Ohio Admin. Code 3701-84-77 - Facilities, equipment, and supplies - pediatric cardiac catheterization service
(A)
Each pediatric cardiac catheterization service will
provide adequate, properly designed space to perform cardiac catheterization
procedures safely and effectively. The amount of space may vary with the types
of procedures performed and the nature of the facility such as interventional
versus diagnostic and teaching versus non-teaching.
(B)
Each pediatric
cardiac catheterization service will ensure the following:
(1)
Procedure
rooms:
(a)
Have
a minimum floor area of four hundred square feet;
(b)
Have the
following minimum clearances:
(i)
If anesthesia is utilized, six feet at the head of the
operating table. This dimension will result in an anesthesia work zone with a
clear floor area of six feet by eight feet.
(ii)
Sufficient
footage on each side of the procedure table to ensure that movement around and
in the sterile field does not compromise or contaminate the sterile
field.
(c)
Utilizing image-guided surgery or procedures requiring
more space for personnel or equipment are sized to accommodate the personnel
and equipment planned to be in the room during procedures and have a minimum
floor area of six hundred square feet with a minimum clear dimension of twenty
feet.
(d)
Classified as hybrid procedure rooms, will meet the
minimum clear floor area and clearances of paragraphs (B)(1)(a) and (B)(1)(b)
of this rule and include at least the additional minimum clear floor area,
clearances, and storage requirements for the imaging equipment contained in the
room.
(e)
Containing fixed encroachments into the minimum clear
floor area are permitted to be include those fixed encroachments when
determining the minimum clear floor area for a procedure room as long as:
(i)
There are no
encroachments into the sterile field;
(ii)
The
encroachments do not extend more than twelve inches into the minimum clear
floor area outside the sterile field; and
(iii)
The
encroachment width along each wall does not exceed ten per cent of the length
of that wall.
(2)
Control rooms, if
utilized, will:
(a)
Accommodate the imaging system control
equipment;
(b)
Be sized and configured in compliance with manufacturer
recommendations for installation, service, and maintenance;
(c)
Have view panels
that provide for a view of the patient and the cardiac catheterization
team;
(d)
Be permitted to serve more than one hybrid procedure
room, provided that manufacturer recommendations for installation, service, and
maintenance are accommodated for all rooms served.
(e)
Be physically
separated from a hybrid procedure room with walls and a door. A door is not
required where a control room serves only one procedure room and is built,
maintained, and controlled the same as the procedure
room.
(3)
Clean utility rooms can sufficiently and efficiently
supply the procedure rooms.
(C)
A general
radiology room cannot be used as a cardiac catheterization procedure
room.
(D)
Each pediatric cardiac catheterization service will
ensure that each laboratory where a cardiac catheterization procedure is
conducted is appropriately equipped to provide high quality pediatric imaging,
physiological monitoring, and provision of emergency care.
(E)
The equipment
described in this paragraph may be replaced by newer technology that has
equivalent or superior capability as determined by the pediatric cardiac
catheterization service. In assessing these new technologies, consideration
should be given to recommendations of recognized professional societies and
accrediting bodies (e.g. the American college of cardiology; the American
academy of pediatrics). All pediatric cardiac catheterization services will
have the following equipment:
(1)
Imaging equipment appropriate to the procedures
performed;
(2)
Immediate replay capabilities;
(3)
A blood gas
analyzer;
(4)
A pulse oximeter;
(5)
An infant warming
device;
(6)
Pacing catheters;
(7)
An external
pacemaker;
(8)
A defibrillator;
(9)
An emergency
cart;
(10)
A comprehensive inventory of pediatric catheters and
devices; and
(11)
Ultrasound.
(F)
Each pediatric
cardiac catheterization service will provide and maintain radiation generating
equipment in accordance with applicable state and federal requirements as
specified in Chapter 3748. of the Revised Code and the rules adopted
thereunder.
Replaces: 3701-84-77
Notes
Promulgated Under: 119.03
Statutory Authority: 3702.11, 3702.13
Rule Amplifies: 3702.11, 3702.12, 3702.13, 3702.14, 3702.141, 3702.15, 3702.16, 3702.18, 3702.19, 3702.20
Prior Effective Dates: 03/01/1997, 03/24/2003, 06/01/2007, 06/21/2012, 08/01/2017
State regulations are updated quarterly; we currently have two versions available. Below is a comparison between our most recent version and the prior quarterly release. More comparison features will be added as we have more versions to compare.
No prior version found.