Tenn. Comp. R. & Regs. 0400-20-07-.71 - FULL CALIBRATION MEASUREMENTS ON GAMMA STEREOTACTIC RADIOSURGERY UNITS
(1) A licensee
authorized to use a gamma stereotactic radiosurgery unit for medical use shall
perform full calibration measurements on each unit:
(a) Before the first medical use of the
unit;
(b) Before medical use under
the following conditions:
1. Whenever
spot-check measurements indicate that the output differs by more than 5 percent
from the output obtained at the last full calibration corrected mathematically
for radioactive decay;
2. Following
replacement of the sources or following reinstallation of the gamma
stereotactic radiosurgery unit in a new location; and
3. Following any repair of the gamma
stereotactic radiosurgery unit that includes removal of the sources or major
repair of the components associated with the source assembly; and
(c) At intervals not exceeding 1
year, with the exception that relative helmet factors need only be determined
before the first medical use of a helmet and following any damage to a
helmet.
(2) To satisfy
the requirements of paragraph (1) of this rule, full calibration measurements
must include determination of: .
(a) The
output within ±3 percent;
(b) Relative helmet factors;
(c) Isocenter coincidence;
(d) Timer accuracy and linearity over the
range of use;
(e) On-off
error;
(f) Trunnion
centricity;
(g) Treatment table
retraction mechanism, using backup battery power or hydraulic backups with the
unit off;
(h) Helmet
microswitches;
(i) Emergency timing
circuits; and
(j) Stereotactic
frames and localizing devices (trunnions).
(3) A licensee shall use the dosimetry system
described in paragraph (1) of Rule 0400-20-07-.68 to measure the output for one
set of exposure conditions. The remaining radiation measurements required in
subparagraph (2)(a) of this rule may be made using a dosimetry system that
indicates relative dose rates.
(4)
A licensee shall make full calibration measurements required by paragraph (1)
of this rule in accordance with published protocols accepted by nationally
recognized bodies.
(5) A licensee
shall mathematically correct the outputs determined in subparagraph (2)(a) of
this rule at intervals not exceeding 1 month for cobalt-60 and at intervals
consistent with 1 percent physical decay for all other radionuclides.
(6) Full calibration measurements required by
paragraph (1) of this rule and physical decay corrections required by paragraph
(5) of this rule must be performed by the authorized medical
physicist.
(7) A licensee shall
retain a record of each calibration in accordance with Rule
0400-20-07-.104.
Notes
Authority: T.C.A. §§ 68-202-101 et seq., 68-202-201 et seq. and 4-5-201 et seq.
State regulations are updated quarterly; we currently have two versions available. Below is a comparison between our most recent version and the prior quarterly release. More comparison features will be added as we have more versions to compare.
No prior version found.