Fla. Admin. Code Ann. R. 64E-5.703 - Operating Requirements
(1) Procedures.
Normal operating procedures shall be written and available to all analytical
x-ray equipment workers. No individual shall be permitted to operate analytical
x-ray equipment in any manner other than that specified in the procedures
unless such individual has obtained the written approval of the radiation
safety officer.
(2) Bypassing. No
individual shall bypass a safety device or interlock unless such individual has
obtained the written approval of the radiation safety officer. Such approval
shall be for a specified period of time. When a safety device or interlock has
been bypassed, a readily discernible sign bearing the words "SAFETY DEVICE NOT
WORKING", or words having a similar intent, shall be placed on the radiation
source housing.
(3) Repair or
Modification of X-ray Tube Systems. Except as specified in (2), above, no
operation involving removal of covers, shielding materials or tube housings or
modifications to shutters, collimators or beam stops shall be performed without
first ascertaining that the tube is off and will remain off until safe
conditions have been restored. The main switch, rather than interlocks, shall
be used for routine shutdown in preparation for repairs.
Notes
Rulemaking Authority 404.051, 404.081, 404.22 FS. Law Implemented 404.051(1), (4), (6), 404.081(1), 404.22 FS.
New 7-17-85, Formerly 10D-91.805.
State regulations are updated quarterly; we currently have two versions available. Below is a comparison between our most recent version and the prior quarterly release. More comparison features will be added as we have more versions to compare.
No prior version found.