Ohio Admin. Code 3701-84-32 - General facilities, equipment, and supplies - adult cardiac catheterization service
(A)
Each provider of an adult cardiac
catheterization service, or "service" shall provide adequate, properly designed
space to perform cardiac catheterization procedures safely and effectively. The
amount of space may vary with the types of procedures performed and the nature
of the facility (e.g., interventional versus diagnostic, teaching versus
non-teaching).
(B) Each service
shall utilize the following:
(1) Procedure
rooms will:
that meet
or exceed minimum space requirements established by the equipment manufacturer
and are at least four hundred square feet or, if constructed or renovated or
cardiac catheterization services initiated after March 20, 1997, at least five
hundred square feet. Each procedure room shall have a designated control
room;
(a)
Have a minimum floor area of four hundred square
feet;
(b)
Have the following minimum clearances:
(i)
If anesthesia is
utilized, six feet at the head of the operating table. This dimension will
result in an anesthesia work zone with a clear floor area of six feet by eight
feet; and
(ii)
Sufficient footage on each side of the procedure table
to ensure that movement around and in the sterile field does not compromise or
contaminate the sterile field.
(c)
Procedure rooms
utilizing image-guided surgery or procedures requiring more space for personnel
or equipment will be sized to accommodate the personnel and equipment planned
to be in the room during procedures and have a minimum floor area of six
hundred square feet with a minimum clear dimension of twenty
feet.
(d)
Hybrid procedure rooms will meet the minimum floor area
and clearances of paragraphs (B)(1)(a) and (B)(1)(b) of this rule and include
at least the additional minimum floor area, clearances, and storage
requirements for the imaging equipment contained in the room.
(e)
Fixed
encroachments into the minimum clear floor area are permitted to be included
when determining the minimum clear floor area for a procedure room as long
as:
(i)
There
are no encroachments into the sterile field;
(ii)
The
encroachments do not extend more than twelve inches into the minimum clear
floor area outside the sterile field; and
(iii)
The
encroachment width along each wall does not exceed ten percent of the length of
that wall.
(2)
Control
Control
rooms, if utilized, will
of at least ninety-six square feet for each
procedure room served. Control rooms that are constructed or renovated or
cardiac catheterization services initiated after March 20, 1997 shall have at
least one hundred and fifty square feet for each procedure room served;
and
:
(a)
Accommodate the
imaging system control equipment;
(b)
Be sized and
configured in compliance with manufacturer recommendations for installation,
service, and maintenance;
(c)
Have view panels that provide for a view of the patient
and the cardiac catheterization team;
(d)
Be permitted to
serve more than one hybrid procedure room, provided that manufacturer
recommendations for installation, service, and maintenance are accommodated for
all rooms served; and
(e)
Be physically separated from a hybrid procedure room
with walls and a door.
(f)
A door is not required where a control room serves only
one procedure room and is built, maintained, and controlled the same as the
procedure room.
(3)
Clean utility rooms that can sufficiently and efficiently supply the procedure
rooms.
(C) A general
radiology room cannot be used as a cardiac catheterization procedure
room.
(D) Equipment described in
this paragraph may be replaced by newer technology that has equivalent or
superior capability as determined by the provider of the service. In assessing
these new technologies, consideration should be given to recommendations of
recognized professional societies and accrediting bodies (e.g. the American
college of cardiology). All services shall
will, at a minimum have the following equipment:
(1) High quality x-ray imaging with mutiaxial
position capability;
(2) A
three-phase, twelve pulse generator with an output of eighty to one hundred
kilowatts or a constant potential generator with an output of at least one
hundred kilowatts at one hundred kilovolts;
(3) Multimode or cesium iodide image
intensifiers;
(4) High-quality
matched optics;
(5) Appropriate
imaging modality for review and storage of images;
(6) Digital imaging equipment;
(7) Carbon-fiber table tops;
(8) A mechanism for continuous monitoring of
a patient's blood pressure and electrocardiogram;
(9) An adequate supply and variety of
catheters, guide wires, and sheaths;
(10) A crash cart with equipment for
ventilatory support;
(11) A
defibrillator;
(12) A temporary
pacemaker; and
(13) An intra-aortic
balloon pump or other percutaneous mechanical circulatory assistance device of
superior capability.
(E)
The provider of a
Each adult cardiac catheterization service
shall
will
provide and maintain radiation generating equipment in accordance with
applicable state and federal requirements as specified in Chapter 3748. of the
Revised Code and the rules adopted thereunder.
Notes
Promulgated Under: 119.03
Statutory Authority: 3702.11, 3702.13
Rule Amplifies: 3702.14, 3702.141, 3702.15, 3702.16, 3702.18
Prior Effective Dates: 03/20/1997, 03/24/2003, 05/15/2008, 04/30/2016, 08/01/2017
State regulations are updated quarterly; we currently have two versions available. Below is a comparison between our most recent version and the prior quarterly release. More comparison features will be added as we have more versions to compare.
No prior version found.