Ohio Admin. Code 3701:1-48-19 - Personnel monitoring
(A) The licensee may not permit any
individual to act as a radiographer or a radiographer's assistant unless, at
all times during radiographic operations, each individual wears, on the trunk
of the body, a direct reading dosimeter, an operating alarm rate meter, and a
personnel dosimeter.
that is processed and evaluated by an accredited national
voluntary laboratory accreditation program (NVLAP) processor. At
permanent radiography installations where other appropriate alarming or warning
devices are in routine use, the wearing of an alarming rate meter is not
required.
(1) Pocket dosimeters must have a
range from zero to two millisievert (two hundred millirem) and must be
recharged at the start of each shift. Electronic personal dosimeters may only
be used in place of ion-chamber pocket dosimeters.
(2) Each personnel dosimeter must be assigned
to and worn by only one individual.
(3) Film badges must be replaced at
periods not to exceed one month
least monthly and all
other personnel dosimeters processed and evaluated
by an accredited NVLAP processor must be replaced at periods not to exceed
three months.
that require replacement must be
replaced at least quarterly. All personnel dosimeters must be evaluated at
least quarterly or promptly after replacement, whichever is more
frequent.
(4) After replacement, each
personnel dosimeter must be processed as soon as possible.
(B) Direct reading
dosimeters such as pocket dosimeters or electronic personal dosimeters, must be
read and the exposures recorded at the beginning and end of each shift, and
records must be maintained in accordance with paragraph (K) of rule
3701:1-48-23 of the
Administrative Code.
(C) Pocket
dosimeters, or electronic personal dosimeters, must be checked at periods not
to exceed twelve months for correct response to radiation, and records must be
maintained in accordance with paragraph (K) of rule
3701:1-48-23 of the
Administrative Code. Acceptable dosimeters must read within plus or minus
twenty per cent of the true radiation exposure.
(D) If an individual's pocket dosimeter is
found to be off-scale, or if his or her electronic personal dosimeter reads
greater than two millisievert (two hundred millirem), and the possibility of
radiation exposure cannot be ruled out as the cause, the individual's personnel
dosimeter that requires processing must be sent
for processing and evaluation within twenty-four
hours. For personnel dosimeters that do not require
processing, evaluation of the dosimeter must be started within twenty-four
hours. In addition, the individual may not resume work associated with
licensed material use until a determination of the individual's radiation
exposure has been made. This determination must be made by the radiation safety
officer for industrial radiography or the
radiation safety officer's
for industrial radiography's designated qualified
individual
designee. The results of this
determination must be included in the records maintained in accordance with
paragraph (K) of rule
3701:1-48-23 of the
Administrative Code.
(E) If the
personnel dosimeter that is required by paragraph (A) of this rule is lost or
damaged, the worker shall cease work immediately until a replacement personnel
dosimeter meeting the requirements in paragraph (A) of this rule is provided
and the exposure is calculated for the time period from issuance to loss or
damage of the personnel dosimeter. The results of the calculated exposure and
the time period for which the personnel dosimeter was lost or damaged must be
included in the records maintained in accordance with paragraph (K) of rule
3701:1-48-23 of the
Administrative Code.
(F) Dosimetry
reports received from the accredited NVLAP personnel
dosimeter processor must be retained in accordance with paragraph (K)
of rule
3701:1-48-23 of the
Administrative Code.
(G) Each alarm
rate meter must:
(1) Be checked to ensure
that the alarm functions properly (sounds) before using at the start of each
shift;
(2) Be set to give an alarm
signal at a preset dose rate of five millisievert (five hundred millirem) per
hour; with an accuracy of plus or minus twenty per cent of the true radiation
dose rate;
(3) Require special
means to change the preset alarm function; and
(4) Be calibrated at periods not to exceed
twelve months for correct response to radiation. The licensee shall maintain
records of alarm rate meter calibrations in accordance with paragraph (K) of
3701:1-48-23 of the Administrative Code.
Notes
Promulgated Under: 119.03
Statutory Authority: 3748.04
Rule Amplifies: 3748.06, 3748.07
Prior Effective Dates: 10/20/2002, 03/22/2007, 01/01/2012, 06/27/2016
State regulations are updated quarterly; we currently have two versions available. Below is a comparison between our most recent version and the prior quarterly release. More comparison features will be added as we have more versions to compare.
No prior version found.