Ariz. Admin. Code § R9-18-408 - Method Criteria and References for Laboratory Analyses
A. In addition to the definitions in A.R.S.
§
36-2850 and
R9-18-101, the definitions in
A.A.C. R9-17-404.03(A)
apply in this Section unless otherwise stated.
B. A technical laboratory director shall
ensure that the marijuana testing facility complies with the requirements in
A.A.C. R9-17-404.03(B) through
(O) when using chemical analytical methods
for any of the analytes in Table 3.1.
C. A technical laboratory director may
release testing results that are scientifically valid and defensible from
analyses using chemical analytical methods, according to
R9-18-410(B)(3) and
(C), with the following data qualifier
notations if:
1. The target analyte detected
in the calibration blank required in A.A.C.
R9-17-404.03(F)(1)(c)
or the method blank specified in A.A.C.
R9-17-404.03(K)(1)
is at or above the limit of quantitation, but the sample result:
a. For potency testing, is below the limit of
quantitation - B1; or
b. When
testing for pesticides, fungicides, growth regulators, mycotoxins, heavy
metals, or residual solvents, is below the maximum allowable concentration in
Table 3.1 for the analyte - B2;
2. The limit of quantitation and the sample
results were adjusted to reflect sample dilution - D1;
3. The relative intensity of a characteristic
ion in a sample analyte exceeded the acceptance criteria in A.A.C.
R9-17-404.03(L)(1) with respect to the reference spectra, indicating
interference - I1;
4. When testing
for pesticides, fungicides, growth regulators, mycotoxins, heavy metals, or
residual solvents, the percent recovery of a laboratory control sample is
greater than the acceptance limits in A.A.C.
R9-17-404.03(K)(2)(d), but the sample's target analytes were not detected above the maximum allowable
concentrations in Table 3.1 for the analytes in the sample - L1;
5. The recovery from the matrix spike in
A.A.C. R9-17-404.03(K)(4)
was:
a. High, but the recovery from the
laboratory control sample in A.A.C.
R9-17-404.03(K)(2) was within acceptance criteria -
M1,
b. Low, but the recovery from
the laboratory control sample in A.A.C.
R9-17-404.03(K)(2)
was within acceptance criteria - M2, or
c. Unusable because the analyte concentration
was disproportionate to the spike level, but the recovery from the laboratory
control sample in A.A.C.
R9-17-404.03(K)(2)
was within acceptance criteria - M3;
6. The analysis of a spiked sample required a
dilution such that the spike recovery calculation does not provide useful
information, but the recovery from the associated laboratory control sample in
A.A.C. R9-17-404.03(K)(2)
was within acceptance criteria - M4;
7. The analyte concentration was determined
by the method of standard addition, in which the standard is added directly to
the aliquots of the analyzed sample - M5;
8. A description of the variance is described
in the final report of testing according to
R9-18-410(B)(3) and
(C) - N1;
9. The relative percent difference for the
laboratory control sample and duplicate exceeded the limit in A.A.C.
R9-17-404.03(K)(3), but the recovery in A.A.C.
R9-17-404.03(K)(2)(d) was within acceptance criteria -
R1;
10. The relative percent
difference for a sample and duplicate exceeded the limit in A.A.C.
R9-17-404.03(O) -
R2; or
11. The recovery from
continuing initial calibration verification standards or continuing calibration
verification standards is greater than the acceptance limits in A.A.C.
R9-17-404.03(H)(2) or (J)(1)(b) as applicable, but the sample's target analytes
were not detected above the maximum allowable concentrations in Table 3.1 for
the analytes in the sample - V1.
D. A technical laboratory director shall
include in the final report of testing from analyses using chemical analytical
methods, according to
R9-18-410(B)(3) and
(C), the following data qualifier notations
if:
1. Sample integrity was not maintained -
Q1;
2. The sample is heterogeneous,
and sample homogeneity could not be readily achieved using routine laboratory
practices - Q2; or
E. For batch analysis of
samples to determine potency, a technical laboratory director may check
precision by using either a duplicate laboratory control sample or a duplicate
sample prepared from the marijuana or marijuana product being tested, according
to requirements in A.A.C.
R9-17-404.03(K)(2) and
(3).
F. A technical laboratory director shall
ensure that the reporting units for:
1.
Pesticides, fungicides, growth regulators, heavy metals, or residual solvents
is in parts per million (ppm); and
2. Mycotoxins are according to A.A.C.
R9-17-404.04(I)(4); and
3. Potency
are:
a. In either:
i. Percent (w/w) relative to the bulk plant
material or marijuana product, as applicable; or
ii. Number of milligrams per designated unit;
and
b. For:
i. Total tetrahydrocannabinol, the sum of
tetrahydrocannabinolic acid (THC-A), multiplied by 0.877, and
delta-9-tetrahydrocannabinol ([DELTA]9-THC); and
ii. Total cannabidiol, the sum of
cannabidiolic acid (CBD-A), multiplied by 0.877, and cannabidiol
(CBD).
G. To perform testing for the microbial
contaminants in Table 3.1, a marijuana testing facility shall:
1. Use an applicable method described in
A.A.C. R9-17-404.04(A)(1)
and validated according to A.A.C.
R9-17-404.04(A)(2), and
2. Comply with A.A.C.
R9-17-404.04(A)(3) and (4), as applicable.
H. A technical laboratory director shall
ensure that the marijuana testing facility complies with the requirements in
A.A.C. R9-17-404.04(B) through
(G) when performing testing for the microbial
contaminants in Table 3.1.
I. A
technical laboratory director shall include in the final report of testing for
the microbial contaminants in Table 3.1, according to
R9-18-410(B)(3) and
(C), the following data qualifier notations
if:
1. The limit of quantitation and the
sample results were adjusted to reflect sample dilution - D1;
2. A description of the variance is described
in the final report of testing according to A.A.C.
R9-17-410(B)(3) and
(C) - N1;
3. Sample integrity was not maintained -
Q1;
4. The sample is heterogeneous,
and sample homogeneity could not be readily achieved using routine laboratory
practices - Q2; or
J. A technical laboratory director shall
ensure that: 3.4. Reporting for
mycotoxins includes:
1. The reporting units for
Escherichia coli are colony forming units per gram
(CFU/g);
2. Reporting for
Salmonella is "Detected" or "Not detected" in one gram;
3. Reporting for
Aspergillus is "Detected" or "Not detected" in one gram;
and
a. Total aflatoxins in
units of micrograms per kilogram (µg/kg), and
b. Ochratoxin A in units of micrograms per
kilogram (µg/kg).