Utah Admin. Code R432-650-12 - Water Quality
(1) Water used for
dialysis purposes shall comply with quality standards established by the
Association for the Advancement of Medical Instrumentation (AAMI) as published
in "Hemodialysis Systems," second edition, which is adopted and incorporated by
reference.
(2) Each ESRD facility
that utilizes in-center water systems must have bacteriologic quality analysis
performed and documented at least monthly by a laboratory that adheres to AAMI
standards.
(3) For home systems,
the ESRD facility must conduct bacteriological quality analysis at least
monthly using an approved home testing methodology as identified in the patient
care plan.
(a) An alternate schedule of
testing may be approved by the attending physician.
(b) The alternate schedule shall be specified
in the patient care plan.
(4) If reverse osmosis or deionization
devices are used for in-center or home systems, the ESRD facility must have
chemical quality analysis performed and documented at least once every 12
months by a laboratory that adheres to AAMI standards.
(5) The ESRD facility must maintain and make
available for Department review all water quality test results. In the case of
home dialysis, test results shall become part of the patient record maintained
by the ESRD facility.
Notes
State regulations are updated quarterly; we currently have two versions available. Below is a comparison between our most recent version and the prior quarterly release. More comparison features will be added as we have more versions to compare.
No prior version found.