Fla. Admin. Code Ann. R. 64E-5.644 - Radiation Surveys for Remote Afterloader and Gamma Stereotactic Radiosurgery Facilities
(1) The licensee
shall perform radiation surveys with an operable radiation survey instrument
calibrated as provided in Rule
64E-5.615, F.A.C., to ensure
that the maximum radiation levels and average radiation levels from the surface
of the main source safe with the source(s) in the shielded position do not
exceed the levels stated in the Sealed Source and Device Registry.
(2) The licensee shall make the survey
specified in subsection
64E-5.644(1),
F.A.C., at the installation of a new source and following repairs to the
source(s) shielding, the source(s) driving unit, or other electronic or
mechanical component that could expose the source, reduce the shielding around
the source(s), or compromise the radiation safety of the unit or the
source(s).
(3) A licensee shall
retain a record of the radiation surveys required by subsection
64E-5.644(1),
F.A.C., for the duration of the license. These records shall include:
(a) The date of the measurements;
(b) The manufacturer's name, model number and
serial number of the treatment unit, source, and instrument used to measure
radiation levels;
(c) Each dose
rate measured around the source while the unit is in the off position and the
average of all measurements; and
(d) The signature of the RSO or authorized
medical physicist who performed the test.