Fla. Admin. Code Ann. R. 64E-5.503 - General Requirements for All Diagnostic X-Ray Systems
In addition to other requirements of this part, all diagnostic x-ray systems shall meet the following requirements:
(1) Warning Label. The main control panel and
all auxiliary control panels of the x-ray system shall bear the equivalent
warning statement, legible and accessible to view, "WARNING: This x-ray unit
may be dangerous to patient and operator unless safe exposure factors and
operating instructions are observed."
(2) Battery Charge Indicator. Visual means
shall be provided on the control panel of battery-powered x-ray generators to
indicate whether the battery is in a state of charge adequate for proper
operation.
(3) Leakage Radiation
from the Diagnostic Source Assembly. The leakage radiation from the diagnostic
source assembly measured at a distance of 1 meter in any direction from the
source shall not exceed 100 milliroentgens (25.8 uC per kg) in 1 hour when the
x-ray tube is operated at its leakage technique factors. Compliance shall be
determined by measurements averaged over an area of 100 square centimeters with
no linear dimension greater than 20 centimeters.
(4) Radiation from Components Other than the
Diagnostic Source Assembly. The radiation emitted by a component other than the
diagnostic source assembly shall not exceed 2 milliroentgens (0.516 uC per kg)
in 1 hour at 5 centimeters from any accessible surface of the component when it
is operated in an assembled x-ray system under any conditions for which it was
designed. Compliance shall be determined by measurements averaged over an area
of 100 square centimeters with no linear dimension greater than 20
centimeters.
(5) Beam Quality.
(a) Half-value Layer. The half-value layer
(HVL) of the useful beam for a given x-ray tube potential shall not be less
than the values shown below. If it is necessary to determine such half-value
layer at an x-ray tube potential which is not listed below, linear
interpolation or extrapolation may be made.
Half-value | ||
Design | Measured | layer |
operating range | potential | (Millimeters of |
(Kilovolts peak) | (Kilovolts peak) | aluminum) |
Below 50 | 30 | 0.3 |
40 | 0.4 | |
49 | 0.5 | |
50 to 70 | 50 | 1.2 |
60 | 1.3 | |
70 | 1.5 | |
Above 70 | 71 | 2.1 |
80 | 2.3 | |
90 | 2.5 | |
100 | 2.7 | |
110 | 3.0 | |
120 | 3.2 | |
130 | 3.5 | |
140 | 3.8 | |
150 | 4.1 |
1. The above
HVL criteria will be considered to have been met if it can be demonstrated that
the aluminum equivalent of the total filtration in the primary beam is not less
than that shown below:
Filtration Required vs. Operating Voltage | |
Total Filtration | |
(inherent plus added) | |
Operating | (millimeters |
Voltage | aluminum |
(kVp) | equivalent) |
Below 50 | 0.5 millimeters |
50-70 | 1.5 millimeters |
Above 70 | 2.5 millimeters |
2.
Beryllium window tubes shall have a minimum of 0.5 millimeters aluminum
equivalent filtration permanently mounted in the useful beam.
3. For capacitor energy storage equipment,
compliance shall be determined with the maximum quantity of charge per
exposure.
4. The required minimum
aluminum equivalent filtration shall include the filtration contributed by all
materials which are always present between the focal spot of the tube and the
patient.
5. In addition to the
requirements of (5)(a)1., above, all intraoral dental radiographic systems
manufactured on and after December 1, 1980, shall have a minimum half-value
layer not less than 1.5 millimeters aluminum equivalent filtration permanently
installed in the useful beam.
(b) Filtration Controls. For x-ray systems
which have variable kVp and changeable filters, and which are used for low
filtration techniques, a positive means shall be provided that will prevent an
exposure unless the minimum required amount of filtration is in the useful beam
for the selected kVp.
(6) Aluminum Equivalent of Material between
Patient and Image Receptor. The aluminum equivalent of each of the items listed
below, which are used between the patient and image receptor, shall not exceed
the indicated limits. This requirement is applicable to the front panel of
cassette holders and film changers provided by the manufacturer for purposes of
patient support or to prevent foreign object intrusions. It does not apply to
such items as a screen and its associated mechanical support panel or grids.
Maximum Aluminum | |
Equivalent | |
Item | (millimeters) |
Front panel of cassette holder (total of all) | 1.0 |
Front panel of film changer (total of all) | 1.0 |
Stationary tabletop | 1.0 |
Moveable tabletop (including stationary subtop) | 1.5 |
Cradle | 2.0 |
(7)
Multiple Tube Heads. Where two or more radiographic tubes are controlled by one
exposure switch, the tube or tubes which have been selected shall be clearly
indicated prior to initiation of the exposure. This indication shall be both on
the x-ray control panel and at or near the selected tube housing
assembly.
(8) Mechanical Support of
Tube Head. The tube housing assembly supports shall be adjusted such that the
tube housing assembly will remain stable during an exposure unless the tube
housing movement is a designed function of the x-ray system.
(9) Technique Indicators.
(a) Each x-ray system shall be equipped with
devices, such as labeled control settings or meters, correctly indicating the
physical factors and modes of operation used for exposures. X-ray systems
utilizing arbitrary number or letter designators for kVp, time and
milliamperage shall be accompanied by a chart giving the value of physical
factors for each arbitrary designator.
(b) The technique factors to be used during
an exposure shall be indicated before the exposure begins, except when
automatic exposure controls are used, in which case the technique factors which
are set prior to the exposure shall be indicated.
(c) On equipment having fixed technique
factors, the requirement in (9)(a), above, may be met by permanent markings.
Indication of technique factors shall be visible from the operator's position
except in the case of fluoroscopy or spot films made by the
fluoroscopist.
(d) Reproducible
technique factor indicators may be relabeled to meet the accuracy requirements
of this part. Such relabeling shall be conspicuous and clearly legible and
shall be utilized by the registrant in setting technique factors.
(10) Accuracy of Technique
Factors. Meters, labeled control settings, exposure time selectors and other
physical factor indicators shall be accurate within the following tolerances:
(a) Milliamperage (mA) - Radiographic
± 10%
(b)
Milliamperage (mA) - Fluoroscopic ±0.2 mA
(c) Kilovolt peak (kVp) ±
5 percent
(d) Timer at settings:
1. Greater than 10 seconds
± 1 second
2.
Ten seconds or less ± ten percent
(11) Timer Reproducibility. When
four timer tests are performed at the same timer settings, the average time
period (T) shall be greater than or equal to 12 times
the maximum time period (Tmax) less the minimum time
period (Tmin). T shall be less than or equal to 0.5
seconds. Expressed mathematically, T >= 12
(Tmax - Tmin).
(12) Exposure Reproducibility. The x-ray
exposure produced by radiographic systems shall be reproducible to within the
following criteria: When all technique factors are held constant and four or
more exposures at the same technique factors are made, the value of the average
exposure (E) shall be greater than or equal to 12
times the quantity of maximum exposure (Emax) minus the
minimum exposure (Emin). Expressed mathematically,
E >= 12 (Emax -
Emin).
(13)
Exposure Linearity. The x-ray output produced by radiographic systems utilizing
means other than automatic exposure controls shall be linear to within the
following criteria:
(a) When a choice of two
or more current settings (mA) or current-time product settings (mAs) may be
selected where X1 and X2 are the
average mR per mAs values obtained from four exposures on each of two mA or mAs
settings at a fixed tube potential (kVp) setting, within the range of 40 to 100
percent of the maximum tube rating, the average of four exposures (mR) for a
given milliampere-second (mAs) product in mR per mAs shall not differ by more
than:
1. Five hundredths times the sum of any
2 consecutive mA or mAs averaged settings; expressed mathematically,
|X1 - X2 |<= 0.05 (X1 + X2); and
2. One tenth
times the sum of any other 2 mA or mAs averaged settings; expressed
mathematically,
|X1 - X2 |<= 0.10 (X1 - X2)
(b) Equipment, which after calibration cannot
be made to meet the requirements of (13)(a), above, may be relabeled to
indicate the effective mA or mAs, 8providing that use of such relabeled
stations will meet the requirements of (13)(a), above.
(c) Equipment, which after calibration cannot
be made to meet the requirements of (13)(a) or (b), above, shall not be used
unless written approval is obtained from the Department. Approval shall not be
granted when the linearity determination exceeds the federal standard for
certified systems.
(14)
Automatic Exposure Controls. When automatic exposure control is provided:
(a) Indication shall be made on the control
panel when this mode of operation is selected.
(b) When the x-ray tube potential is greater
than or equal to 50 kVp, the minimum exposure time for field emission equipment
rated for pulsed operation shall be less than or equal to a time interval
equivalent to 2 pulses.
(c) The
minimum exposure time for all equipment other than that specified in (14)(b),
above, shall be less than or equal to 1/60 second or a time interval required
to deliver 5 mAs, whichever is greater.
(d) Either the product of peak x-ray tube
potential, current, and exposure time shall be limited to not more than 60 kWs
per exposure, or the product of x-ray tube current and exposure time shall be
limited to not more than 600 mAs per exposure except when the x-ray tube
potential is less than 50 kVp, in which case the product of x-ray tube current
and exposure time shall be limited to not more than 2,000 mAs per
exposure.
(e) A visible signal
shall indicate when an exposure has been terminated at the limits described in
(14)(d), above, and manual resetting shall be required before further
automatically timed exposures can be made.
(f) Exposure Linearity. When a choice of two
or more tube current settings (mA) may be selected, the average of four
consecutive exposures (E1) made at any one tube current
setting minus the average of four consecutive exposures
(En) made at any other tube current setting shall be
less than or equal to 0.05 times the sum of the two averages. Expressed
mathematically, |E1 - En |<=
0.05 (E1 + En). Measuring
compliance for the above shall be based on the following criteria:
1. An attenuation block as described in
subsection
64E-5.501(5),
F.A.C., shall be in the useful beam.
2. Exposure (mR) shall be measured on the
exit side of the attenuation block.
3. The tube potential (kVp) shall be
maintained at a fixed setting within the range of 40 percent to 100 percent of
the maximum tube rating.
(15) Beam Limiting Devices.
(a) Beam limiting devices capable of
restricting the useful beam to the area of clinical interest shall be used
during exposures.
(b) Beam limiting
devices shall provide a degree of attenuation not less than that required for
the tube housing.
(16)
Remote Exposure Switches. Where an x-ray control is equipped with two or more
remote exposure switches, each remote switch shall serve a single x-ray tube,
and exposures with any tube shall be possible only by the remote switch with
which that particular tube is associated.
(17) Electrical Power Supply. The electrical
power supply and service lines to x-ray systems shall be of sufficient capacity
to permit operation without significant variation in voltage or machine
output.