs shall not differ by more than:
1. Five hundredths times the sum of any 2
consecutive mA or mAs averaged settings; expressed mathematically,
|X1 - X2 |[LESS
THAN EQUAL TO]= 0.05 (X1 + X2);
and,
2. One tenth times the
sum of any other 2 mA or mAs averaged settings; expressed mathematically,
|X1 - X2 |[LESS
THAN EQUAL TO]= 0.10 (X1 -
X2)
(b) Equipment, which after calibration cannot
be made to meet the requirements of paragraph (13)(a), above, may be relabeled
to indicate the effective mA or mAs, 8providing that use of such relabeled
stations will meet the requirements of paragraph (13)(a), above.
(c) Equipment, which after calibration cannot
be made to meet the requirements of paragraph (13)(a) or (b), above, shall not
be used unless written approval is obtained from the Department. Approval shall
not be granted when the linearity determination exceeds the federal standard
for certified systems.
(14) Automatic Exposure Controls. When
automatic exposure control is provided:
(a)
Indication shall be made on the control panel when this mode of operation is
selected.
(b) When the x-ray tube
potential is greater than or equal to 50 kVp, the minimum exposure time for
field emission equipment rated for pulsed operation shall be less than or equal
to a time interval equivalent to 2 pulses.
(c) The minimum exposure time for all
equipment other than that specified in paragraph (14)(b), above, shall be less
than or equal to 1/60 second or a time interval required to deliver 5 mAs,
whichever is greater.
(d) Either
the product of peak x-ray tube potential, current, and exposure time shall be
limited to not more than 60 kWs per exposure, or the product of x-ray tube
current and exposure time shall be limited to not more than 600 mAs per
exposure except when the x-ray tube potential is less than 50 kVp, in which
case the product of x-ray tube current and exposure time shall be limited to
not more than 2, 000 mAs per exposure.
(e) A visible signal shall indicate when an
exposure has been terminated at the limits described in paragraph (14)(d),
above, and manual resetting shall be required before further automatically
timed exposures can be made.
(f)
Exposure Linearity. When a choice of two or more tube current settings (mA) may
be selected, the average of four consecutive exposures
(E
1) made at any one tube current setting minus the
average of four consecutive exposures (E
n) made at any
other tube current setting shall be less than or equal to 0.05 times the sum of
the two averages. Expressed mathematically, |E
1 -
E
n |[LESS THAN EQUAL TO]= 0.05
(E
1 + E
n). Measuring compliance
for the above shall be based on the following criteria:
1. An attenuation block as described in
subsection
64E-5.501(5),
F.A.C., shall be in the useful beam.
2. Exposure (mR) shall be measured on the
exit side of the attenuation block.
3. The tube potential (kVp) shall be
maintained at a fixed setting within the range of 40 percent to 100 percent of
the maximum tube rating.
(15) Beam Limiting Devices.
(a) Beam limiting devices capable of
restricting the useful beam to the area of clinical interest shall be used
during exposures.
(b) Beam limiting
devices shall provide a degree of attenuation not less than that required for
the tube housing.
(16)
Remote Exposure Switches. Where an x-ray control is equipped with two or more
remote exposure switches, each remote switch shall serve a single x-ray tube,
and exposures with any tube shall be possible only by the remote switch with
which that particular tube is associated.
(17) Electrical Power Supply. The electrical
power supply and service lines to x-ray systems shall be of sufficient capacity
to permit operation without significant variation in voltage or machine
output.