In addition to the requirements of sections
33.1-10-06-03 and
33.1-10-06-04, the requirements
of this section apply to x-ray equipment and associated facilities used for
dental radiography. Criteria for extraoral dental radiographic systems are
covered in section
33.1-10-06-06. Only systems
meeting the requirements of this section shall be used.
1.
Source-to-skin distance.
X-ray systems designed for use with an intraoral image receptor shall be
provided with means to limit source-to-skin distance to not less than:
a. Eighteen centimeters if operable above
fifty kilovolts peak.
b. Ten
centimeters if operable at fifty kilovolts peak only.
2.
Beam limitation. Radiographic
systems designed for use with an intraoral image receptor shall be provided
with means to limit the x-ray beam such that:
a. The x-ray beam, at the minimum
source-to-skin distance, shall be containable in a circle having a diameter of
no more than seven centimeters.
b.
An open-ended shielded position indicating device shall be used. The shielding
shall be equivalent to the requirements of subsection 4 of section
33.1-10-06-04.
3.
Radiation exposure control.
a. Exposure initiation.
(1) Means shall be provided to initiate the
radiation exposure by a deliberate action on the part of the operator, such as
the depression of a switch. Radiation exposure shall not be initiated without
such an action.
(2) It shall not be
possible to make an exposure when the timer is set to a "zero" or "off"
position if either position is provided.
b. Exposure indication. Means shall be
provided for visual indication observable at or from the operator's protected
position whenever x-rays are produced. In addition, a signal audible to the
operator shall indicate that the exposure has terminated.
c. Exposure termination.
(1) Means shall be provided to terminate the
exposure at a preset time interval, preset product of current and time, a
preset number of pulses, or a preset radiation exposure to the image
receptor.
(2) An x-ray exposure
control shall be incorporated into each x-ray system such that an exposure can
be terminated by the operator at any time, except for exposures of one-half
second or less.
(3) Termination of
an exposure shall cause automatic resetting of the timer to its initial setting
or to "zero".
d. Exposure
duration (timer) linearity. For systems having independent selection of
exposure time settings, the average ratios (X
1) of
exposure to the indicated timer setting, in units of coulombs per kilogram per
second [milliroentgen per second], obtained at any two clinically used timer
settings shall not differ by more than ten hundredths times their sum.
Click
here to view image
where X1 and X2
are the average values.
e.
Exposure control location and operator protection.
(1) Stationary x-ray systems shall be
required to have the x-ray exposure control permanently mounted in a protected
area, so that the operator is required to remain in that protected area during
the entire exposure and so the operator can view the patient while making the
exposure.
(2) Mobile and portable
x-ray systems which are:
(a) Used for greater
than one week in the same location, i.e., a room or suite, shall meet the
requirements of paragraph 1.
(b)
Used for less than one week in the same location shall be provided with either
a protective barrier at least two meters [6.5 feet] high for operator
protection, or means to allow the operator to be at least two and seven-tenths
meters [9 feet] from the tube housing assembly while making
exposures.
(3) Hand-held
dental x-ray equipment:
(a) Operators shall
use all safety devices and follow safety procedures according to the
manufacturer.
(b) Operators shall
wear a protective apron during exposure in accordance with subparagraph b of
paragraph 6 of subdivision a of subsection 1.
4.
Reproducibility. When the equipment is operated on an adequate
power supply as specified by the manufacturer, the estimated coefficient of
variation of radiation exposures shall be no greater than five hundredths for
any specific combination of selected technique factors.
5.
mA/mAs linearity. The
following requirements apply when the equipment is operated on a power supply
as specified by the manufacturer for any fixed x-ray tube potential within the
range of forty percent to one hundred percent of the maximum rated.
a. Equipment having independent selection of
x-ray tube current (mA). The average ratios (X
1) of
exposure to the indicated milliampere seconds product, in units of coulombs per
kilogram per milliampere second (or milliroentgen per milliampere seconds),
obtained at any two consecutive tube current settings shall not differ by more
than ten hundredths times their sum:
Click
here to view image
where X1 and X2
are the average values obtained at any two consecutive tube current settings,
or at two settings differing by no more than a factor of two where the tube
current selection is continuous.
b. Equipment having a combined x-ray tube
current exposure time product (mAs) selector, but not a separate tube current
(mA) selector. The average ratios (X
1) of exposure to
the indicated milliampere-seconds product, in units of coulombs per kilogram
per milliampere second (or mR/mAs), obtained at any two consecutive mAs
selector settings shall not differ by more than ten hundredths times their sum:
Click
here to view image
where X1 and X2
are the average values obtained by any two mAs selector settings, or at two
settings differing by no more than a factor of two where the mAs selector
provides continuous selection.
c. Measuring compliance. Determination of
compliance shall be based on ten exposures taken within a time period of one
hour, at each of the two settings. These two settings may include any two focal
spot sizes except where one is equal to or less than forty-five hundredths
millimeters and the other is greater than forty-five hundredths millimeters.
For purposes of this requirement, focal spot size is the nominal focal spot
size specified by the x-ray tube manufacturer.
6.
Accuracy. Deviation of
technique factors from values for kVp and exposure time (if time is
independently selectable) shall not exceed the limits specified for that system
by its manufacturer. In the absence of manufacturer's specifications, the
deviation shall not exceed ten percent of the indicated value for kVp and
twenty percent for time.
7.
kVp limitations. Dental x-ray machines with a nominal fixed kVp of
less than fifty kVp shall not be used to make diagnostic dental radiographs of
humans.
8.
Beam quality.
All dental x-ray systems are subject to the filtration requirements of
subdivision a of subsection 5 of section
33.1-10-06-04
9.
Administrative controls.
a. Patient and film holding devices shall be
used when the techniques permit.
b.
The tube housing and the position indicating device shall not be handheld
during an exposure unless the system was specifically designed as a hand-held
dental x-ray machine.
c. The x-ray
system shall be operated in such a manner that the useful beam at the patient's
skin does not exceed the requirements of subdivision a of subsection
2.
d. Dental fluoroscopy without
image intensification shall not be used.
e. Security for portable dental systems. A
means shall be provided to prevent unauthorized use whenever the x-ray system
is not under the control and constant surveillance of the registrant or an
authorized operator.
10.
Structural shielding requirements (see appendix C).