(1) Scope. This part establishes
requirements, for which a registrant is responsible, for use of x-ray equipment
by or under the supervision of an individual authorized in accordance with
State statutes to engage in the healing arts. The provisions of this part are
in addition to, and not in substitution for, other applicable provisions of
these regulations.
(2) General
Requirements.
(a) Training of Operators who
Administer X-ray in the Healing Arts.
1. The
registrant shall assure the Department that all radiation machines and
associated equipment under his control are operated only by individuals
instructed in safe operating procedures.
2. The registrant shall require persons
operating his radiation machine and associated equipment to receive, at a
minimum, six hours of instruction. The following subject categories shall be
covered:
(i) Protection Against Radiation
(I) Protective Clothing
(II) Patient Holding
(III) Time, Distance, Shielding
(IV) Radiation Protection Standards
(ii) Dark Room Techniques
(I) Developing Chemicals
(II) Film Protection
(III) Cassettes
(IV) Screens
(iii) Patient Protection
(I) Beam Limitation
(II) Setting Up Techniques
(III) Biological Effects of
Radiation
(iv) Machine
Safety
(I) Machine Functions
(II) Safety Procedures
(III) Recognizing Problems
3. Instruction required
by .04(2)(a)2. shall begin within 30 days after employment and shall be
completed no later than 90 days after date of employment. The registrant shall
maintain a record of all training for each operator. Such record shall be made
available for Departmental inspection. This rule shall take effect 180 days
after the effective date of these regulations.
4. Persons who show written proof that they
have received the required instruction are considered to meet the requirements
of .04(2)(a)2.
(b) Except
for patients who cannot be moved out of the room, only the staff and ancillary
personnel required for the medical procedure or training shall be in the room
during the radiographic exposure. Other than the patient being examined:
1. All individuals shall be positioned such
that no part of the body will be struck by the useful beam, unless protected by
at least 0.5 millimeter meter lead equivalent material; and
2. Staff and ancillary personnel who must
remain in areas because of their required presence during an x-ray procedure,
shall be protected from direct scatter radiation by protective aprons or whole
body protective barriers of not less than 0.25 millimeter lead equivalent;
and
3. Patients, other than the one
being radiographed, who cannot be removed from the x-ray room shall be
protected by a barrier of at least .25 mm Pb equivalent or be at least 2 meters
from the tube head and the image receptor.
(c) Except for dental intraoral radiography,
veterinary, or portable x-ray use, the operator's position at the controls
shall be in a protected area that will meet the radiation protection
requirements of rule
.03(2)(a)1. of
these regulations.
(d) Except for
therapy exposures, gonad shielding of not less than 0.25 millimeter lead
equivalent shall be available and shall be used when the gonads are in the
useful beam except when its use will interfere with the diagnostic information
on the image receptor.
(e)
Individuals shall only be exposed to the useful beam for healing arts purposes
except as required by law enforcement officials or their designated
representatives in the interest of public safety. This provision specifically
prohibits deliberate exposure of persons for non-productive x-ray procedures
such as for training, demonstration, or for other non-healing arts
purposes.
(f) When a patient or
film must be provided with auxiliary support during a radiation exposure:
1. Mechanical holding devices shall be used
when the technique permits. Holding shall be used only when other means of
support cannot be utilized.
2. No
individual shall be used routinely to hold film or patients.
3. When holding is required, the person
holding shall be provided with protective clothing and shall be positioned so
that no part of the body is struck by the useful beam.
(g) Portable equipment shall be used only for
examinations where it impractical, for medical purposes, to transfer the
patient to the x-ray suite.
(3) Information and Maintenance Records and
Associated Information.
(a) The registrant
shall maintain the following information for each radiation machine for
inspection by the Department:
1. Model and
serial numbers of x-ray tube housing and generator; and
2. Records of surveys, calibrations,
maintenance, and modifications performed on the radiation machine(s) with the
names of persons who performed such services.
(b) The vendor shall supply the registrant
with a record of all maintenance performed, or parts replaced or installed,
written in a clear and legible manner.
(4) Light Fields. When used for aligning or
centering an x-ray field, a light field shall have a clearly defined perimeter
and have illumination intensity equal to the needs for collimation or
alignment. For collimators equipped with beam defining lights, this requirement
will be deemed to be met if the illumination at the receptor is visible to the
x-ray operator under normal room illumination in all quadrants of the light
field.
(5) Darkroom and Film
Processors.
(a) Darkrooms used for film
processing and/or developing shall be light tight.
(b) Each darkroom shall be equipped with a
safelight which will meet or exceed the requirements of the radiographic film.
This will be deemed to have been met if the film manufacturer's recommendations
are followed.
(c) Except for
automatic developing systems, each darkroom shall have and use a solution
thermometer and timing device. Sight development shall be prohibited.
(d) The chemical solution used for manual
film development shall not be used for periods in excess of two (2) months.
Records of solution changes shall be maintained.
(e) When automatic film processing is used it
shall be maintained in accordance with the manufacturer's recommendations and a
record of cleaning and developer change shall be maintained.
(f) Unexposed film shall not be subject to
radiation levels in excess of 0.2 mR during the period of storage.
(g) Unexposed film which is outdated shall
not be used for human radiographic procedures.
(6) General Requirements for all Diagnostic
Radiation Machines. In addition to other requirements of this part, all
diagnostic radiation machines shall meet the following requirements:
(a) Warning Label. The control panel
containing the main power switch shall bear the following warning statement, in
a manner legible and accessible to view:
"WARNING: This x-ray unit may be dangerous to patient and
operator unless safe exposure factors and operating instructions are
observed"
(b) Battery Charge
Indicator. On battery-powered x-ray generators, visual means shall be provided
on the control panel to indicate whether the battery is in a state of charge
adequate for proper operation.
(c)
Leakage Radiation from the Diagnostic Source Assembly. The leakage radiation
from the diagnostic source assembly, measured at a distance of 1 meter in any
direction from the source, shall not exceed 100 milliroentgens in 1 hour when
the x-ray tube is operated at its leakage technique factors.
(d) Beam Quality.
1. Half-value layer.
(i) The half-value layer of the useful beam
for a given x-ray tube potential shall be no less than the values shown in
Table I. If it is necessary to determine the half-value layer at an x-ray tube
potential which is not listed in Table I, linear interpolation or extrapolation
may be made.
TABLE I
|
Design Operating Range (Kilovolts Peak)
|
Measured Potential (Kilovolts Peak)
|
Half- value layer (Millimeters of Aluminum)
|
|
Below 50
|
30
|
0.3
|
|
40
|
0.4
|
|
49
|
0.5
|
|
50-70
|
50
|
1.2
|
|
60
|
1.3
|
|
70
|
1.5
|
|
Above 70
|
71
|
2.1
|
|
80
|
2.3
|
|
90
|
2.5
|
|
100
|
2.7
|
|
110
|
3.0
|
|
120
|
3.2
|
|
130
|
3.5
|
|
140
|
3.8
|
|
150
|
4.1
|
(ii)
The requirements of .04(6)(d)1. (i) will be considered to have been met if it
can be demonstrated that the aluminum equivalent of the total filtration in the
primary beam is not less than that shown in Table II.
TABLE II
Filtration Required vs. Operating Voltage
|
Operating Voltage (kVp)
|
Total Filtration
(inherent plus added)
(millimeters aluminum equivalent)
|
|
Below 50
|
0.5 millimeters
|
|
50 to 70
|
1.5 millimeters
|
|
Above 70
|
2.5 millimeters
|
(iii)
Beryllium window tubes shall have a minimum of 0.5 millimeter aluminum
equivalent filtration permanently installed in the useful beam.
(iv) For capacitor energy storage equipment,
compliance with the requirements of .04(6)(d)1. (i) shall be determined with
the maximum quantity of charge per exposure.
(v) The required minimal aluminum equivalent
filtration shall include the filtration contributed by all materials which are
always present between the source and the patient.
2. Filtration Controls. For radiation
machines which have variable kVp and variable filtration for the useful beam, a
device shall link the kVp selector with the filter(s) and shall prevent an
exposure unless the minimum amount of filtration required by .04(6)(d)1. (i) or
(ii) is in the useful beam for the given kVp which has been selected.
(7) Multiple
Tubes. Where two or more radiographic tubes are controlled by one exposure
switch, the tube or tubes which have been selected shall be clearly indicated
prior to initiation of the exposure. This indication shall be both on the x-ray
control panel and at or near the tube housing assembly which has been
selected.
(8) Mechanical Support of
Tube Head. The tube housing assembly supports shall be adjusted such that the
tube housing assembly will remain stable during an exposure unless the tube
housing movement is a designed function of the radiation machine.
(9) Technique Indicators. The technique
factors to be used during an exam shall be indicated prior to any exposure.
This requirement may be met by permanent markings on equipment having fixed
technique factors. Indication of technique factors shall be visible from the
operator's position except in the case of spot films made by the
fluoroscopist.
(10) Exposure
Timing.
(a) Except in fluoroscopy a device
shall be used to terminate and accurately reproduce the exposure at a preset
time interval, preset product of current and time, a preset number of pulses,
or a preset radiation exposure to the image receptor.
(b) Except for fluoroscopy, dental intraoral
and panographic, veterinary, and procedures requiring the use of portable
barriers, the exposure switch shall be so located that it cannot be
conveniently operated outside of a shielded area.
(c) Except during serial radiography, the
operator shall be able to terminate the exposure at any time during an exposure
of greater than one-half second.
(d) During serial radiography, the operator
shall be able to terminate the x-ray exposure(s) at any time, but means may be
provided to permit completion of any single exposure of the series in
process.
(e) Automatic exposure
controls.
1. When an automatic exposure
control is provided, indication shall be made on the control panel when this
mode of operation is selected.
2.
When an automatic exposure control is provided, a backup timer shall be
required. The backup timer shall be capable of terminating the exposure at a
preset time should the automatic exposure control fail. The preset time shall
be consistent with the technique used.
(f) The x-ray production shall be controlled
by a deadman switch.
(g) It shall
not be possible to make an exposure when the time is set to a zero or off
position if either position is provided.
(h) Termination of an exposure shall cause
automatic resetting of the timing device to its initial setting or to
zero.
(11) Hand-held
fluoroscopic screens are prohibited except for law enforcement or forensic
requirements, and then only upon approval by the Department.
(12) Fluoroscopic Radiation Machines. All
fluoroscopic radiation machines shall meet the following requirements:
(a) Limitation of Useful Beam.
1. Primary Barrier.
(i) Image intensification shall be used with
all fluoroscopic machines. The fluoroscopic imaging assembly shall be provided
with a primary protective barrier which intercepts the entire cross section of
the useful beam at any SID.
(ii)
The x-ray tube used for fluoroscopy shall not produce x-rays unless the barrier
and image intensifier are in position to intercept the entire useful
beam.
2. X-Ray Field.
(i) For image-intensified fluoroscopic
equipment, neither the length nor the width of the plane of the image receptor
shall exceed that of the visible area of the image receptor by more than 3
percent of the SID. The sum of the excess length and the excess width shall be
no greater than 4 percent of the SID.
(b) Spot film devices which are certified
components shall meet the following additional requirements:
1. Means shall be provided between the source
and the patient for adjustment of the x-ray field size, in the plane of the
film, to the size of that portion of the film which has been selected on the
spot film selector; and
2. It shall
be possible to adjust the x-ray field size in the plane of the film to a size
smaller than the selected portion of the film; and
3. The center of the x-ray field in the plane
of the film shall be aligned with the center of the selected portion of the
film to within 2 percent of the SID.
(c) Pre-certified fluoroscopic machines are
exempt from the requirements of .04(12)(a) and .04(12)(b) provided that:
1. The machine was in service prior to the
date of adoption of these regulations and meets all other applicable
requirements for fluoroscopic machines. However, these machines shall be
brought up to standards referenced in .04(12)(a) and (b) within three years
from the date of adoption of these regulations or be taken out of service and
electronically disabled.
2. The
shutter mechanism is adjusted so that the x-ray field diameter is limited to
the dimensions of the film cassette used during spot filming at a 35
centimeters (14 inches) table-to-image-receptor distance.
3. When spot films are either unnecessary or
not required during a portion of the exam, the leading edge of the shutters
shall be restricted to the edge of the image intensifier.
(d) Activation of the Fluoroscopic Tube.
X-ray production in the fluoroscopic mode shall be controlled by a dead man
switch. When recording serial fluoroscopic images, the fluoroscopist shall be
able to terminate the x-ray exposure(s) at any time, but means may be provided
to permit completion of any single exposure of the series in process.
(e) Exposure Rate Limits.
1. Entrance Exposure Rate Allowable Limits.
(i) When the automatic brightness control is
used, the exposure measured at the point where the center of the useful beam
enters the patient shall not exceed 10 roentgens per minute, except during
recording of fluoroscopic images.
(ii) When provided with optional high level
control, the equipment shall not be operable at any combination of tube
potential and current which will result in an exposure rate in excess of 5
roentgens per minute at the point where the center of the useful beam enters
the patient unless the high level control is activated.
(I) Special means of activation of high level
controls shall be required. The high level control shall only be operable when
a continuous secondary level of pressure is provided by the operator.
(II) When the high level control is activated
the entrance exposure rate shall not exceed 10 R/min. except in the recording
of fluoroscopic images.
(III) A
continuous signal audible to the fluoroscopist shall indicate that the high
level control is being employed.
(iii) In addition to the other requirements
of .04(12)(e)1. (i) and (ii), certified equipment which does not incorporate an
automatic exposure control shall not be operable at any combination of tube
potential and current which will result in an exposure rate in excess of 5
roentgens per minute at the point where the center of beam enters the patient,
except during recording of fluoroscopic images or when provided with an
optional high level control.
(iv)
Non-certified equipment shall not operate at any combination of tube potential
and current which will result in an exposure in excess of 10 R/min.
2. Compliance with the
requirements .04(12)(e)1. shall be determined as follows:
(i) Movable grids and compression devices
shall be removed from the useful beam during the measurement;
(ii) With the source below the table, the
exposure rate shall be measured 1 centimeter above the tabletop or
cradle;
(iii) With the source above
the table, the exposure rate shall be measured at 30 centimeters above the
tabletop with the end of the beam-limiting device or spacer positioned as
closely as possible to the point of measurement;
(iv) In a C-arm type of fluoroscope, the
exposure rate shall be measured 30 centimeters from the input surface of the
fluoroscopic imaging assembly.
(f) Periodic measurement of Entrance Exposure
Rate. The registrant shall cause periodic measurement of entrance exposure
rate, including the exposure rate at staff positions around the table and
panel, to be made for each fluoroscope by an individual competent to make such
measurements. Results of these measurements shall be posted where any
fluoroscopist may have ready access to them. An adequate period for such
measurements shall be annually or after any maintenance of the unit if such
maintenance might affect the exposure rate. Results of the measurements shall
include the maximum possible R/minute of the fluoroscope at the maximum kVp and
mA used. The posted data shall indicate the technique factors used to determine
the data along with the name of the person and/or company performing the
measurements and the date the measurements were performed.
1. Fluoroscopes that incorporate automatic
exposure control shall have sufficient material placed in the useful beam to
produce a milliamperage typical of the use of the x-ray machine; and
2. Fluoroscopes that do not incorporate an
automatic exposure control shall utilize a milliamperage typical of the
clinical use of the radiation machine.
(g) Barrier Transmitted Radiation Rate
Limits. The exposure rate due to transmission through the primary protective
barrier with the attenuation block in the useful beam shall not exceed 2
milliroentgens per hour at 10 centimeters from any accessible surface of the
fluoroscopic imaging assembly beyond the plane of the image receptor for each
roentgen per minute of entrance exposure rate.
(h) Indication of Potential and Current.
During fluoroscopy and cine-fluorography, the kV and the mA shall be
continuously indicated.
(i)
Source-Skin Distance. The source to skin distance shall not be less than:
1. 38 centimeters (15 inches) on stationary
fluoroscopes;
2. 30 centimeters (12
inches) on all mobile fluoroscopes;
3. 20 centimeters (8 inches) for image
intensified fluoroscopes, used for specific surgical application;
4. 30 centimeters (12 inches) on stationary
precertified fluoroscopes.
(j) Fluoroscopic Timer. Means shall be
provided to preset the cumulative "on" time of the fluoroscopic tube. The
maximum cumulative time of the timing device shall not exceed 5 minutes without
resetting. Termination of the exposure or a signal audible to the fluoroscopist
shall indicate the completion of any preset cumulative on time. Such signal
shall continue to sound while x-rays are produced or until the timing device is
reset. Audible signals are recommended.
(k) Radiation Therapy Simulation Machines.
Radiation therapy simulation shall be exempt from all the requirements of
.04(12)(a), .04(12)(e), .04(12)(f) and of .04(12)(j) provided that:
1. Such machines are designed and used so
that no individual other than the patient is in the x-ray room during
radiography procedures; and
2. Such
machines which do not meet the requirements of .04(12)(j) are provided with a
means of indicating the cumulative exposure time for each individual patient.
Procedures shall require in each case that the timer be reset between
examinations.
(13) Radiographic Machines Other Than
Fluoroscopic, Dental Intraoral, and Veterinary.
(a) Beam Limitation. The useful beam shall be
limited to the area of clinical interest, and shall not be greater than the
dimensions of the image receptor.
(b) General Purpose Stationary and Mobile
Radiation Machines.
1. Means for stepless
independent adjustment in both the longitudinal and transverse direction of the
x-ray field and a light for visually defining the perimeter of the x-ray field
shall be provided.
2. Means shall
be provided to permit adequate light intensity at the film plane when the light
field intersects with the image receptor at a 100 cm SID. This will be deemed
to be met if a visual outline of the light field is visible at the
receptor.
3. Congruence of the
x-ray and light fields shall not have a misalignment in excess of 2% of the SID
in any one direction and not more than 3% of the SID when measured as the sum
of the absolute misalignment in the longitudinal and transverse
direction.
(c) Additional
Requirements for Stationary General Purpose Radiation Machines. In addition to
the requirements of .04(13)(b), all stationary radiation machines shall meet
the following requirements:
1. Means shall be
provided to indicate when the axis of the x-ray beam is perpendicular to the
film plane with respect to the center of the image receptor to within 2 percent
of the SID, and to indicate the SID to within 2 percent; and
2. The beam-limiting device shall numerically
indicate the field size in the plane of the image receptor to which it is
adjusted.
(d) Machines
Designed for or Provided With Special Attachments for Mammography.
1. Radiographic machines designed only for
mammography and general purpose radiographic machines, when special attachments
for mammography are in service, shall be provided with means to limit the
useful beam such that the x-ray field at the plane of the image receptor does
not extend beyond any edge of the image receptor at any designated SID, except
the edge of the image receptor designed to be adjacent to the chest wall where
the x-ray field may not extend beyond this edge by more than 2 percent of the
SID.
2. This requirement can be met
with a machine which performs as prescribed in .04(13)(e).
3. Each image receptor support intended for
installation on a system designed only for mammography shall have clear and
permanent markings to indicate the maximum image receptor size for which it is
designed.
(e) Special
Purpose Radiation Machines. Radiation machines which are limited by design to
radiographic examinations of a specific anatomical region shall meet the
following requirements:
1. The x-ray field in
the plane of the image receptor shall be limited such that the field does not
exceed each dimension of the image receptor by more than 2 percent of the SID
when the axis of the x-ray beam is perpendicular to the plane of the image
receptor;
2. The center of the
x-ray field shall be aligned with the center of the image receptor to within 2
percent of the SID, or means shall be provided to both size and align the x-ray
field such that the x-ray field at the plane of the image receptor does not
extend beyond any edge of the image receptor;
3. Section
.04(13)(e)2. may
be met with a machine that meets the requirements for a general purpose
radiation machine as specified in .04(13)(b) or, when alignment means are also
provided;
4. For special purpose
cephalometric use, an assortment of removable, fixed-aperture, beam-limiting
devices sufficient to limit the beam to areas of clinical interest may be used.
Each such device shall have clear and permanent markings to indicate the image
receptor size and the SID for which it is designed;
5. Special purpose radiographic units will be
exempt from the primary barrier requirements of .01(8)(b)1. provided that the
tube housing assembly is electronically interlocked to a primary protective
barrier, or the tube housing assembly is mechanically fixed such that the
entire cross section of the useful beam is always intercepted by a primary
barrier sufficient to attenuate the useful beam to the limits specified in
.03(2). Secondary barriers shall meet the shielding requirements of
.01(8)(b)1.
(f) Radiation
Exposure Control Device.
1. Each x-ray
control shall meet the following requirements:
(i) stationary radiation machines shall have
the exposure switch permanently mounted in such a way as to prevent the
operator from leaving the protected area of the operator's barrier during the
exposure;
(ii) except for unique
situations such as those found in intensive care units or operating room
suites, mobile and portable radiation machines which are used for greater than
1 week in 1 location, (i.e., 1 room or suite) shall meet the requirements of
.04(13)(f)1. (i).
(iii) The x-ray
control device shall provide audible or visual indication observable at or from
the operator's protected position whenever x-rays are produced. For certified
radiation machines, a signal audible to the operator shall indicate that the
exposure has terminated.
2. Portable Equipment.
(i) Provisions of .04(13)(f) apply except for
exposure switch location.
(ii) The
exposure switch shall be so arranged that the operator can stand at least 1.8
meters (six feet) from the patient, the x-ray tube, and the useful beam unless
there is shielding sufficient to assure compliance with .03(2)(a).
(iii) The source-to-skin distance shall be
limited to not less than 30 centimeters (12 inches).
(iv) Protective aprons of at least 0.25 mm
lead equivalent shall be available and their use shall be required of the
operator.
(v) Personnel monitoring
is required of all operators.
(vi)
Mobile or portable Radiation machines which are used for greater than one week
in one location, i.e., (one room or suite of rooms) shall meet the requirements
of .01(8).
(g)
Structural Shielding.
1. In addition to the
requirements in .01(8), diagnostic radiation machines routinely used in one
location shall meet the following requirements for structural shielding:
(i) All areas of the walls, floors, and
ceiling exposed to the primary beam shall have primary barriers; and
(ii) Secondary protective barriers shall be
provided in all wall, floor, and ceiling areas not having primary barriers or
where primary barrier requirements are less than secondary barrier
requirements.
2. For
stationary radiation machines and mobile or portable equipment routinely used
in one location:
(i) Except for those unique
situations found in such uses as the intensive care unit, operating suite,
etc., the operator's station at the controls shall be behind a protective
barrier which will intercept any radiation that has been scattered only
once.
(ii) The operator's
protective barrier shall be equipped with a glass window of lead equivalency
equal to that required of the adjacent barrier, or a mirror system so placed
that the entire patient can be seen by the operator while the exposure is
made.
(iii) Facilities constructed
or modified after the effective date of these regulations shall have built-in
operator's protective barriers which will ensure that the limits specified in
.03(2)(a) are not exceeded.
(h) Source-to-Skin Distance.
1. All radiographic machines, except as
provided for in .04(13) (h)2., shall be provided with means to limit the
source-to-skin distance to not less than 30 centimeters (12 inches).
2. A radiographic machine intended for
specific surgical and dental application may be used with an SSD less than 30
centimeters, (12 inches), but in no case less than 20 centimeters (8
inches).
(i) Radiation
from Capacitor Energy Storage Equipment in Standby Status. Radiation emitted
from the x-ray tube when the exposure switch or timer is not activated shall
not exceed a rate of 2 milliroentgens per hour at 5 centimeters from any
accessible surface of the diagnostic source assembly, with the beam-limiting
device fully open.
(14)
Intraoral Dental Radiographic Machines.
(a)
Source-to-Skin Distance. Radiation machines designed for use with an intraoral
image receptor shall be provided with means to limit source-to-skin distance,
i.e., SSD, to not less than 18 centimeters (7 inches), if operable above 50
kVp, or 10 centimeters (4 inches), if not operable above 50 kVp.
(b) Field Limitation.
1. Radiographic machines designed for use
with an intra-oral image receptor shall be provided with means to limit the
x-ray beam such that the diameter of the useful beam at the end of the cylinder
shall not be greater than 7.0 centimeters (2.75 inches). For intraoral
rectangular collimation the useful beam at the end of the spacer shall not have
a diagonal measurement greater than 7.0 centimeters (2.75 inches). Positioning
devices should be used to assure beam alignment.
2. An open ended shielded cylinder, or other
open ended shielded spacers that will meet the requirements of .04(14)(a) and
(b)1. shall be used.
(c)
Structural Shielding.
1. The provisions of
.01(8) shall apply, except that National Council on Radiation Protection and
Measurements Report No. 35, "Dental X-Ray Protection," or its current revision
or replacement, shall be referenced by the Department.
2. When dental x-ray units are installed in
adjacent rooms or areas, protective barriers sufficient to reduce the exposure
to the requirements of .03(2) shall be provided between the rooms and/or
areas.
(d) Operating
Procedures.
1. Patient and film holding
devices shall be used when the techniques permit.
2. Neither the tube housing nor the position
indicating device shall be hand-held during an exposure.
3. Mechanical support of the tube head shall
maintain the exposure position without drift.
4. Dental fluoroscopy shall not be used
without image intensification and shall meet the requirements of
.04(12).
5. Only persons required
for the radiographic procedure shall be in the x-ray room during exposure. All
persons shall be adequately protected.
6. The operator shall be able to view the
patient during an exposure.
7.
During each exposure, the operator shall stand at least 1.8 meters (6 feet)
from the patient and tube head and outside the path of the useful beam or
behind a barrier that meets the requirements of .03(2).
(e) The total filtration in the useful beam
shall not be less than the appropriate values stated in .04(6)(d)1.(i) or
(ii).
(15) Veterinary
Radiographic Installations.
(a) Equipment.
1. The tube housing shall be of the
diagnostic type.
2. The primary
beam for diagnostic purposes in radiography and fluoroscopy should not be
larger than clinically necessary and shall not be greater than the image
receptor. Cones, diaphragms, or adjustable collimators capable of restricting
the primary beam to the area of clinical interest shall be used and shall
provide the same degree of protection as is required in the tube
housing.
3. Means shall be provided
to terminate the exposure at a preset time interval, preset product of current
and time, a preset number of pulses, or a preset radiation exposure to the
image receptor.
4. The exposure
switch shall be of a dead-man type.
5. The total filtration permanently in the
useful beam shall not be less than the appropriate value stated in
.04(6)(d)1.(i) or (ii).
6. A means
shall be provided for aligning the center of the x-ray beam with the center of
the image receptor prior to an x-ray examination.
7. An easily discernible indicator which
shows whether or not x-rays are being produced shall be on the control
panel.
8. The installation shall be
so arranged that the operator can stand at least six feet from the animal, the
x-ray tube and out of the useful beam.
9. Leaded gloves and aprons shall be
available for use, and shall be used by all personnel in the room during an
exposure.
10. The effectiveness of
protective equipment (i.e., gloves, aprons, etc.), shall not be
impaired.
(b) Operating
Procedures.
1. Only persons whose presence is
necessary shall be in the radiographic area during exposure. Protective
clothing of at least 0.25 mm lead equivalent shall be provided and shall be
worn by all individuals required to be in controlled areas, except when the
individuals are entirely behind protective barriers while the equipment is
energized.
2. Patient support:
(i) When an animal patient or film must be
held in position for radiography, mechanical supporting or restraining devices,
or other means of immobilization, shall be used unless human holding is
required by the technique.
(ii) If
an animal patient must be held or positioned manually, the individual holding
the animal shall wear protective gloves having at least 0.5 mm lead equivalency
and a protective apron of at least 0.25 mm lead equivalency;
(iii) Personnel monitoring devices shall be
used if radiation measurements indicate potential exposure in excess of 25
percent of the applicable values specified in Section
.03(2)(a)1. to the
head, or trunk of the body.
(c) Fluoroscopy.
1. The provisions of .04(12) shall apply to
fluoroscopic equipment.
(d) Structural Shielding. The provisions of
.01(8) shall apply except that the National Council on Radiation Protection and
Measurements Report No. 36, "Radiation Protection in Veterinary Medicine," or
its current revision or replacement, shall be referenced by the
Department.
(16)
Therapeutic Radiation Machines of Less Than One MeV.
(a) Leakage Requirements. When the tube is
operated at its leakage technique factors, the leakage radiation shall not
exceed the value specified, at the distance specified for the classification of
that radiation machine.
1. Contact Therapy
Machines. Leakage radiation shall not exceed 100 milliroentgens (.0258 mC/Kg)
an hour at five (5) centimeters from the surface of the tube housing
assembly.
2. 0-150 kVp Machines.
(i) Machines which were manufactured or
installed prior to the date of adoption of these regulations shall not permit
radiation leakage in excess of 1 Roentgen (.258 mC/Kg) in one (1) hour at one
(1) meter from the source.
(ii) In
machines manufactured on, or after the date of adoption of these regulations,
leakage radiation shall not exceed 100 mR (.0258 mC/Kg) in one (1) hour at one
(1) meter from the source.
3. 151 to 999 kVp Systems. The leakage
radiation does not exceed one (1) roentgen (.258 mC/Kg) in one (1) hour at one
(1) meter from the source except systems that operate in excess of 500 kVp may
have a leakage radiation at 1 meter from the source not to exceed 0.1 percent
of the useful beam one meter from the source.
(b) Permanent Beam Limiting Devices. The
registrant shall be responsible for assuring that permanent fixed diaphragms or
cones used for limiting the useful beam shall provide at least the same
protection as required by the tube housing assembly.
(c) Removable and Adjustable Beam Limiting
Devices.
1. Removable beam limiting devices
shall, for the portion of the useful beam to be blocked by these devices,
transmit not more than 1 percent of the useful x-ray beam at the maximum
kilovoltage and with maximum treatment filter.
2. Adjustable beam limiting devices installed
after the effective date of these regulations shall transmit not more than 1
per cent of the useful x-ray beam.
3. Adjustable beam limiting devices installed
before the effective date of these regulations shall transmit not more than 5
percent of the useful x-ray beam.
(d) Filter System.
1. The filter system shall be so designed
that the filters cannot be accidentally displaced from the useful beam at any
possible tube orientation; and
2.
The radiation at 5 centimeters from the filter insertion slot opening does not
exceed 30 roentgens (7.74 mC/Kg) per hour under any operating conditions;
and
3. Each filter shall be
conspicuously inscribed as to its material of construction and its thickness.
For wedge filters, the wedge factor and wedge angle shall appear on the wedge
or wedge tray.
(e) Tube
Immobilization. The tube housing assembly shall be capable of being immobilized
during stationary treatments.
(f)
Focal Spot Marking. The tube housing assembly shall be so marked that it is
possible to determine the location of the focal spot to within 5 millimeters,
and such marking shall be readily accessible for use during calibration
procedures.
(g) Timer.
1. A timer shall be provided which has a
display at the treatment control panel. The timer shall be graduated in minutes
and fractions of minutes.
2. The
timer shall have a preset time selector and an elapsed time
indicator.
3. The timer shall be a
cumulative timer which activates with the radiation and retains its reading
after irradiation is interrupted or terminated.
4. After irradiation is terminated and before
irradiation can be reinitiated, it shall be necessary to cycle the present time
selector through zero time.
5. The
timer shall permit accurate presetting and determination of exposure times as
short as 1 second.
6. The timer
shall not permit an exposure if set at zero.
7. The timer shall not activate until the
shutter is opened when irradiation is controlled by a shutter
mechanism.
(h) Control
Panel Functions. The control panel, in addition to the displays required in
other provisions of .04(16), shall have:
1. an
indication of x-ray production; and
2. means for indicating kV and x-ray tube
current; and
3. means for
terminating an exposure at any time; and
4. a locking device which will prevent
unauthorized use of the radiation machine; and
5. for radiation machines installed after the
date of adoption of these regulations, a positive display of specific filter(s)
in the beam.
(i)
Source-to-Skin Distance. There shall be means of determining the SSD distance
to within 1 centimeter.
(j) Low
Filtration X-Ray Tubes. Each radiation machine equipped with a beryllium or
other low-filtration window shall be clearly labeled as such upon the tube
housing assembly and at the control panel.
(k) Calibrations and Spot Checks.
1. Calibrations.
(i) The calibration of therapeutic radiation
machines shall be performed at intervals not to exceed 1 year and after any
change or replacement of components which could cause a change in the radiation
output.
(ii) The registrant shall
insure that such calibration is performed by an individual competent to perform
such work.
(iii) Records of
calibrations performed shall be maintained by the registrant for at least 5
years after completion of the calibration.
(iv) A copy of the most recent radiation
machine calibration shall be available at the control panel.
(v) The radiation machine shall not be used
in the administration of radiation therapy unless the calibrations required by
.04(16)(k)1. (i) -(iv) have been met.
2. Spot Calibration Checks. Spot calibration
checks on radiation machines capable of operation at greater than 150 kVp shall
be performed in accordance with written procedures. A record of such checks
shall be maintained for a two (2) year period after completion of the
spot-check measurements.
(17) Additional Facility Design Requirements
for Therapy Radiation Machines Capable of Operating Above 50 kVp and less than
1 MeV.
(a) Voice Communication. Provision
shall be made for two-way aural communication between the patient and the
operator at the control panel.
(b)
Viewing Systems.
1. Windows, mirrors,
closed-circuit television, or an equivalent system shall be provided to permit
continuous observation of the patient during irradiation and shall be so
located that the operator can observe the patient from the control
panel.
2. When the primary viewing
system is by electronic means, an alternate viewing system, which may also be
electronic, shall be available for use in the event of electronic
failure.
3. In the event of total
failure of patient viewing, therapy shall be discontinued until the system is
functioning.
(c)
Structural Shielding. In addition to the provisions of .01(8):
1. For existing equipment operating above 125
kVp the required operator's barrier(s) shall be an integral part of the
building;
2. For all therapeutic
machines operating below 150 kVp, built or modified after the effective date of
these regulations, the operator's barrier(s) shall be an integral part of the
building;
3. For equipment
operating above 150 kVp, the control panel shall be within a protective booth
equipped with an interlocked door, or located outside the treatment
room.
(d) Additional
Requirements for Radiation Machines Capable of Operation Above 150 kVp and less
than 1 MeV.
1. All necessary shielding,
except for any beam interceptor, shall be provided by fixed barriers;
2. The control panel shall be outside the
treatment room;
3. All doors of the
treatment room shall be electrically connected to the control panel such that
x-ray production cannot occur unless all doors are closed;
4. When the treatment room door is opened
during any exposure, the exposure shall terminate immediately;
5. After termination of the exposure, it
shall be possible to restore the radiation machine to full operation only upon
closing the door, and subsequently reinitiating the exposure at the control
panel.
(e) Operating
Procedures.
1. Therapeutic radiation machines
shall not be left unattended unless the machine is secured.
2. When a patient must be held in position
for radiation therapy, mechanical supporting or restraining devices shall be
used.
3. The tube housing assembly
shall not be held by an individual during exposures.
4.
(i) For
radiation machines operating above 150 kVp, no individual other than the
patient shall be in the treatment room during exposures.
(ii) For machines operating below 150 kVp, no
individual other than the patient shall be in the treatment room unless such
individual is protected by a barrier sufficient to meet the requirements of
these regulations.
(18) X-Ray and Electron Therapy Machines with
Energies of One MeV and Above.
(a) Scope. This
part applies to medical facilities using therapy machines with energies of 1
MeV and above. Additional requirements for these machines are found in Section
111-8-90-.05 entitled "Radiation
Safety Requirements for Particle Accelerators".
(b) Requirements for Equipment.
1. Leakage Radiation to the Patient Area.
(i) New equipment shall meet the following
requirements:
(I) For operating conditions
producing maximum leakage radiation, the absorbed dose in rads (grays) due to
leakage radiation, including x-rays, electrons, and neutrons, at any point in a
circular plane of 2 meters radius centered on and perpendicular to the central
axis of the beam at the isocenter or normal treatment distance and outside the
maximum useful beam size shall not exceed 0.1 percent of the maximum absorbed
dose in rads (grays) of the unattenuated useful beam measured at the point of
intersection of the central axis of the beam and the plane surface.
Measurements excluding those for neutrons shall be averaged over an area up to
but not exceeding 100 square centimeters at the positions specified.
Measurements of the portion of the leakage radiation dose contributed by
neutrons shall be averaged over an area up to but not exceeding 200 square
centimeters.
(II) For each machine,
the registrant shall determine or obtain from the manufacturer the leakage
radiation existing at the positions specified in (.04)(18)(b)l. (i) (I) for the
specified operating conditions. Records of leakage radiation measurements shall
be maintained for inspection by the Department.
(ii) Existing equipment shall meet the
following requirements:
(I) For operating
conditions producing maximum leakage radiation, the absorbed dose in rads
(grays) due to leakage radiation excluding neutrons at any point in a circular
plane of 2 meters radius centered on a perpendicular to the central axis of the
beam 1 meter from the virtual source, and outside the maximum size useful beam,
shall not exceed 0.1 percent of the maximum absorbed dose in rads (grays) of
the unattenuated useful beam measured at the point of intersection of the
central axis of the beam and the surface of the circular plane. Measurements
shall be averaged over an area up to but not exceeding 100 square centimeters
at the positions specified.
(II)
For each machine, the registrant shall determine or obtain from the
manufacturer the leakage radiation existing at the positions specified in
.04(18)(b)1. (ii)(I) for the specified operating conditions. Records of leakage
radiation measurements shall be maintained for inspection by the
Department.
2.
Leakage Radiation Outside the Patient Area for New Equipment.
(i) The absorbed dose in rads (grays) due to
leakage radiation except in the area specified in .04(18)(b)1. (i) (I) when
measured at any point 1 meter from the path of the charged particle, before the
charged particle strikes the target or window, shall not exceed 0.1 percent for
x-ray leakage nor 0.05 percent for neutron leakage of the maximum absorbed dose
in rads (grays) of the unattenuated useful beam measured at the point of
intersection of the central axis of the beam and the circular plane specified
in .04(18)(b)1. (i) (I).
(ii) The
registrant shall determine or obtain from the manufacturer, the actual leakage
radiation existing at the positions specified in .04(18)(b)2. (i) for specified
operating conditions. Radiation measurements excluding neutrons shall be
averaged over an area up to but not exceeding 100 square centimeters. Neutron
measurements shall be averaged over an area up to but not exceeding 200 square
centimeters.
3. The
registrant shall assure that adjustable or interchangeable beam limiting
devices are provided and that such devices shall transmit no more than 2
percent of the useful beam at the normal treatment distance for the portion of
the useful beam which is to be attenuated by the beam limiting device.
Documentation of the transmission factors shall be maintained at the facility
for inspection by the Department. The neutron component of the useful beam
shall not be included in this requirement.
4. Filters.
(i) Each filter which is removable from the
system shall be clearly marked with an identification number. Documentation
available at the control panel shall contain a description of the filter. For
wedge filters, the wedge angle shall appear on the wedge or wedge
tray.
(ii) For equipment
manufactured after the effective date of these regulations which utilizes a
system of wedge filters, interchangeable field flattening filters, or
interchangeable beam scattering filters:
(I)
irradiation shall not be possible until a selection of a filter has been made
at the treatment control panel;
(II) an interlock system shall be provided to
prevent irradiation if the filter selected is not in the correct
position;
(III) a display shall be
provided at the treatment control panel indicating the filter(s) in
use;
(IV) an interlock shall be
provided to prevent irradiation if any filter selection operation carried out
in the treatment room does not agree with the filter selection operation
carried out at the treatment control panel.
5. Beam Symmetry. In equipment manufactured
after the effective date of these regulations, inherently capable of producing
useful beams with asymmetry exceeding 5 percent, the asymmetry of the radiation
beam in two orthogonal directions shall be monitored before the beam passes
through the beam limiting device. Facilities shall be provided so that, if the
difference in dose rate between one region and another region symmetrically
displaced from the central axis of the beam exceeds 5 percent of the central
axis dose rate, indication of this condition is made at the control panel; and
if this difference exceeds 10 percent, the irradiation is terminated. It shall
be the registrant's responsibility to assure that the above requirements are
met and that records of confirming tests are maintained for Departmental
inspection.
6. Beam Monitors. All
therapy accelerator machines shall be provided with radiation detectors in the
radiation head.
(i) New equipment shall be
provided with at least two radiation detectors. The detectors shall be
incorporated into two separate dose monitoring systems.
(ii) Existing equipment shall be provided
with at least one radiation detector. This detector shall be incorporated into
a primary dose monitoring system.
7. Selection and Display of Dose Monitor
Units.
(i) Irradiation shall not be possible
until a selection of a number of dose monitor units has been made at the
treatment control panel.
(ii) The
preselected number of dose monitor units shall be displayed at the treatment
control panel until reset manually for the next irradiation.
(iii) After termination of irradiation, it
shall be necessary to zero before subsequent treatment can be
initiated.
(iv) For equipment
manufactured after the effective date of these regulations, it shall be
necessary after termination of irradiation to manually reset the preselected
dose monitor units before irradiation can be initiated.
8. Interruption Switches. It shall be
possible to interrupt irradiation and equipment movements at any time from the
operator's position at the treatment control panel. Following an interruption,
it shall be possible to restart irradiation by operator action without any
reselection of operating conditions. If any change is made of a preselected
value during an interruption, irradiation and equipment movements shall be
automatically terminated.
9.
Termination of Irradiation by the Dose Monitoring System or Systems.
(i) Each of the required monitoring systems
shall be capable of independently terminating irradiation.
(ii) Each primary system shall terminate
irradiation when the preselected number of dose monitor units has been detected
by the system.
(iii) If original
design of the equipment included a second dose monitoring system, that system
shall be capable of terminating irradiation when not more than 15 percent or 40
dose monitoring units above the preselected number of dose monitor units set at
the control panel has been detected by the second dose monitoring
system.
10. Termination
Switches. It shall be possible to terminate irradiation and equipment movements
at any time from the operator's position at the treatment control
panel.
11. Timer.
(i) A timer shall be provided which has a
display at the treatment control panel. The timer shall have a preset time
selector and an elapsed time indicator.
(ii) The timer shall be a cumulative timer
which activates with the production of radiation and retains its reading after
irradiation is interrupted or terminated. After irradiation is terminated and
before irradiation can be reinitiated, it shall be necessary to reset the
elapsed time indicator to zero.
(iii) For equipment manufactured after the
effective date of these regulations after termination of irradiation and before
irradiation can be reinitiated, it shall be necessary to manually reset the
preset time selector.
(iv) The
timer shall terminate irradiation when a preselected time has elapsed if the
dose monitoring systems fail to do so.
12. Selection of Radiation Type. Equipment
capable of both x-ray therapy and electron therapy shall meet the following
additional requirements:
(i) Irradiation
shall not be possible until a selection of radiation type has been made at the
treatment control panel.
(ii) An
interlock system shall be provided to ensure that the equipment can emit only
the radiation type which has been selected.
(iii) An interlock system shall be provided
to prevent irradiation if any selected operations carried out in the treatment
room do not agree with the selected operations carried out at the treatment
control panel.
(iv) An interlock
system shall be provided to prevent irradiation with x-rays except to obtain a
port film when electron applicators are fitted.
(v) An interlock system shall be provided to
prevent irradiation with electrons when accessories specific for x-ray therapy
are fitted.
(vi) The radiation type
selected shall be displayed at the treatment control panel before and during
irradiation.
13.
Selection of Energy. Equipment capable of generating radiation beams of
different energies shall meet the following requirements:
(i) Irradiation shall not be possible until a
selection of energy has been made at the treatment control panel.
(ii) An interlock system shall be provided to
prevent irradiation if any selected operations to be carried out in the
treatment room do not agree with those selected operations carried out at the
treatment control panel.
(iii) The
nominal energy value selected shall be displayed at the treatment control panel
before and during irradiation.
(iv)
For new equipment, an interlock system shall be provided to terminate
irradiation if the energy of the electrons striking the x-ray target or
electron window deviates by more than 20 percent or 3 MeV, whichever is
smaller, from the selected nominal energy.
14. Selection of Stationery Beam Therapy or
Moving Beam Therapy. Equipment capable of both stationary beam therapy and
moving beam therapy shall meet the following requirements:
(i) Irradiation shall not be possible until a
selection of stationery beam therapy or moving beam therapy has been made at
the treatment control panel.
(ii)
An interlock system shall be provided to insure that the equipment can operate
only in the mode which has been selected.
(iii) An interlock system shall be provided
to prevent irradiation if any selected operations to be carried out in the
treatment room do not agree with those selected operations carried out at the
treatment control panel.
(iv) The
mode of operation shall be displayed at the treatment control panel.
(v) For new equipment, an interlock system
shall be provided to terminate irradiation if:
(I) movement of the gantry occurs during
moving stationary beam therapy; or
(II) movement of the gantry stops during
moving beam therapy unless such stoppage is a preplanned function.
(vi) Moving beam therapy shall be
controlled to obtain the selected relationships between incremental dose
monitor units and incremental angle of movement.
(I) For new equipment, an interlock system
shall be provided to terminate irradiation if the number of dose monitor units
delivered in any 10 degrees of arc differs by more than 20 percent from the
selected value.
(II) for new
equipment, where gantry angle terminates the irradiation in arc therapy, the
dose monitor units shall differ by less than 5 percent from the value
calculated from the absorbed dose per unit angle relationship.
(c) Facility
and Shielding Requirements. In addition to Section
.01(8) of these
regulations, the following design requirements shall apply:
1. The treatment control panel shall be
located outside the treatment room; and
2. Except for entrance doors or beam
interceptors, all the required barriers shall be fixed; and
3. Windows, mirrors, closed-circuit
television, or other equivalent viewing devices shall be provided to permit
continuous observation of the patient during irradiation and shall be so
located that the operator may observe the patient from the treatment control
panel. When the viewing system is by electronic methods, a secondary viewing
system, which may also be electronic, shall be available for use in the event
of failure of the primary system; and
4. Provision shall be made for two-way aural
communication between the patient and the operator at the treatment control
panel; and
5. The entrance to the
treatment room shall be equipped with a steady, red warning light which
operates when, and only when, radiation is being produced; and
6. Interlocks shall be provided such that all
entrance doors shall be closed before treatment can be initiated or continued.
If the radiation beam is interrupted by any door opening, it shall be possible
to restore the machine to operation only by closing the door and reinitiating
exposure by manual action at the control panel.
(d) Calibrations and Spot Checks.
1. Calibration.
(i) A calibration of all new machines and
existing machines not previously surveyed shall be performed prior to the
initial irradiation of a patient and thereafter at time intervals not to exceed
12 months, and after any change which might significantly alter the
calibration, spatial distribution, or other characteristics of the therapy
beam. It shall be the responsibility of the registrant to ensure that the
individual performing the calibration is competent to perform such
calibrations.
(ii) The calibration
of a particle accelerator machine shall be performed in accordance with a
calibration protocol such as that published by the American Association of
Physicists in Medicine in Volume 10, number 6, issue of Medical Physics, or its
current revision or replacement.
(iii) Any calibration protocol used must
contain the following minimum measurement criteria:
(I) full calibration measurements shall be
performed using a dosimetry system that has been calibrated by the National
Bureau of Standards or by a Regional Calibration Laboratory accredited by the
American Association of Physicists in Medicine. The dosimetry system shall have
been calibrated within the previous two years and after any servicing that may
have affected system calibration.
(II) spot-check measurements shall be
performed using a dosimetry system that has been calibrated in accordance with
.04(18)(d)1. (iii)(I) of this rule. Alternatively, a dosimetry system
spot-check measurements may be calibrated by direct intercomparison with a
system that has been calibrated in accordance with .04(18)(d)1. (iii)(I) of
this rule. This alternative calibration method shall have been performed within
the previous one year and after each servicing that may have affected system
calibration. Dosimetry systems calibrated by this alternative method shall not
be used for full calibration measurements.
(iv) The full calibration of the therapy beam
shall include but not be limited to the following determinations:
(I) verification that the equipment is
operating in compliance with the design specifications concerning the light
localizer, the side light and back-pointer alignment with the isocenter when
applicable, variation in the axis of rotation for the table, gantry and jaw
system, and beam flatness and symmetry at the specified depth.
(II) the absorbed dose rate at various depths
of water for the range of field sized used, for each effective energy, and for
each treatment distance used for radiation therapy.
(III) the uniformity of the radiation field
and any dependency upon the direction of the useful beam.
(IV) verification of depth-dose data and
isodose curves applicable to the specific machine.
(V) verification of transmission factors for
all accessories such as wedges shadow trays, etc.
(VI) records of full calibration measurements
and dosimetry system calibrations shall be preserved for 5 years after
completion of the full calibration.
(VII) a copy of the latest full calibration
performed as described in .04(18)(d)1. (iv)(I)-(VI) shall be available at the
accelerator facility.
2. Spot-Calibration Checks.
(i) Spot-calibration checks shall be
performed on machines subject to .04(18)(b) during calibrations and thereafter
at intervals not to exceed one month.
(ii) Such spot-calibration checks shall be in
accordance with written procedures and shall include absorbed dose measurements
in a phantom at intervals not to exceed one week.
(iii) Where a system has built-in devices
which provide a measurement of any parameter during irradiation, such
measurement shall not be utilized as a spot-check; and
(iv) Records of spot-check measurements
performed pursuant to .04(18)(b) shall be maintained by the registrant for a
period of 2 years after completion of the spot-check measurements and any
necessary corrective actions.
(e) Qualified Expert. The registrant shall
determine if a person is an expert qualified by training and experience to
calibrate a therapy machine and establish procedures for (and review the
results of) spot-check measurements. The registrant shall determine that the
person calibrating their therapy machine:
1.
is certified by the American Board of Radiology in Therapeutic Radiological
Physics, Radiological Physics, Roentgen-Ray and Gamma-Ray Physics, or x-ray and
Radium Physics; or
2. has the
following minimum training and experience:
(i)
a Master's or Doctor's degree in physics, biophysics, radiological physics or
health physics;
(ii) one year of
full-time training in therapeutic radiological physics; and
(iii) one year of full-time experience in a
radiotherapy facility including personal calibration and spot-check of at least
one therapy machine.
(f) Operating Procedures.
1. No individual other than the patient shall
be in the treatment room during treatment of a patient.
2. If a patient must be held in position
during treatment, mechanical supporting or restraining devices shall be
used.
3. The machine shall not be
used in the administration of radiation therapy unless the requirements of
.04(18)(d) have been met.