Fla. Admin. Code Ann. R. 64E-5.640 - Dosimetry Equipment Used With Remote Afterloading Units, Teletherapy Units, or Gamma Stereotactic Radiosurgery Units
(1) Except for low dose-rate remote
afterloader source output or where the activity is determined by the
manufacturer, a licensee shall have a dosimetry system available for use
calibrated by paragraph (a) or (b), below.
(a) The system must have been calibrated
using a system or source traceable to the NIST and published protocols accepted
by nationally recognized bodies; or by a calibration laboratory accredited by
the AAPM. The calibration must have been performed within the previous 2 years
and after any servicing that may have affected system calibration.
(b) The system shall have been calibrated
within the previous 4 years and shall have been intercompared 18 to 30 months
after the calibration at an intercomparison meeting with another dosimetry
system that has been calibrated within the previous 2 years by the NIST or by a
calibration laboratory accredited by the AAPM. The intercomparison meeting
shall be sanctioned by a calibration laboratory or radiologic physics center
accredited by the AAPM. The calibration factor of the licensee's system shall
not have changed by more than 2 percent. The licensee shall not use the
intercomparison result to change the calibration factor. When intercomparing
dosimetry systems to be used for calibrating sealed sources for therapeutic
units, the licensee shall use a comparable unit with beam attenuators or
collimators, as applicable, and sources of the same radionuclide as the source
used at the licensee's facility.
(2) The licensee shall have available for use
a dosimetry system for spot-check measurements. The spot-check system shall be
the same system used to meet the requirement in subsection
64E-5.640(1),
F.A.C., or shall be a system that has been compared with a system that has been
calibrated as provided in subsection
64E-5.640(1),
F.A.C. This comparison shall have been performed within the previous year and
after each servicing that may have affected system calibration.
(3) The licensee shall maintain a record of
each calibration, intercomparison, and comparison for the duration of the
license. For each calibration, intercomparison, or comparison, the record shall
include:
(a) The date, the manufacturer's
name, model numbers and serial numbers of the instruments that were calibrated,
intercompared, or compared as required by subsections
64E-5.640(1) and
(2), F.A.C.;
(b) The correction factors that were
determined;
(c) The names of the
individuals who performed the calibration, intercomparison, or comparison;
and
Notes
Rulemaking Authority 404.022, 404.051, 404.061, 404.071, 404, 081, 404.141 FS. Law Implemented 404.022, 404.051(1), (4), (5), (6), (8), (9), (10), (11), 404.061(2), (3), 404.071(1), 404.081, 404.141 FS.
New 8-25-91, Formerly 10D-91.759, Amended 2-11-10.
State regulations are updated quarterly; we currently have two versions available. Below is a comparison between our most recent version and the prior quarterly release. More comparison features will be added as we have more versions to compare.
No prior version found.