Fla. Admin. Code Ann. R. 64E-5.641 - Full Calibration Measurements on Teletherapy Units
(1) A licensee authorized to use a
teletherapy unit for medical use shall perform full calibration measurements on
each teletherapy unit:
(a) Before the first
medical use of the unit;
(b) Before
medical use whenever spot-check measurements indicate that the output differs
by more than 5 percent from the output obtained at the last full calibration
corrected mathematically for radioactive decay;
(c) Before medical use following replacement
of the source or following reinstallation of the teletherapy unit in a new
location;
(d) Before medical use
following any repair of the teletherapy unit that includes removal of the
source or major repair of the components associated with the source exposure
assembly; and,
(e) At least every
12 months.
(2) Full
calibration measurements shall include the determination of:
(a) The output within 3 percent for the range
of field sizes and for the distance or range of distances used for medical
use;
(b) The coincidence of the
radiation field and the field indicated by the light beam localizing
device;
(c) The uniformity of the
radiation field and its dependence on the orientation of the useful
beam;
(d) Timer constancy and
linearity over the range of use;
(e) On-off error; and,
(f) The accuracy of all distance measuring
and localization devices in medical use.
(3) A licensee shall use the dosimetry system
described in Rule 64E-5.640, F.A.C., to measure
the output for one set of exposure conditions. The remaining radiation
measurements required in paragraph
64E-5.6411(2)(a),
F.A.C., may be made using a dosimetry system that indicates relative dose
rates.
(4) A licensee shall make
full calibration measurements required by subsection
64E-5.6411(1),
F.A.C., using the manufacturer's published protocols, published protocols as
accepted by nationally recognized bodies or equivalent procedures that have
been submitted to the department. An example of a nationally recognized body is
the American Association of Physicists in Medicine.
(5) A licensee shall correct mathematically
the outputs determined in paragraph
64E-5.6411(2)(a),
F.A.C., for physical decay monthly for cobalt 60 and at least every 6 months
for cesium 137.
(6) Full
calibration measurements required by subsection
64E-5.6411(1),
F.A.C., and physical decay corrections required by subsection
64E-5.6411(5),
F.A.C., shall be performed by the authorized medical physicist.
(7) A licensee shall maintain a record of
each calibration of each teletherapy unit for three years. The record shall
include:
(a) The date of the
calibration;
(b) The manufacturer's
name, model number, and serial number for both the teletherapy unit and the
source;
(c) The model numbers and
serial numbers of the instruments used to calibrate the teletherapy
unit;
(d) The results and an
assessment of the full calibration to include the following:
1. The tables that describe the output of the
unit over the range of field sizes and for the range of distances used in
radiation therapy;
2. A
determination of the coincidence of the radiation field and the field indicated
by the light beam localizing device;
3. The measured timer accuracy for a typical
treatment time;
4. The calculated
on-off error;
5. The estimated
accuracy of each distance measuring or localization device; and,
6. The signature of the authorized medical
physicist.
Notes
Rulemaking Authority 404.022, 404.051, 404.061, 404.071, 404.081, 404.141 FS. Law Implemented 404.022, 404.051(1), (4), (5), (6), (8), (9), (10), (11), 404.061(2), (3), 404.071(1), 404.081, 404.141 FS.
New 8-25-91, Amended 5-12-93, Formerly 10D-91.760, Amended 2-11-10.