Utah Admin. Code R392-304-18 - Designated Swimming Area Water Quality and Disinfection
(1) The operator shall ensure that the DSA of
an artificial swimming lagoon has sufficient clarity to easily see the deepest
part of the DSA when the water is static. This may be measured by a Secchi disk
or a four-inch by four-inch marker tile.
(2) The operator shall ensure that the DSA of
an artificial swimming lagoon is treated by a disinfectant that complies with
Subsections R392-302-25(1)(a)
through R392-302-25(1)(f),
except that:
(a) the minimum concentrations
for the DSA treatment chemicals are found in Table 2 in this rule rather than
Table 3 in Rule R392-302; and
(b)
the maximum concentration for cyanuric acid described in Subsection
R392-302-25(1)(f)(ii)(B)
is reduced to 50 milligrams per liter.
(3) The operator shall install automated,
manual, or remotely managed chemical feed controllers that comply with
Subsections R392-302-24(1)(a),
R392-302-24(3)(c),
and R392-302-24(4)(b).
(4) The operator shall install an ORP
controller that complies with Subsection
R392-302-24(2)
to monitor the oxidizing potential of the water if an oxidizer is used for
disinfection.
(5) The operator
shall provide a portable poolside water quality test kit that is:
(a) compatible with the disinfectant in
use;
(b) maintained in and stored
according to the manufacturer's instructions; and
(c) protected from extreme heat and cold,
exposure to water, chemicals, or any other element or environment that could
adversely affect the efficacy of the test kit.
(6) The operator may use a remote water
quality testing system, but may not substitute the remote testing system
results for onsite operation and maintenance.
(7)
(a) The
operator shall perform testing at a frequency that verifies the DSA is meeting
the water quality conditions required in Table 2.
(b) At a minimum, tests for disinfectant
levels and pH in each DSA shall be performed before the area is open for bather
use and at least every 24 hours of operation.
(8) The operator shall:
(a) collect a water sample from each DSA at
least once per month;
(b) submit
the sample to a laboratory approved under Rule R444-14, Certification of
Environmental Laboratories, to perform total coliform and heterotrophic plate
count testing; and
(c) ensure that
the collected sample is analyzed in a laboratory for total coliform and
heterotrophic plate count using methods allowed under Section
R444-14-4.
(9) The operator shall review the sample
results to determine if the sample has failed the bacteriological quality
standard as determined by the following sample failure criteria:
(a) the sample contains more than 200 colony
forming units (CFUs) per milliliter, as determined by the heterotrophic plate
count; or
(b) the sample indicates
the presence of coliform bacteria or contains more than one CFU of coliform
bacteria per 100 milliliters.
(10)
(a)
For any test that does not meet the chemical, bacteriological, or physical
parameters described in this section, the operator shall repeat the test within
24 hours of receiving the initial test results.
(b) If the repeat testing shows that the
water continues to be outside of the parameters described in this section, the
manager shall close any DSA to swimming and contact recreation until additional
testing results show that the water quality parameters are within the required
range.
|
TABLE 2 Chemical parameters in each DSA of an ASL |
||
|
Parameter |
Minimum Value |
Maximum Value |
|
Free chlorine |
0.5 ppm |
10.0 ppm |
|
Bromine |
2.5 ppm |
8.0 ppm |
|
Combined chlorine |
None |
0.5 ppm |
|
pH |
7.0 |
7.8 |
|
Cyanuric acid |
None |
50 ppm |
Notes
State regulations are updated quarterly; we currently have two versions available. Below is a comparison between our most recent version and the prior quarterly release. More comparison features will be added as we have more versions to compare.
No prior version found.