Fla. Admin. Code Ann. R. 64E-5.647 - Five Year Inspection for Teletherapy and Gamma Stereotactic Radiosurgery Units
(1) A licensee
shall have each teletherapy unit and gamma stereotactic radiosurgery unit fully
inspected and serviced during source replacement or at least every 5 years,
whichever comes first, to assure proper functioning of the source exposure
mechanism.
(2) This inspection and
servicing shall only be performed by persons specifically licensed to do so by
the department, an agreement state, or the U.S. Nuclear Regulatory
Commission.
(3) A licensee shall
maintain a record of the inspection and servicing for the duration of the
license. The record shall contain:
(a) The
inspector's name;
(b) The
inspector's radioactive materials license number;
(c) The date of inspection;
(d) The manufacturer's name and model number
and serial number for both the treatment unit and source;
(e) A list of components inspected;
(f) A list of components serviced and the
type of service;
(g) A list of
components replaced; and
(h) The
signature of the inspector.