(1) A licensee
providing mobile remote afterloader service for medical use shall perform the
following:
(a) Check survey instruments
before medical use at each address of use or on each day of use, whichever is
more frequent; and
(b) Account for
all sources before departure from a client's address of
use.
(2) In addition to
the periodic spot-checks required by Rule
64E-5.6421, F.A.C., a licensee
authorized to use mobile afterloaders for medical use shall perform checks on
each remote afterloader unit before use at each address of use. At a minimum,
checks must be made to verify the operation of the following:
(a) Electrical interlocks on treatment area
access points;
(b) Source exposure
indicator lights on the remote afterloader unit, on the control console, and in
the facility;
(c) Viewing and
intercom systems;
(d) Applicators,
source transfer tubes, and transfer tube-applicator interfaces;
(e) Radiation monitors used to indicate room
exposures;
(f) Source positioning
(accuracy); and
(g) Radiation
monitors used to indicate whether the source has returned to a safe shielded
position.
(3) In addition
to the requirements for checks in subsection
64E-5.6423(2),
F.A.C., a licensee shall ensure overall proper operation of the remote
afterloader unit by conducting a simulated cycle of treatment before use at
each address of use.
(4) If the
results of the checks required in subsection
64E-5.6423(2),
F.A.C., indicate the malfunction of any system, a licensee shall lock the
control console in the off position and not use the unit except as may be
necessary to repair, replace, or check the malfunctioning system.
(5) The licensee shall keep a copy of each
check for mobile remote afterloader unit required by subsection
64E-5.6423(2),
F.A.C., for three years. The records shall include:
(a) The date of the check;
(b) The manufacturer's name, model number,
and serial number of the remote afterloader unit;
(c) Notations accounting for all sources
before the licensee departs from a facility;
(d) Notations indicating the operability of
each entrance door electrical interlock, radiation monitors, source exposure
indicator lights, viewing and intercom system, applicators, source transfer
tubes, and transfer tube applicator interfaces, and source positioning
accuracy; and
(e) The signature of
the individual who performed the check.