Fla. Admin. Code Ann. R. 64E-5.316 - Control of Access to High Radiation Areas
(1) The licensee or registrant shall ensure
that each entrance or access point to a high radiation area has one or more of
the following features:
(a) A control device
that upon entry into the area causes the level of radiation to be reduced below
that level at which an individual might receive a deep dose equivalent of 0.1
rem (1 millisievert) in 1 hour at 30 centimeters from the source of radiation
from any surface that the radiation penetrates;
(b) A control device that energizes a
conspicuous visible or audible signal so that the individual entering the high
radiation area and the supervisor of the activity are made aware of the entry,
or
(c) Entryways that are locked
except during periods when access to the areas is required with positive
control over each individual entry.
(2) The licensee or registrant can substitute
continuous direct or electronic surveillance that is capable of preventing
unauthorized entry in place of the controls required by subsection
64E-5.316(1),
F.A.C., for a high radiation area.
(3) The licensee or registrant can apply to
the department for approval of alternative methods for controlling access to
high radiation areas.
(4) The
licensee or registrant shall establish the controls required by subsections
64E-5.316(1) and
(3), F.A.C., in a way that does not prevent
individuals from leaving a high radiation area.
(5) The licensee is not required to control
each entrance or access point to a room or other area that is a high radiation
area solely because of the presence of radioactive materials prepared for
transport and packaged and labeled as specified in the regulations of the U.S.
Department of Transportation if:
(a) The
packages do not remain in the area longer than 3 days; and,
(b) The dose rate at 1 meter from the
external surface of any package does not exceed 0.01 rem (0.1 millisievert) per
hour.
(6) The licensee is
not required to control entrance or access to rooms or other areas in hospitals
solely because of the presence of patients containing radioactive material if
there are personnel in attendance who are taking the necessary precautions to
prevent the exposure of individuals to radiation or radioactive material in
excess of the established limits in this part and to operate within the ALARA
provisions of the licensee's radiation protection program.
(7) The registrant is not required to control
entrance or access to rooms or other areas containing sources of radiation
capable of producing a high radiation area as described in Rule
64E-5.316, F.A.C., if the
registrant has met all the specific requirements for access and control
specified in other applicable parts of these rules, such as Part IV for
industrial radiographic operations, Part V for x-rays in the healing arts, and
Part VIII for particle accelerators.
Notes
Rulemaking Authority 404.051, 404.081 FS. Law Implemented 404.022, 404.051(1), (4), 404.081 FS.
New 1-1-94, Formerly 10D-91.447.
State regulations are updated quarterly; we currently have two versions available. Below is a comparison between our most recent version and the prior quarterly release. More comparison features will be added as we have more versions to compare.
(1) The licensee or registrant shall ensure that each entrance or access point to a high radiation area has one or more of the following features:
(a) A control device that upon entry into the area causes the level of radiation to be reduced below that level at which an individual might receive a deep dose equivalent of 0.1 rem (1 millisievert) in 1 hour at 30 centimeters from the source of radiation from any surface that the radiation penetrates;
(b) A control device that energizes a conspicuous visible or audible signal so that the individual entering the high radiation area and the supervisor of the activity are made aware of the entry, or
(c) Entryways that are locked except during periods when access to the areas is required with positive control over each individual entry.
(2) The licensee or registrant can substitute continuous direct or electronic surveillance that is capable of preventing unauthorized entry in place of the controls required by subsection 64E-5.316(1), F.A.C., for a high radiation area.
(3) The licensee or registrant can apply to the department for approval of alternative methods for controlling access to high radiation areas.
(4) The licensee or registrant shall establish the controls required by subsections 64E-5.316(1) and (3), F.A.C., in a way that does not prevent individuals from leaving a high radiation area.
(5) The licensee is not required to control each entrance or access point to a room or other area that is a high radiation area solely because of the presence of radioactive materials prepared for transport and packaged and labeled as specified in the regulations of the U.S. Department of Transportation if:
(a) The packages do not remain in the area longer than 3 days; and,
(b) The dose rate at 1 meter from the external surface of any package does not exceed 0.01 rem (0.1 millisievert) per hour.
(6) The licensee is not required to control entrance or access to rooms or other areas in hospitals solely because of the presence of patients containing radioactive material if there are personnel in attendance who are taking the necessary precautions to prevent the exposure of individuals to radiation or radioactive material in excess of the established limits in this part and to operate within the ALARA provisions of the licensee's radiation protection program.
(7) The registrant is not required to control entrance or access to rooms or other areas containing sources of radiation capable of producing a high radiation area as described in Rule 64E-5.316, F.A.C., if the registrant has met all the specific requirements for access and control specified in other applicable parts of these rules, such as Part IV for industrial radiographic operations, Part V for x-rays in the healing arts, and Part VIII for particle accelerators.